Walk-in High Low Temperature and Low Pressure Test Chamber
Get a quote:
Whats/App: +86 185 0176 3615
Web: qinsun-lab.com
TheWalk-in High/Low Temperature and Low Pressure Test Chamberis a precision environmental simulation instrument designed to replicate high-temperature, low-temperature, and low-pressure composite conditions. You can use it to perform single-parameter or multi-parameter coupled environmental tests on large-scale equipment, ensuring reliability and performance verification under extreme conditions.
Standards
GB/T 10590-2006 Technical Specifications for High/Low Temperature and Low Pressure Test Chambers
GB/T 2423.25 Environmental Testing for Electrical and Electronic Products. Part 2: Test Methods. Test Z/AM: Combined Low Temperature/Low Pressure Test
GB/T 2423.26 Environmental Testing of Electrical and Electronic Products. Part 2: Test Methods. Test Z/BM: Combined High Temperature/Low Pressure Test
GJB 150.2A Laboratory Environmental Test Methods for Military Equipment. Part 2: Low Pressure (Altitude) Test
GJB 367.2 General Technical Conditions for Military Communication Equipment. Environmental Test Methods
Parameters
| Item | Specifications |
|---|---|
| Nominal Internal Volume | 12.1 m³ |
| Chamber Internal Dimensions | 2500 mm (L) × 2300 mm (W) × 2100 mm (H) |
| Chamber Structure | Integral construction required for the equipment chamber |
| Chamber Material | Inner liner made of stainless steel with sandblasted surface treatment |
| Temperature Range | -45°C ~ +130°C (maintainable under both normal pressure and vacuum conditions) |
| Temperature Fluctuation | ≤±0.5°C (under normal pressure) |
| Temperature Uniformity | ≤2°C (under normal pressure) |
| Temperature Deviation | ≤±2°C (under normal pressure) |
| Pressure Range | 101.3 kPa ~ 20 kPa |
| Pressure Change Rate | ≤10 kPa/min (adjustable); from normal pressure to 1 kPa ≤30 min |
| Pressure Recovery Rate | ≤10 kPa/min (adjustable and controllable) |
| Pressure Deviation | ±2 kPa (normal pressure ~ 40 kPa)±5% (40 kPa ~ 4 kPa)±0.1 kPa (4 kPa ~ 1 kPa) |
| Pressure Reduction Time | From normal pressure to 20 kPa ≤30 min |
| Simulated Altitude | ≥10.000 meters |
Send product request
Other supplier products
| Filter Efficiency Testing Equipment EN 1822 ISO 29463 | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com TheFilter Efficiency Testing Equipmentis designed to evaluate filtration efficienc... | |
| Air Flow Resistance Tester | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Application The Air Flow Resistance Tester is suitable for airflow evaluation acr... | |
| Automatic dry cleaning tester | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Automatic dry cleaning testers have wide applications in multiple fields, primaril... | |
| Dry microbial penetration resistance tester | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Dry microbial penetration resistance performance tester is professionaly used to d... | |
| Taber Abrasion Tester (Double-Head) | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Standards DIN Standards: DIN 53754 (plastics), DIN 53109 (paper/board), DIN 52347... |