Dry microbial penetration resistance tester
DryMicrobial Penetration Resistance Testeris designed to determine determine the penetration resistance of material to bacterial on dry particulates at a size range of human dandruff.
The resistance to dry microbial penetration test (ISO 22612) evaluates resistance characteristics of protective equipment to dry bacteria-carrying particles.Testing is most often performed on equipment such as surgical gowns, drapes, and clean air suits when used as medical devices. This test is part of a test suite required by the EN 13795 guidance, which determines performance requirements and performance levels for gowns, drapes, and clean air suits when used as medical devices. It is also included in the test suite required by EN 14126, for personal protective equipment. Testing in compliance to EN 13795 is part of following the European Union Medical Device Directive 93/42/EEC, and is important when seeking a CE mark for products marketed in Europe. Nelson Laboratories personnel are experts in barrier testing, and we offer a one-stop shop for barrier tests required for marketing gowns and drapes in both the US and Europe.
Standards:
ISO 22612; YY / T 0506.5-2009
Test principle:
When testing, the samples were seperately fixed on a container.In these containers, five containers carrying the bacillus subtilis and 1 container carrying uncontaminated talc powder used as control.Insert 1 petri dish at the bottom of each container and beneathe the bottom of the specimen.
The supporting vessel's equipment is oscillated by a gas ball type oscillator, and all of the talcum powder passing through the sample falls onto the petri dish. Remove the petri dish for culture. The experiment results were evaluated by counting the colonies.
Scope of application:
Dry microbial penetration resistance performance tester is professionaly used to determine the bacterial penetration resistance performance of materials to dry particulates in the size range of human body dandruff.Whenn testing, the samples were fixed on a container respectively.In these containers, five containers containing subtilisin talc, one container added with Not stained talcum powder was used as a control.A Petri dish was inserted at the bottom of each container at a distance from the bottom of the timer.
The equipment supporting the container is oscillated by a Gas ball oscillator, and all the talc penetrating the sample Falled on the culture dish , cultured without the culture vessel.The test results were evaluated by counting the growing colonies.
Main parameters:
Vibration form: pneumatic ball vibrator
Vibration frequency: 20800 times / min
Vibration force: 650N
Worktable size: 40cm × 40cm × 10mm
Workbench media: marble plate
Experimental container: stainless steel test container
Work stations: 6
Cabinet negative pressure range: -50 Pa ~ -200 Pa
High efficiency filter filtration efficiency: better than 99.99%
Negative pressure cabinet ventilation flow: ≥ 5m3 / min
Data storage capacity: 100,000 group
Working voltage: 220V 50HZ
Features:
Reliable design, stable test;
Front switch glass doors, easy to observe the operation of the experimental staff;
Removable bracket, height adjustable bracket;
Support, mobile dual casters;
Negative pressure experimental system, with the fan exhaust system and the introduction of high efficiency filter to ensure the safety of operators;
Embedded high - speed industrial microcomputer control;
Dedicated operating software, with software parameters calibration, fault detection automatically
Vibration form: pneumatic ball vibrator
Vibration frequency: 20800 times / min
Vibration force: 650N
Worktable size: 40cm × 40cm × 10mm
Workbench media: marble plate
Experimental container: stainless steel test container
Work stations: 6
Cabinet negative pressure range: -50 Pa ~ -200 Pa
High efficiency filter filtration efficiency: better than 99.99%
Negative pressure cabinet ventilation flow: ≥ 5m3 / min
Data storage capacity: 100,000 group
Working voltage: 220V 50HZ
Features:
Reliable design, stable test;
Front switch glass doors, easy to observe the operation of the experimental staff;
Removable bracket, height adjustable bracket;
Support, mobile dual casters;
Negative pressure experimental system, with the fan exhaust system and the introduction of high efficiency filter to ensure the safety of operators;
Embedded high - speed industrial microcomputer control;
Dedicated operating software, with software parameters calibration, fault detection automatically

在线联系供应商
Other supplier products
| Electronic Twist Tester | Electric Twist Tester Standards: ASTM D1422/1423,BS 2085,DIN 53832,ISO 2061,IWTO 25-70,GB/T 2543.1-2001/2543.2-2001 Applications: TheElectric Twi... | |
| rain testing machine | TRANSLATE with x English Arabic Hebrew Polish Bulgarian Hindi Portuguese Catalan Hmong Daw Romanian ... | |
| Horizontal flame test chamber | The Horizontal Flammability Tester is designed for determining the comparative burn rates and burn resistance of textiles, particularly those for a... | |
| Alcohol abrasion resistance tester | Brief Introduction: The test machinecomplies withUL817 and other relevant standards "special rubber fixed to the machine friction rub hammer surfac... | |
| 纺织品气流阻力测试仪 | Textile Airflow resistance measurementApplications:Quality control in production process.Testing in research and development.Fabric Airflow resista... |
Same products
| PH-2398 pH/EC/ORP/温度 多参数测试仪 | 卖方: 科立龙电子股份有限公司 | 大显示,可同时进行水质PH、氧化还原电位、电导率、温度的测量; 带背光LCD显示清晰,功能齐全,精度高; 仪表采用微电脑测量与控制技术,具有自动校准功能,方便实用。 内置蓝牙,可以与电脑或... | |
| ORP-2026S 防水型 土壤ORP&温度 二合一测试仪 | 卖方: 科立龙电子股份有限公司 | 范围 ORP -999~999mV Temp 0~50℃ 分辨率 ORP 1mV... | |
| 防水型 EC /TDS/温度三合一参数测试仪 | 卖方: 科立龙电子股份有限公司 | 范围 EC ~19.9mS TDS Temp 0~50℃ 分辨率 ... | |
| EC-98330 电导率营养测试棒 | 卖方: 科立龙电子股份有限公司 | 本营养测试仪适用于水培和农业应用,能够准确测量水中的营养液电导率(EC)单位为mS/cm,以及总溶解固体(TDS)转换为百万分之一(ppm)。 适合需要较高EC值的植物,该测试仪具有高达6.0... | |
| PH-808防水笔式土壤PH计 | 卖方: 科立龙电子股份有限公司 | 两点按键校准,自动识别pH标准缓冲液 自动温度补偿确保全量程范围的精确测量。 自动电源关闭,有效节省电池能耗。 可替换型电极降低仪表维护与使用成 全防水性能保障在恶劣环境中对仪表的完整保... |