High precision environmental simulation chamber
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Application
HVAC and Indoor Airflow Research: Study indoor airflow patterns, temperature distribution, and pollutant concentration under different ventilation strategies for research and teaching purposes.
Human Thermal Comfort Testing: Simulate real indoor conditions with various airflow organizations to compare thermal comfort under different ventilation scenarios.
Electronics and Component Reliability: Evaluate products, parts, and materials in real airflow conditions to assess operational reliability and performance.
Medical and Military Studies: Research cross-infection and aerosol transmission under different air circulation conditions within controlled indoor environments.
Standards
Internal compliance with laboratory safety, electrical, and environmental simulation standards.
Air exchange rates, temperature, and humidity control meet industrial and research-grade environmental chamber specifications.
Custom configurations may comply with client-specific ventilation or airflow standards.
Parameters
| No. | Item | Technical Specifications |
|---|---|---|
| 1 | Chamber Dimensions | External: 5.1 m (L) × 5.6 m (W) × 3.2 m (H) Internal: 4.9 m (L) × 5.4 m (W) × 2.6 m (H) Standard insulated door: 0.8 m × 2.0 m with double-glass observation window Internal lighting: 1 daylight lamp; power outlet: 2.5 kW Digital display panel for temperature and humidity Optional additional window available at extra cost |
| 2 | Temperature Range | 10–40°C ±0.5°C adjustable |
| 3 | Humidity Range | 35–90% ±5% adjustable |
| 4 | Air Change Rate | 3–16 ACH; optional single-pass fresh air supply available (installation-dependent, extra cost) |
| 5 | Electrical Control System | Centralized control panel; Supply: AC 380V/50Hz, max power ≤28.5 kW; Equipped with short-circuit and over-temperature protection |
| 6 | Airflow Organization Modes | 9 modes: Mixed Ventilation:(1) Top diffuser with side-lower return, (2) Side supply-side return, (3) Side supply-top return Displacement Ventilation:(1) Top return, (2) Side-top return Floor Ventilation:(1) Top return, (2) Side-top return Stratified Ventilation:(1) Top return, (2) Side-lower return Optional quick-connect supply fittings available (extra cost) |
| 7 | Working Environment | Standard laboratory conditions |
| 8 | Other Requirements | Refrigeration unit piping and electrical lines default 5 m; site preparation including leveling, wall openings, rain shelter, water supply, drainage, and main power line to control box to be provided by user; chamber panels: 100 mm PU double-color steel |
| 9 | Special Notes | Default specifications are factory standards; custom requirements exceeding these parameters require additional consultation and pricing |
Accessories
Digital display panel for temperature and humidity monitoring.
Standard insulated chamber door with double-glass observation window.
Internal lighting (daylight lamp).
Optional quick-connect supply fittings for airflow customization.
Test Procedures
Chamber Preparation: Confirm power, water, and drainage connections; ensure refrigeration system is operational (~10 min).
Set Environmental Conditions: Program desired temperature (10–40°C), humidity (35–90%), and airflow mode via control panel or software (~5 min).
Load Samples/Subjects: Place equipment, test materials, or participants inside chamber; ensure proper positioning (~10 min).
Run Test: Activate test program; monitor environmental conditions in real-time; adjust airflow mode if needed (~variable duration).
Data Recording and Export: Record chamber temperature, humidity, and airflow data; save and export for analysis (~5 min).
Completion and Retrieval: End program; allow chamber to stabilize at safe conditions; remove samples or subjects safely (~10 min).
Maintenance Information
Regularly clean observation window, internal lighting, and control panel.
Inspect refrigeration unit, piping, and electrical connections for wear or damage.
Verify airflow systems and sensors for calibration and accuracy.
Maintain chamber in a clean, dust-free, and temperature-stable environment.
在线联系供应商
Other supplier products
| ASTM needle flame tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The needle flame tester is widely used in various fields, including but not li... | |
| Affordable flame spread testing machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The Flame Spread Calorimeter is suitable for: Building materials: Evaluate fl... | |
| Catheter Vacuum Leak Tester | The Catheter Vacuum Leak Tester is a professional device used to detect leaks in medical catheter systems under vacuum conditions. Its primary purp... | |
| Conveyor Belt Drum Friction Testing Machine | Main Purposes of Conveyor Belt Drum Friction Testing Machine The conveyor belt drum friction testing machine has extensive application value in mu... | |
| Mask Visual Field Tester | pplication Mask Visual Field Tester to determine whether field of vision under the mask meets the standard requirement or not. Standards GB 2890... |
Same products
| Tabletop Xenon Arc Weathering Tester (SunMaster) | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Water-Cooled Xenon Lamp Weat... | |
| Blow-By Test Rig | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test system for the characterisa... | |
| Air filter System(ISO 16890/EN 779) | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Filter test system for general a... | |
| Oil Thread Leak Test System | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test System for oil thread leak ... | |
| Manual HEPA/ULPA Filter Scanning Test System | 卖方: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Manual HEPA/ULPA Filter Scan... |