Dry microbial penetration resistance meter
DryMicrobial Penetration Resistance Testeris designed to determine determine the penetration resistance of material to bacterial on dry particulates at a size range of human dandruff.
The resistance to dry microbial penetration test (ISO 22612) evaluates resistance characteristics of protective equipment to dry bacteria-carrying particles.Testing is most often performed on equipment such as surgical gowns, drapes, and clean air suits when used as medical devices. This test is part of a test suite required by the EN 13795 guidance, which determines performance requirements and performance levels for gowns, drapes, and clean air suits when used as medical devices. It is also included in the test suite required by EN 14126. for personal protective equipment. Testing in compliance to EN 13795 is part of following the European Union Medical Device Directive 93/42/EEC, and is important when seeking a CE mark for products marketed in Europe.
Standards:
ISO 22612; YY / T 0506.5-2009
Test principle:
When testing, the samples were seperately fixed on a container.In these containers, five containers carrying the bacillus subtilis and 1 container carrying uncontaminated talc powder used as control.Insert 1 petri dish at the bottom of each container and beneathe the bottom of the specimen.
The supporting vessel's equipment is oscillated by a gas ball type oscillator, and all of the talcum powder passing through the sample falls onto the petri dish. Remove the petri dish for culture. The experiment results were evaluated by counting the colonies.
Scope of application:
Dry microbial penetration resistance performance tester is professionaly used to determine the bacterial penetration resistance performance of materials to dry particulates in the size range of human body dandruff.Whenn testing, the samples were fixed on a container respectively.In these containers, five containers containing subtilisin talc, one container added with Not stained talcum powder was used as a control.A Petri dish was inserted at the bottom of each container at a distance from the bottom of the timer.
The equipment supporting the container is oscillated by a Gas ball oscillator, and all the talc penetrating the sample Falled on the culture dish , cultured without the culture vessel.The test results were evaluated by counting the growing colonies.
Main parameters:
Vibration form: pneumatic ball vibrator
Vibration frequency: 20800 times / min
Vibration force: 650N
Worktable size: 40cm × 40cm × 10mm
Workbench media: marble plate
Experimental container: stainless steel test container
Work stations: 6
Cabinet negative pressure range: -50 Pa ~ -200 Pa
High efficiency filter filtration efficiency: better than 99.99%
Negative pressure cabinet ventilation flow: ≥ 5m3 / min
Data storage capacity: 100.000 group
Working voltage: 220V 50HZ
Features:
Reliable design, stable test;
Front switch glass doors, easy to observe the operation of the experimental staff;
Removable bracket, height adjustable bracket;
Support, mobile dual casters;
Negative pressure experimental system, with the fan exhaust system and the introduction of high efficiency filter to ensure the safety of operators;
Embedded high - speed industrial microcomputer control;
Dedicated operating software, with software parameters calibration, fault detection automatically
Send product request
Other supplier products
| Helmet Strap Tensile meter | Purpose Apply a certain initial load to the helmet strap system for a certain period of time. Then, let a weight of a certain mass fall from a cer... | |
| Fire Propagation Test Machine | Description This machine is developed in accordance to BS 476-6. It is popularly used for Many British and Hong Kong's building materials or const... | |
| Sulfur Dioxide and Hydrogen Sulfide Corrosion Test Chamber for Automotive Connectors | Meeting Standards: Carry out various sulfur dioxide gas corrosion tests in accordance with GB10587 - 89, GB2423.9 - 93, GJB150, QC/T417.1 - 2001 a... | |
| Particle Strength Tester | Standard ASTM D4179: Standard Test Method for Single-Pellet Pressure Resistance of Catalysts and Catalyst Carriers ASTM D6175: Standard Test Me... | |
| Face Mask Tensile meter | Introduction: The test machine uses the integrated structure of the motor and the speed control system to drive the synchronous belt deceleration ... |
Same products
| Custom Stainless Components for OEM Applications | Seller: Yongjia County Yaokang Technology Co., Ltd. | Our stainless steel float balls are designed for float valves, water tanks, and industrial liquid... | |
| FH-77 Dissolved hydrogen meter | Seller: Kelilong Electron Co.,Ltd | Pen type dissolved hydrogen detector / dissolved hydrogen meter: used to measure the dissolved hy... | |
| Xenon Arc Weather Ometer | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App: +86 181 0160 0864 SI4000Xenon Arc Weather Ometer utilizes a xenon lamp ... | |
| Xenotest | Light Fastness Tester Si220 (SunTest) | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App:+86 180 0160 0864 A Light Fastness Tester, or Xenon Tester, Accelerated ... | |
| Ozone Aging Chamber | Seller: Standard International Group (HK) Limited | Get a quote: Whats/App:+86 156 1874 6768 The Ozone Aging Chamber is a precision testing instrum... |