Building Material Monomer Combustion Tester EN 13823 Manufacturer in China
Get a quote:
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Features:
1. the computer data acquisition system can automatically record the oxygen concentration, concentration of carbon dioxide, smoke density, flue temperature, environment temperature and humidity (data collection interval 3 s), can be automatically calculated the heat release quantity, THR600s, gas production rate index and smoke production rate;
2. The combustion chamber shall be installed with a screw fixing cabinet, which can be disassembled; The thermal insulation exhaust duct, the insulation gas cover and the thermal collector adopt double layer insulation function to ensure the accuracy of data collection.
3. The combustion control system adopts the gas solenoid valve to ensure the safety of the equipment;
4. High degree of automation and easy to learn;
5. System modular design for easy maintenance.
Electronic hardware:
The control core adopts imported omron PLC and temperature expansion module to reduce interference and improve stability.
The output terminal adopts the new solid-state contactless switch device of schneider, with high reliability, low noise and fast switching speed.
The data interface adopts rs-232 standard serial port communication, stable data, high reliability and easy to use.
Test process safety: the combustion control system adopts the gas solenoid valve, and the igniter adopts the automatic ignition method.
Software part:
Friendly software interface, easy to operate.
The software control system includes automatic control and manual control.
Automatic control mode can run automatic control equipment, automatic detection, automatic data collection, automatic display test curve, completion test, at the same time also can automatically generate test results, the automatic generation of test report, etc.
The manual control mode can help the debugging function of the equipment.
Parameters:
| Item | Specification |
|---|---|
| Combustion laboratory size | 3000mm * 3000mm * 2800mm |
| Outline dimensions of smoke pipe in combustion chamber | 5200mm * 3000mm * 1800mm |
| Control box outline dimension | 800mm * 1200mm * 1500mm |
| Smoke extraction speed | 0.50m³/s ~ 0.65m³/s |
| Temperature measurement device | Imported K-type armored thermocouple, 3mm diameter |
| Temperature measurement precision | ±0.5℃ |
| Accuracy of pressure sensor | (Not specified) |
| Test time | 1560s; Time recording system accuracy: 0.1s |
| Oxygen analysis (paramagnetism) range | (0-25)% |
| Oxygen analysis response time | <12s |
| Oxygen analysis noise drift (within 30min) | ≤100 x 10⁻⁶ |
| Oxygen analysis data collection and output resolution | >100 x 10⁻⁶ |
| Carbon dioxide (IR) measurement range | (0-10)% |
| Carbon dioxide measurement linearity | >1% of full range |
| Carbon dioxide data collection and output resolution | >100 x 10⁻⁶ |
| Detector color standard accuracy | ±5% |
| Detector output linearity (transmittance) | <3% |
| Detector absolute penetration rate | <1% |
| Test environment temperature | (10 ~ 30)℃ |
| Control cabinet power supply voltage | AC 220V 50HZ |
| Maximum equipment power | 11.5KW |
| Alarm condition | Flue gas pipeline temperature >400℃ |
| Sample adjustment status temperature | (21-25)℃ |
| Sample adjustment status humidity | (45-55)% RH |
| Distance of test piece from opposite side | 2100mm |
Test principle:
A sample consisting of two vertically angled vertical wings is exposed to the flame produced by the main burner at the bottom of the right angle, the flame is produced by the combustion of propane gas, the propane gas passes through the sandbox burner and produces (30.7 ± 2.0) kW of heat output The
The combustion performance of the sample was evaluated by the entire test procedure of 20 min. Performance parameters include: heat release, smoke production, flame propagation and burning dripping and particulate matter.
Prior to igniting the main burner, the heat output and smoke generation of the burner itself should be measured for a short time using the auxiliary burner remote from the sample.
Some parameters can be measured automatically, while others can be obtained by visual inspection. The smoke pipe is equipped with a sensor for measuring the temperature, light attenuation, the mole fractions of O2 and CO2. and the airflow in the pipeline. These values are automatically recorded and used to calculate the volumetric flow rate, heat release rate (HRR) and smoke rate (SPR).
The transverse propagation and burning of the flame and the particulate matter can be measured using a visual method.
Send product request
Other supplier products
| Acid and alkali protective clothing splash test chamber manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Acid and Alkali Protective Clothing Splash and Spray Liquid Tightness Tes... | |
| Testing apparatus for back and butt forms evaluation | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com 5th Percentile Female SAE Butt Form The 5th Percentile Female SAE Butt Form ... | |
| Foam Repeated Compression Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Foam repeated compression testing machine will accurately measure the thickness of the test... | |
| Automotive Bed Liner Strength Tester for Vehicle Sleeper Bed and Interior Structure Testing | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Parameters FeatureSpecification Drop/Load Capacity Drop fixture ... | |
| Limited Oxyen Index Tester,Automatic oxygen index test machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction STD Limited Oxyen Index Tester for the determination under the test condition... |
Same products
| MMA-120 Handheld Arc Welder | Seller: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | Seller: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |