Suture Needle Cutting Force Tester
The Suture Needle Cutting Force Tester is a precision instrument specifically used to detect the force required by medical suture needles during the cutting process.
Applications
Evaluating Cutting Performance: By measuring the cutting force required for suture needles to penetrate simulated tissues, the cutting performance and stability of suture needles are evaluated. This is crucial for ensuring the effectiveness and safety of suture needles in actual surgeries. Quality Control: In the production process of medical devices, as a key tool for quality control, it is used to detect whether the cutting force of suture needles meets relevant standards and regulations.
R & D Support: During the R & D process of medical suture needles, the tester can provide key data support, helping researchers understand the impact of different materials, designs, and processes on the cutting force of suture needles, thus optimizing product performance. Product Certification: It provides the data support needed for product certification for manufacturers of medical suture needles, ensuring that products comply with relevant international or domestic standards and regulations.
Standards
YY/T 0043-2016:Medical suture needle
YY 0166-2002:Surgical suture needles with thread
Features
1. High Precision and High Repeatability
The testing instrument is equipped with high-precision sensors that can accurately capture force changes at the millinewton level. Combined with an advanced control system that precisely adjusts testing parameters, it ensures consistency of test data across multiple tests and different scenarios. This provides reliable data support for enterprise product quality control (eliminating defective products, reducing quality risks, and ensuring patient safety) and R&D optimization (scientific analysis and improved product performance).
2. Strong Flexibility
The testing parameters (speed, force) and clamp gripping force can be flexibly adjusted according to different specifications and designs of surgical needles, such as triangular and round needles. This allows for accurate evaluation of the cutting performance of various surgical needles while avoiding damage to the needle body. It is compatible with testing both conventional and new special surgical needles, assisting enterprises in efficient quality inspection and new product development.
3. User-Friendly Human-Machine Interface
It features an intuitive touchscreen/computer control interface, making operation simple and easy to understand; it supports automatic generation of standardized reports containing key data (puncture force, cutting force, etc.) and trend charts, saving labor and reducing errors. Some advanced models also support real-time data monitoring and remote transmission, facilitating timely problem handling and cross-regional data sharing.
Technical Parameters
| Serial No. | Parameter Item | Parameter Details |
|---|---|---|
| 1 | Operation Interface | Offers simplified Chinese/English selection; user-friendly and easy to operate |
| 2 | Measurement Range | 0 - 80N, with an error not exceeding ±0.1N (ensures measurement accuracy) |
| 3 | Testing Speed | ≤10mm/s (freely settable), simulating cutting speed in actual surgeries |
| 4 | Resolution | 0.01N (supports high-precision measurement for fine evaluation of suture needle cutting force) |
| 5 | Sensor | High-precision force value sensor (ensures measurement stability and reliability) |
| 6 | Display Screen | 7-inch high-resolution touch screen (directly displays test data and curves) |
| 7 | Programmable Controller | PLC/ARM (supports multiple testing modes and data processing functions) |
| 8 | External Dimensions | 430×290×250mm (specific dimensions may vary by model) |
| 9 | Power Supply | AC90V - 240V/50Hz (adaptive wide voltage, convenient for use across countries/regions) |
Accessoriess
Main unit × 1
Testing software × 1 (PC version)
Instruction manual / Operation manual × 1
Certificate of conformity / Warranty card × 1
Power cord × 1 (AC 90V-240V/50Hz adaptive)
Matching fixtures × 1 set (including various specifications)
Standard weights × 1 set (for calibration)
Printer paper × 1 roll
Suture thread × 1 roll (usually 4-0 or 3-0 gauge)
Wrench set × 1 (for fixture adjustment)
Test Procedures
1. Preparation Before Operation
Equipment and Environment Check:
Connect the power supply (compatible with AC90V-240V/50Hz wide voltage), start the device and confirm that the 7-inch touch screen (Chinese and English dual-language interface) is working properly; place the device on a vibration-free workbench at 22±2℃, ensuring that the testing environment meets the requirements for medical device testing.
Sample and Consumable Preparation:
Prepare surgical needles that comply with the standards of "YY/T 0043 Medical Suturing Needles" and "YY 0166 Suturing Needles with Thread" (no deformation, clean surface), and standard test sutures (such as 4-0 gauge, cut to 10cm length and folded into a U-shape).
