Medical needle toughness tester
Get a quote:
Whats/App: +86 185 0176 3615
Web: qinsun-lab.com
The Medical Needle Toughness Tester is designed to evaluate the mechanical strength and flexibility of medical needles under bending stress. It features a 5.7-inch LCD touch display (120 × 90 mm) and supports testing of three types of needle walls: normal, thin, and ultra-thin. The device is manufactured in compliance with GB18457 (“Stainless steel needle for medical equipment manufacturing”), GB15811 (“Sterile disposable injection needle”), and ISO 9626 standards.
This tester is ideal for R&D, quality control, and production verification of disposable medical needles.
Application
(1) Medical disposable injection needles – evaluating bending durability before packaging.
(2) Insulin pen needles – testing ultra-thin walls for high flexibility without fracture.
(3) Vacutainer blood collection needles – ensuring mechanical reliability during insertion.
(4) IV infusion needles and cannulas – verifying structural integrity under repeated bending stress.
(5) Specialty surgical needles – quality assurance for orthopedic or microsurgical applications.
Standards
The device meets and supports testing according to the following international and national standards:
(1) GB15811 – Sterile disposable injection needle
(2) GB18457 – Stainless steel needle for medical equipment manufacturing
(3) ISO 9626 – Stainless steel needle for medical use
Features
(1) High-precision touch LCD interface for intuitive operation.
(2) Adjustable bending angles for normal, thin, and ultra-thin needle walls.
(3) Automatic bi-directional bending for standardized toughness testing.
(4) High-brightness 4-digit LED display for accurate parameter monitoring.
(5) Supports multiple syringe diameters (0.2–3.4 mm).
(6) Compact and robust design with net weight of 10 kg.
(7) Free perpetual upgrades according to national standards.
(8) Complete accessory kit included for immediate testing.
Test Procedure
(1) Fix one end of the needle tube firmly in the jig.
(2) Adjust the span according to needle wall type:
Normal wall: 25° bending
Thin wall: 20° bending
Ultra-thin wall: 15° bending
(3) Apply force at the specified span position, bi-directionally, 20 times at a frequency of 0.5 Hz.
(4) Observe the needle for breakage, deformation, or failure.
Principle: One end of the needle is fixed. A force is applied from the fixed point to the prescribed span, bending in one direction and then in the opposite direction, repeated for the defined number of cycles.
Technical Parameters
| Parameter | Specification |
|---|---|
| Syringe nominal specifications | 0.2 – 3.4 mm |
| Test frequency | 0.5 Hz |
| Rocker arm bending angle | 15°, 20°, 25° (switch selectable, displayed on high-brightness 4-digit LED) |
| Bending distance | 6 – 95 mm, accuracy ±0.1 mm (10 selectable spans) |
| Number of experiments | Bi-directional force 20 times |
| Net weight | 10 kg |
| Overall dimensions (L × W × H) | 300 × 240 × 240 mm |
| Standards | GB15811, GB18457, ISO 9626 |
| Control & Display | 5.7-inch LCD touch menu (120 × 90 mm) |
| Upgrade | Free perpetual upgrade according to national standards |
| Power supply | AC 220 V, 50 Hz |
| Part List | Host (1 set), Printing paper (1 roll), Fixture (1 set), Weight (1 set), Operating Instructions |
在线联系供应商
Other supplier products
| Cork Puncture and Chip Detection Instrument | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The main purpose of the rubber stopper puncture debris detector is to evaluate the... | |
| Automated Steam Quality Analysis System V2 | Technical Parameters ParameterSpecification Model STD-100-V2 Water Flow Rate 2.7 ± 0.1 L/min Steam Dynamic Pressure ... | |
| Water-cooled Xenon Lamp Aging Test Chamber - lab instrument | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Email: serving@qinsun-lab。com Introduction: SI4000Xenon Arc Weather Ometer utili... | |
| 8 Ram Static Load and Pull Tester | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com 1. Product Function and Application The 8 Ram Static Load and Pull Tester is desi... | |
| Smoke Density Chamber | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Email: serving@qinsun-lab。com The main principle is to measure smoke density in a... |
Same products
| Benchtop Thermal Conductivity Tester for Insulation Materials | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument complies with a... | |
| High-Accuracy Guarded Heat Flow Meter Thermal Conductivity Measurement System | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument is designed and... | |
| Heat Flow Method for Thermal Conductivity Measurement | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument complies with a... | |
| Benchtop Thermal Conductivity Tester | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument is designed and... | |
| Guarded Heat Flow Measurement System | 卖方: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASTM E1530: Standard Test Meth... |