Limiting Oxygen Index Tester
Производитель: Тысяча сплошная точный электромеханический прибор контрольно-измерительной ЛТД
Все товары поставщика
Все товары поставщика
Introduction
To automatically determine the minimum quantity of oxygen required for combustion. The apparatus determines the flammability characteristics of the materials used in the production of electric cables. The oxygen percentage is detected by a paramagnetic transducer (O2 accuracy <0.1%), and the combustion occurs inside the glass column. The operator is fixing only the O2ratio, while the N2flux is kept constant by proportional valve of the electronic system. Two units separated, one to check the measurements and one to feed oxygen and nitrogen to the column where the testing sample is placed.
Technical Parameter:Main unit:360 x 300 x 240 h (mm)
Colum unit:200 x 300 x 50 h (mm)
Colum:Type A ,Ø100 x 450 h (mm)
Overall dimension:360 x 300 x 740 h (mm)
Net weight:13 kg
Gross weight:19kg
Power supply:230 V 50Hz,single phase
Gas:oxygen, nitrogen, propane, dry industrial air 2 bar
ASTM D 2863Standard Test Method for Measuring the Minimum Oxygen Concentration to Support Candle-Like Combustion of Plastics (Oxygen Index)
ISO 4589-2Plastics-Determination of burning behaviour by oxygen index –Part 2: Ambient-temperature test
CEI 20-22/4Tests on electric cables under fire conditions-Part 3: Tests on bunched wires or cables
ISO 4589-2Plastics-Determination of burning behaviour by oxygen index –Part 2: Ambient-temperature test
CEI 20-22/4Tests on electric cables under fire conditions-Part 3: Tests on bunched wires or cables
Configuration include:
· Support for self supporting specimens
· Support for non-self supporting specimens
· Support for thin specimens
· Column Pyrex glass with cap
· Burner for the ignition of the sample with manual regulator and gas hose
· Clamp for specimen positioning
· Support for self supporting specimens
· Support for non-self supporting specimens
· Support for thin specimens
· Column Pyrex glass with cap
· Burner for the ignition of the sample with manual regulator and gas hose
· Clamp for specimen positioning
· Determination of the percentage of oxygen through a paramagnetic transducer (accuracy O2<0.1% )
· Direct reading on color display 3.2" (% Oxygen, Flow, Air)
· Automatic flow through Mass flow
· Automatic mixing oxygen / nitrogen
· Management of equipment Microprocessor
· Management mixture by addition of air to more economic measures with high concentration oxygen
· Automatic calibration (0 and 100%)
· Distributor flow inside the column
· USB output to connecting PC
· Direct reading on color display 3.2" (% Oxygen, Flow, Air)
· Automatic flow through Mass flow
· Automatic mixing oxygen / nitrogen
· Management of equipment Microprocessor
· Management mixture by addition of air to more economic measures with high concentration oxygen
· Automatic calibration (0 and 100%)
· Distributor flow inside the column
· USB output to connecting PC
Optional:Software testing report data, easy to use, customize your testing procedures and final report.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
Тестер бактериальной эффективности ASTM F2101 (BFE) | Эффективность бактериальной фильтрации для лицевых масок Тестер BFE предназначен для определения эффективности бактериальной фильтрации (BFE) медиц... | |
coatings hardness test machine scratch tester apparatus | With the Shear / Scratch Tester, you can evaluate resistance to shearing, scratching, gouging, marring, scraping and engraving on a variety of flat... | |
Multi-Functional Scratch Tester | Introductionscratches, plastic scraper three kinds of automotive industry common test methods into one.It is suitable for a variety of automotive i... | |
High Efficiency Filter Material Filtration Performance Tester | TheAutomatedFilter Tester Model 8130is designed to comply with commercial respirator mask standards around the globe and has a proven track record ... | |
Multi Finger Scratch/Mar Tester | Introduction: Multi Finger Scratch/Mar Tester uses tungsten carbide as a scratch to accurately detect surface damage caused by unconventional ph... |