Boil Test Apparatus for Safety Glazing Materials in Buildings
Function: Determining the effect of exposure of safety glazing materials to temperature and humidity condition by boiling.
Related Standard:
- EN ISO 12543 – 4 Glass in building -Laminated glass and laminated safety glass – Part 4: Test methods for durability – Clause 5 High-temperature test
- AS/NZS 2208 Safety glazing materials in buildings – Appendix F Boil Test
- ANSI Z97.1 Safety Glazing Materials Used in Buildings – Safety Performance Specifications and Methods of Test – Clause 5.3.2 – ( 1 ) Boil Test
Technical Data:
Outside Size: 500mm*300mm*580mm
Inner Size: 395mm*260mm*495mm
Working Voltage and Frequency : AC220V/50HZ
Power: 3KW
Apparatus:The apparatus include two water baths and one temperature control device, each water baths is of sufficient size to hold the test specimen fully immersed vertically on edge for the duration test.
Principle:Specimens are immerse in hot water and boiling water for specified period of time and inspect bubbles and other defects
Working Environment:
Temperature: 20℃±5℃
Atmospheric Pressure: 8.60×104Pa-1.06×105Pa
Relative Humidity: 40%-80%
Instructions:
1, Place the boil test equipment stably, and put the sample holder into the box;
2, Pour water into box, control the water level is about 50-80mm from the top of the box, cover the box;
3, Switch on the power supply, set the temperature-controlled meter to required temperature (63~69°C).
4, When the temperature reach the required temperature (63~69°C), place the samples vertically on the sample holder inside the box,3 minutes later, take out the samples and quickly put them into another boiling test box, after 2 hours, take out the samples and observe.
5, Disconnect the power supply, drain the water after cooling.
Matters needing attention:
1, check whether the box is deformed or damaged before use.
2, check the water level in the box before switching on the power supply.
3, keep the box dry and maintain regularly when it is not used for a long time.
4, it is normal to have a little scale after heating for a long time. Please clean up in time so as not to affect the heating effect.
Packing details:
Plywood case (Meet import and export requirements)
Packing Size: 650mm*535mm*900mm
Total Weight: 49kg
Send product request
Other supplier products
| Quality Control Test Equipment for Safety Glass | Function: Determining the resistance of glazing material to impact by the penetration method. Principle: The specimen is supported vertically in a... | |
| Drop Ball Shock Tester | Function: Determination of whether the safety glazing material has a certain minimum strength and cohesion under impact from a small hard object Su... | |
| BS EN 12600 Pendulum Impact Testing Machine For Glass | BS EN 12600 Pendulum Impact Testing Machine For Glass Reference Standard: BS EN 12600:2002 Glass in building – Pendulum test – Impact t... | |
| Customized Non-standard Sheet Metal Fabrication of Aluminum Profile Frame | Name: Aluminum profile frame 40*40/50*50/60*60 Size: 960 mm*730 mm*1000 mm(Customized) Material: Aluminum profile Angle pieces: aluminum alloy Scre... | |
| GB / T9962-1999 Shot - Bag Impact Testing Machine For Laminated Glass | GB / T9962-1999 Shot - Bag Impact Testing Machine For Laminated Glass Function: Determining the resistance of glazing material to impact by the pen... |
Same products
| Optical Brightener OB-1 (OBA 393) | Seller: Jinan Ogilvy Chemical Co., Ltd | Chemical Name: 2-[(Ethylsulfanyl)methyl]phenyl methylcarbamateC.I.: 393CAS: 1533-45-5Molucular Fo... | |
| PH-A63 Waterproof PH/EC/Temperature Meter | Seller: Kelilong Electron Co.,Ltd | Model PH-A63 Range PH ~14.00PH EC 0~19.99mS/c... | |
| PH-A6 Waterproof pH and Temperature Meter | Seller: Kelilong Electron Co.,Ltd | Model PH-A6 Range PH ~14.00PH Temp 0~50℃ ... | |
| PH-A8 Waterproof Soil PH Meter | Seller: Kelilong Electron Co.,Ltd | Model PH-A8 Range PH ~14.00PH Temp 0~50℃ ... | |
| TH021WIFI Internet of Things TEMP&HUMIDITY Controller | Seller: Kelilong Electron Co.,Ltd | Product description: The electronic thermostat can be used for a variety oftemperature control... |