2. Instrument Calibration (Ensuring Test Accuracy)
No-load Zero Point Calibration:
Ensure that there are no samples in the fixture, enter the device's "Calibration Interface" and perform zero point calibration, repeating 2-3 times to ensure that the sensor zero point error is ≤±0.1N (matching the instrument's measurement accuracy requirements).
Range Verification (Optional):
Verify the accuracy of the sensor measurement using standard weights (such as 5N, 10N) to ensure that the data meets the 0-80N range requirements.
3. Sample and Test Piece Installation
Fixture Adaptation:
Select the corresponding dedicated fixture according to the type of surgical needle (triangular needle/round needle), ensuring that the fixture matches the needle type.
Surgical Needle Fixing:
Install the surgical needle firmly in the fixture with the blade facing downwards, avoiding needle body shaking; if it is a suturing needle with thread, the root of the suture should be fixed simultaneously.
Test Wire Installation:
Hang the U-shaped test wire on the upper fixture, so that the wire is precisely aligned with the front 1/3 of the surgical needle blade, avoiding contact with other parts of the device.
4. Test Parameter Settings
Enter the touch screen operation interface and select the "Surgical Needle Cutting Force Test" mode;
Set the core parameters: test speed ≤10mm/s (simulating actual surgical cutting speed), number of tests (3-5 times are recommended, taking the average to improve data stability), ensuring that the parameters meet the requirements of the YY series standards;
Save the task after confirming the parameters. 5. Test Execution
Ensure personnel are away from the test area, then click "Start Test";
The instrument automatically drives the fixture to move, and a high-precision force sensor collects cutting force data in real time (including peak cutting force);
The equipment automatically stops after a single test is completed and records the data; repeat the test 3-5 times to obtain multiple sets of valid data.
6. Data Processing and Application
Data Viewing:
The touchscreen displays the peak cutting force, average cutting force, and force-displacement curve, allowing for intuitive assessment of the suture needle's cutting performance stability.
Report Generation:
Export a report containing test data that conforms to the YY series standards, for use in quality control (checking whether the cutting force meets the standards), R&D support (optimizing suture needle design), or product certification (providing compliance data).
Data Archiving:
Save the test report and raw data to complete record keeping.
7. Equipment Maintenance
After the test, turn off the power and clean the fixture and workbench (remove residual thread debris);
Regularly check the wear of the sensor and fixture to ensure the long-term stable operation of the equipment.
在线联系供应商
Other supplier products
| Cigarette Draw Resistance Tester | The Cigarette Draw Resistance Tester is designed to measure the airflow resistance encountered by a cigarette during the smoking process. By simula... | |
| UV Sterilizer | Product introduction of Environmental test chamber Ultraviolet disinfection technology is the latest generation of disinfection technology emerg... | |
| Solid Auto - ignition Point Tester | This tester determines the self - ignition characteristics of solid chemical products by placing the substances in a 20mm - side - length cubic wir... | |
| Hot Tack Tester / Heat Seal Tester | The Hot Tack Tester/Heat Seal Tester is specifically designed for testing the heat-adhesion and heat-sealing properties of packaging materials such... | |
| Taber Linear Abraser | TheTABER Linear Abraser (Abrader) - Model 5750has been engineered to evaluate product’s abrasion resistance, scratch resistance (single or mu... |
Same products
| Air Permeability Tester STD- F8 | 卖方: Standard International Group (HK) Limited | The Air Permeability Tester is designed to quickly and accurately measure the breathability and p... | |
| Surface Performance Tester STD-FB4 - A | 卖方: Standard International Group (HK) Limited | Introduction TheSurface Tester analyzes finger movements-particularly, strokes-performed by arti... | |
| Compression Tester STD - FB3 - A | 卖方: Standard International Group (HK) Limited | TheCompression Tester analyzes hand movements-particularly, pushing with a finger-performed by ar... | |
| Single - Strand Bending Tester STD - FB2 - SH | 卖方: Standard International Group (HK) Limited | Standards ASTM A370. ASTM A416. ISO6934.4 The Single Fiber Bending Tester makes it possible to ... | |
| Bending Tester STD- FB2 - S | 卖方: Standard International Group (HK) Limited | Compliance Standards ASTM D790 (Plastics Flexural Properties) ISO 178 (Flexural Strength of Rig... |