Catheter Leakage and Compliance Tester
The Catheter Leakage and Compliance Tester is a professional testing device used to evaluate the expansion capacity (compliance) and leak resistance of medical tubing (such as infusion lines, vascular catheters, and respiratory tubing) under varying pressure conditions.
Applications
The catheter leakage and compliance tester is mainly used to detect the leakage and compliance of medical catheters (such as the breathing circuits of ventilators and anesthesia machines). Specifically, it can: Leakage Test: By applying a certain pressure inside the catheter and monitoring the rate or extent of pressure drop, it determines whether there is leakage in the catheter. This is crucial for ensuring the sealing and safety of the catheter during medical procedures. Compliance Test: Compliance refers to the ability of the catheter to deform when subjected to external forces. The tester applies a certain pressure inside the catheter and measures the volume change of the catheter to evaluate its compliance. This is of great significance for understanding the flexibility and elasticity of the catheter, as well as its applicability in medical applications.
Standards
ISO 13988-1:2017: Intravascular catheters - Part 1
ISO 8536-4:2016: Medical infusion equipment
YY/T 0285.4-2017: Intravascular catheters - Part 4
YY/T 1551.1-2017: Non-intravascular catheters
Features
Color LCD large screen display: Real-time display of test data and statistical values, intuitive and clear.
Intelligent touch control operation: Simple human-machine interface for quick experiment switching.
High-precision electronic flow meter: Achieves precise flow control, avoiding testing errors.
Top-tier sensor system: High-sensitivity data acquisition ensures stable and consistent test results.
Data security management: Four-level access control and encrypted storage prevent unauthorized operations and ensure the integrity and traceability of results.
Convenient result output: Equipped with a micro printer, supporting instant printing of single test results and statistical data.
Technical Parameters
| Parameter | Specification |
|---|---|
| Pressure Range | Determined by test object & application scenarios:- e.g., 0 - 300kPa for medical breathing circuits- Narrower range (e.g., 0 - 50kPa) for routine compliance testing |
| Pressure Accuracy | Key for test accuracy; generally requires ±0.1% or higher (e.g., 0 - 50kPa pressure sensor can reach one-thousandth accuracy) |
| Flow Range | Covers various flow situations of tested catheters; common flowmeter range: 0 - 100L/min |
| Flow Accuracy | Important indicator; generally requires ±1% or higher (e.g., ±1%FS for 2L/min mass flow controller) |
| Volume Measurement Range and Accuracy | - Range: From a few milliliters to several liters (adapts to different catheter volumes)- Accuracy: Generally needs to reach ±0.1ml or higher (to accurately calculate catheter compliance) |
| Test Time Range | - Leakage tests: Flexible setting (a few seconds to several hours)- Compliance tests: Controls pressure change rate & time interval to capture catheter volume changes at different pressure stages |
Accessoriess
Standard accessories include: specialized connectors for different sizes of medical tubing (such as ventilator/anesthesia machine breathing circuits), pressure/flow test tubing, tubing fixing clamps (to ensure sample stability during testing), standard test samples for equipment calibration (including standard leakage and compliance components), operating manual, and calibration certificate.
Test Procedures
Sample Pretreatment
Tubing Aging Treatment: Immerse in physiological saline at 37°C for 24 hours to relieve material stress;
Connection Sealing Inspection: Apply soapy water to the interfaces to confirm no initial leakage.
Typical Test Modes
Static Compliance Test:
Pressurize to 100kPa, maintain pressure for 30 seconds, and record the volume change ΔV1;
Release pressure to 0kPa, and record the volume recovery ΔV2;
Compliance calculation: C = (ΔV1 - ΔV2)/ΔP (unit: mL/kPa).
Dynamic Leakage Test:
Apply a sinusoidal pressure wave with a frequency of 1Hz and an amplitude of ±50kPa for 10 minutes;
Collect the leaked fluid and calculate the leakage rate using the weighing method (≤0.5mL/10min is considered acceptable).
Result Evaluation Standards (Refer to YY/T 0285.4-2017)
Compliance Deviation: ≤±15% compared to the nominal value;
Leakage: ≤1mL/minute under positive pressure of 100kPa, and no visible bubbles under negative pressure of -50kPa.
在线联系供应商
Other supplier products
| Infusion device Leakage negative pressure tester | The Infusion Device Leakage Negative Pressure Tester is designed to evaluate the airtightness of syringes, infusion sets, and related medical devic... | |
| Medical Face Mask Synthetic Blood Penetration Test machine | Applicable Standards: GB 19083-2010 YY/T 0691-2008 YY 0469-2011 ISO 22609 ASTM F1862-07 Features: 1. Mask fixture is metal human head mold, ... | |
| Mask Bacterial Filtration Efficiency Tester | Medical Masks bacterial filtration efficiency tester also called BFE test machine is to determine the bacterial filtration efficiency (BFE) of medi... | |
| Solid Combustion Rate Tester | Product Introduction This product complies with "Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria" 33.2.1.4 Test... | |
| Oxidizing Solids Tester | Conforms to UN Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria 34.4.1 Test O.1: Test for oxidizing solids, 34.4.... |
Same products
| Air Permeability Tester STD- F8 | 卖方: Standard International Group (HK) Limited | The Air Permeability Tester is designed to quickly and accurately measure the breathability and p... | |
| Surface Performance Tester STD-FB4 - A | 卖方: Standard International Group (HK) Limited | Introduction TheSurface Tester analyzes finger movements-particularly, strokes-performed by arti... | |
| Compression Tester STD - FB3 - A | 卖方: Standard International Group (HK) Limited | TheCompression Tester analyzes hand movements-particularly, pushing with a finger-performed by ar... | |
| Single - Strand Bending Tester STD - FB2 - SH | 卖方: Standard International Group (HK) Limited | Standards ASTM A370. ASTM A416. ISO6934.4 The Single Fiber Bending Tester makes it possible to ... | |
| Bending Tester STD- FB2 - S | 卖方: Standard International Group (HK) Limited | Compliance Standards ASTM D790 (Plastics Flexural Properties) ISO 178 (Flexural Strength of Rig... |