Custom Squirrel Cage Smoke Toxicity Testing Equipment for R&D

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The test apparatus complies with the following national and international standards:

(1) GB/T 20285-2006 – Classification for Toxic Hazard of Smoke Generated by Materials

(2) ISO 19700:2006 – Controlled Equivalence Ratio Method for the Determination of Hazardous Components of Fire Effluents

(3) DIN 53436 – Testing of Thermal Decomposition Products of Materials under Ventilated Conditions and Their Toxicological Evaluation

Parameters

ItemSpecification
Quartz Tube & Quartz Boat Made of quartz glass
Quartz Tube Nominal Diameter (36 ± 1) mm
Quartz Tube Wall Thickness (2 ± 0.5) mm
Quartz Tube Length 1000–1300 mm
Quartz Tube Support Adjustable horizontally and vertically
Smoke Collection & Distribution Assembly Three-way stopcock, dilution gas inlet pipe, gas distribution elbow
Gas Flow Pipe Nominal Diameter (36 ± 1) mm
Mouse Rotating Cage Aluminum plate stamping
Rotating Cage Mass (60 ± 10) g
Exposure Chamber Transparent organic glass
Exposure Capacity Up to 10 mice simultaneously
Temperature Control System Thermocouple, cold-junction compensation, temperature controller, digital display
Furnace Movement Drive Stepper motor
Movement Speed (10 ± 0.1) mm/min
Adjustable Speed Range 5–50 mm/min
Maximum Travel Distance > 600 mm
Carrier & Dilution Gas Compressed air
Air Flow Rate 0–20 L/min adjustable

Features

Designed specifically according to GB/T 20285-2006 for smoke toxicity hazard classification of materials.

Ring furnace adopts an energy-saving full-fiber furnace tube structure with stable thermal performance.

High-precision temperature control system ensures furnace temperature meets the requirements of Chapter 7 of GB/T 20285.

Stepper-motor-driven furnace displacement system provides low noise, high accuracy, and flexible speed selection.

Quartz tube and joints are precision ground and paired during processing to ensure excellent sealing performance.

Waste gases generated during testing are directly discharged into the ventilation duct through plastic hoses, preventing smoke leakage in the working area.

Transparent exposure chamber allows visual observation of test animals during exposure.

Mouse rotating cages are designed to rotate freely, preventing fixed postures and reducing the risk of mechanical injury.

Integrated carrier gas and dilution gas supply system ensures stable smoke concentration control.

Computer-based control system enables high automation and real-time monitoring of test parameters.

Accessories

(1) Ring furnace heating system

(2) Quartz tube

(3) Quartz boat

(4) Smoke collection and distribution assembly

(5) Transparent organic glass exposure chamber

(6) Mouse rotating cages

(7) Temperature control and furnace displacement system

(8) Carrier and dilution gas supply components

(9) Computer control software system

Test Procedures

Specimen Installation

Prepared specimens are placed into the quartz boat and positioned at the specified location inside the quartz tube.

Animal Placement

Experimental mice are placed into the aluminum rotating cages, ensuring free rotation without fixed stationary points to avoid injury or restricted movement.

System Startup

The industrial computer and control software are started, and a preset test program is selected.

The ring furnace is heated to the required temperature, and once stable, the furnace begins uniform scanning movement to heat the specimen.

Exposure and Monitoring

Smoke is diluted by carrier air and introduced into the exposure chamber, where mice are exposed for 30 minutes.

The software displays temperature curves, airflow rate, and furnace displacement in real time while animal behavior is observed.

Data Collection

Smoke concentration and related parameters are automatically recorded.

After the test, the system generates and prints test data, including time-temperature curves and mouse survival information.

Maintenance Information

Regular inspection of the quartz tube, sealing joints, and gas pipelines is recommended to maintain proper sealing and safe operation. The temperature control system and furnace displacement mechanism should be periodically checked to ensure accuracy and smooth movement. After testing, the exposure chamber and quartz components should be cleaned to prevent residue accumulation. Software data should be backed up regularly, and sensor calibration should follow laboratory quality management requirements.


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Textile snagging test equipment ICI Mace method
Textile snagging test equipment ICI Mace method Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The ICI Mace Snag Tester is a laboratory device used to evaluate the tendency...
G678W-L cold resistance testing machine for shoes
G678W-L cold resistance testing machine for shoes Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction: Shoes cold resistance testing machine is used to test finished shoes...
Advanced Automatic Liquid Dosing System for Fabric Testing
Advanced Automatic Liquid Dosing System for Fabric Testing Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Automatic Liquid Dosing System for Textile Color Fastness is designed for...
Stapler Blade Sharpness Tester
Stapler Blade Sharpness Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The sharpness tester for stapler blades is mainly used to assess the sharpness...
Solar Simulator Tester
Solar Simulator Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Recommended related instruments: Solar Simulators The Solar Simulator is a brand-new solar...
供应产品

Same products

MMA-120 Handheld Arc Welder
MMA-120 Handheld Arc Welder 卖方: 871183 The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con...
SHX Series Circulating Fluidized Bed Steam Boiler
SHX Series Circulating Fluidized Bed Steam Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi...
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ...
DHL Series Coal-Fired Steam Boilers
DHL Series Coal-Fired Steam Boilers 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b...
SZL Series Biomass Steam Boiler
SZL Series Biomass Steam Boiler 卖方: Hebei Genjux Boiler Co.,Ltd Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi...