Advanced catheter power injection flow and pressure tester equipment
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
The catheter flow and pressure tester for power injection is an important medical device, mainly used to detect key parameters such as flow and pressure of medical catheters (especially micro - catheters) during power injection. These parameters are crucial for evaluating the safety, reliability, and therapeutic effects of catheters. The tester can simulate clinical usage scenarios, conduct pressure resistance tests and flow measurements on catheters, ensuring that catheters will not rupture or leak under high - pressure injection, while maintaining the stability and accuracy of the flow.
Standards
YY 0285.1 - 2017 Disposable Sterile Intravascular Catheters, etc.
Technical Parameters
| Technical Parameter | Specification |
|---|---|
| Test Pressure | Generally adjustable within 0 - 10MPa, with an error not exceeding ±2%. Some devices may have a smaller pressure range, such as 0 - 28MPa. |
| Test Flow | Generally 0 - 500ml, with an error not exceeding ±2% |
| Water Bath Temperature | Usually 37°C, with an error not exceeding ±1°C |
| Data Storage | It can store multiple groups of test data, such as 200 groups or 500 groups, etc. |
| Accuracy | High - precision sensors and data acquisition systems ensure the accuracy of test results |
Accessoriess
Main unit, printer, temperature control device, balance
Test Procedures
1. Operating Procedure
Sample Preparation: Select a clean, undamaged catheter and connect it to the testing instrument interface.
Parameter Setting: Input parameters such as pressure, flow rate, and temperature according to testing requirements. For example, when testing a central venous catheter, set the pressure to 5 MPa, the flow rate to 300 ml/min, and the temperature to 37°C.
Test Start: Start the test via the touchscreen. The device automatically simulates the injection process and records data.
Result Analysis: Generates a flow-pressure curve to determine if the catheter performance meets the standards. For example, if the curve fluctuation exceeds the set range, retesting or adjustment of the catheter design is required.
2. Maintenance Points
Regular Calibration: Calibrate the pressure and flow sensors every 6 months to ensure testing accuracy.
Cleaning and Maintenance: Keep the testing instrument's water tank and tubing clean to avoid residual liquid affecting test results.
Lubrication of Transmission Devices: Regularly lubricate the pump and valves to reduce wear and extend equipment life.
在线联系供应商
Other supplier products
| Mask Air Tightness Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com TheAir TightnessTesterismainly used for air tightness testing of masks and respirators unde... | |
| Multi-function thermal conductivity testing platform for materials | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The Multi-Function Thermal Conductivity Measurement Platform is suitable for t... | |
| Anti-siphon machine for shoelace manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications (1)Testing waterproof leather and synthetic leather used in sho... | |
| Athletic Footwear Shock Attenuating Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard SATRA TM142 ASTM F1614 ASTM F1976 Product Advantages 1. Uncompromising Precis... | |
| Carpet Peel & Tuft Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Details For tufted carpets or traditional carpets, an important parameter is the r... |
Same products
| MMA-120 Handheld Arc Welder | 卖方: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |