Laboratory AC/DC Wire Temperature and Voltage Drop Tester Manufacturer
Get a quote:
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Features
(1) Dual AC/DC testing capability with precise current and voltage control.
(2) Real-time recording of conductor temperature, current, and voltage drop.
(3) Cyclic testing function for evaluating long-term reliability.
(4) Thermocouple-based multi-channel temperature monitoring.
(5) Environmental chamber with precise temperature control and data communication.
(6) Software-enabled curve generation for current vs. temperature and temperature vs. time.
(7) Automatic cut-off at preset time for safety and test repeatability.
(8) Excel-compatible report generation for documentation and analysis.
Technical Parameters
| Parameter | Specification |
|---|---|
| Input Voltage | AC 380V ±10%, 50Hz |
| Load Power Supply | DC: Switching power supply; AC: Constant current source |
| Operating Current | AC: 0–1000 A, open-circuit voltage ≤10 V; DC: 10–1000 A, open-circuit voltage ≤10 V |
| Current Output Control | Computer programmable |
| Current Meter Accuracy | ±1% + 3 digits |
| DC Voltage Drop Measurement | 0.01 mV–199.99 mV / 0.1 mV–1999.9 mV (two ranges) |
| Timing Range | 0–9999.99 H/M/S, automatic cut-off at set time |
| Temperature Measurement | 8-channel standard, T1C, T2C, T3C, T1S, and ambient; Type-T 30AWG thermocouples |
| Temperature Control | Environmental chamber with 485 communication for monitoring and control |
| Chamber Effective Space | 3000 × 500 × 500 mm |
| Chamber Door | Top-hinged lever structure |
| Built-in Connection | Copper busbars for sample connection |
| Communication Interface | 485 interface, Excel report generation, storage, and printout |
| Software | Data acquisition software for AC/DC testing, recording current, voltage, and temperature curves |
| Computer | Brand PC included |
Accessories
(1) Environmental chamber with copper busbar connections.
(2) Thermocouple sensors (T-type, 30AWG) for multi-point temperature monitoring.
(3) Brand PC with pre-installed test software.
(4) Data acquisition and analysis software for AC/DC current and temperature curves.
(5) Communication cable for 485 interface.
Maintenance Information
(1) Calibrate current sensors and thermocouples regularly.
(2) Inspect copper busbars and connection clamps to ensure reliable contact.
(3) Keep chamber clean and free of moisture to maintain accurate temperature control.
(4) Periodically check the software for updates and backups of test protocols.
(5) Verify AC/DC source output stability before each testing session.
在线联系供应商
Other supplier products
| Heated plate drying rate tester with digital display | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards AATCC 201-2013 Fabric Drying Rate: Hot Plate Method FZ/T 01176-2024 Textiles - ... | |
| Electric Toothbrush Air Tightness Testing Equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Electric Toothbrush Air Tightness Testing Equipment is a dedicated instrument used to e... | |
| Food Sachet Pressure Tester | ISO 11607 Flexible Packaging Test Equipment | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The pressure resistance tester for sauce packets is a device specifically des... | |
| Thermostat testing device | Whats/App: +86 156 1874 6768 Web: standard-groups.com TheThermostat Testing Deviceis a dedicated system for life cycle testing of snap-action the... | |
| Wet leakage current test system of photovoltaic module | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Wet Leakage Current Test System is designed for high-precision evaluation of insulation... |
Same products
| MMA-120 Handheld Arc Welder | 卖方: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | 卖方: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |