Cleveland Open Cup Flash Point Tester

The instrument can automatically test the flash point and ignition point of petroleum products (the ignition point of oil exceeding 80 ° C, except crude oil and petroleum fuel), and can be compared with the test results of manual instruments, very simple and convenient.

Meets the ISO 2592 standard. Its programmable logic controller and touch screen (5.5 ") help automate testing, and managing test results is easy.

Application

Covering Multiple Industries, Driving Safety and Efficiency Upgrades

Petrochemical Industry

Refining and Quality Inspection: Monitoring flash point changes in processes such as crude oil distillation and lubricant blending to ensure product quality. For example, a refinery reduced the lubricant rejection rate from 2% to 0.5% by real-time flash point detection.

Asphalt Production: Testing asphalt flash point to meet JTJ052 standards and prevent spontaneous combustion risks during transportation.

Power and Energy Sector

Transformer Oil Testing: Determining the flash point of transformer oil according to DL/T 596 standard to assess its insulation performance and aging degree. For example, a power plant found that the flash point of a transformer oil had dropped to 130℃ through testing, and timely replacement prevented equipment failure.

New Energy Research and Development: Studying the flash point characteristics of new fuels such as biodiesel and hydrogen carriers to promote the application of clean energy.

Transportation Industry

Railways and Aviation: Detecting the flash point of locomotive oil and aviation lubricants to ensure operational safety. For example, an airline found that the flash point of a batch of aviation fuel was too low through testing, and suspended the use of that batch of fuel to avoid flight accidents.

Standard

ASTM D92:Standard Test Method for Flash and Fire Points by Cleveland Open Cup

IP 36. 403:Determination of Flash and Fire Points by Cleveland Open Cup Method (IP 36/403)

ISO 2592:Determination of flash and fire points - Cleveland open cup method

AFNOR T60-118:Petroleum products - Determination of flash and fire points - Cleveland open cup method (AFNOR T60-118)

EN 22592:Petroleum products - Determination of flash and fire points - Cleveland open cup method (EN 22592)

DIN 51376:Testing of mineral oils; determination of flash and fire point by Cleveland open cup

Feature

Automates the testing process for both flash point and ignition point, reducing operator intervention and potential for error.

Provides highly repeatable and reproducible test results, crucial for reliable product characterization.

Designed to comply with the ISO 2592 standard (Cleveland Open Cup Method).

Features an intuitive 5.5-inch color touch screen and programmable logic controller (PLC) for easy operation and test management.

Simplifies data processing and result management, enhancing laboratory efficiency.

Offers a convenient and efficient alternative to traditional manual testing methods.

Equipped with safety features necessary for handling flammable petroleum products.

Product Features:

Automatic Test Procedure: Conducts the heating, test flame application, flash detection, and ignition point determination automatically.

Programmable Logic Controller (PLC): Enables precise control over the test parameters and sequences.

Touch Screen Interface (5.5"): Provides a user-friendly interface for setting up tests, monitoring progress, and viewing results.

Data Processing and Management: Capable of processing test data and facilitating easy management of results.

Micro Printer: For direct printing of test results for record-keeping.

Integrated Heating System: Precisely controls the sample temperature rise rate.

Automatic Ignition System: Applies the test flame at specified intervals.

Flash/Ignition Detection System: Reliably detects the flash and sustained ignition.

Parameters

DescriptionDetails
Dimensions 550(W) x 400(D) x 320(H)mm
Table size 1000(W) x 600(D) x 800(H)mm (assuming the "1.000" is a typo and should be "1000")
Power supply AC 220V 50/60Hz, 15A
Weight 30kg
Instruction manual Available
Exhaust Minimum 50L/s
Tools Propane gas, vacuum cleaner

Accessoriess

Thermal fuse

Gas hose

Heating plate

AC power cord

Instruction manuals, etc.

Test Procedures

Sample Preparation: Heat the sample until it becomes fluid, then pour it into the cup up to the marked line, avoiding contamination of the cup walls.

Heating Control: In the initial stage, heat rapidly at 14-17℃/min, then slow down when the temperature reaches 56℃ below the expected flash point.

Critical Stage: When the temperature is 28℃ below the expected flash point, control the heating rate to 5.5℃/min ± 0.5℃.

Ignition Determination: Sweep a flame across the cup opening every 2℃ increase in temperature. Record the temperature when a transient blue flame appears and immediately extinguishes.

Test Completion: Clean the equipment and switch off the power supply.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

Photovoltaic module insulation resistance tester
Photovoltaic module insulation resistance tester Standard IEC 61215: Terrestrial Photovoltaic (PV) Modules – Design Qualification and Type Approval UL 61730: Safety Standard for Photovolta...
Micro Rapid Closed Cup Flash Point Tester(Pensky Martens)
Micro Rapid Closed Cup Flash Point Tester(Pensky Martens) Automatic Flash Point Tester designed for accurate and repeatable determination of the flash point of petroleum products using the Pensky-Martens c...
Prefilled Syringe Plunger Force Tester
Prefilled Syringe Plunger Force Tester The Push - force Tester for Prefilled Syringes is widely used in the medical device industry, mainly for testing the push - force performance of pr...
Battery Electrode Resistance Tester
Battery Electrode Resistance Tester Standard ASTM B193: Standard Test Method for Resistivity of Electrical Conductor Materials ISO 18516: Conductive Coatings – Measurement o...
Glass Scratch/Abrasion Tester
Glass Scratch/Abrasion Tester The Glass Scratch/Abrasion Tester is used to test the scratch and abrasion resistance of glass. This instrument uses a counter to set the number of...
All supplier products

Same products

Gas Oxidizing Property Tester
Gas Oxidizing Property Tester Seller: Standard International Group (HK) Limited Product Description Conforms to ISO 10156-2010 “Gases and Gas Mixtures Selection of Cylind...
Battery Powder Compaction Density Measurement System
Battery Powder Compaction Density Measurement System Seller: Standard International Group (HK) Limited The Battery Powder Compaction Density Measurement System is a precision testing system designed...
Battery Powder resistance tester
Battery Powder resistance tester Seller: Standard International Group (HK) Limited The Powder Resistivity Meter is a precision testing system designed to measure electrical resisti...
KES-SESRU surface roughness/friction tester
KES-SESRU surface roughness/friction tester Seller: Standard International Group (HK) Limited TheRoughness/Friction Tester analyzes hand movements-particularly, strokes-performed by artisans ...
Air Permeability Tester STD- F8
Air Permeability Tester STD- F8 Seller: Standard International Group (HK) Limited The Air Permeability Tester is designed to quickly and accurately measure the breathability and p...