Pendulum and charpy impact test machine
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
This advanced Pendulum Impact Tester is specifically tailored for evaluating the impact resistance of various materials, including plastic film, composite film, metallic foil, and more. With applications ranging from scientific research and academic education to laboratory use, research institutes, and quality control departments, Pendulum Impact Tester is a versatile solution for precise and reliable testing.
Film:It is suitable for testing the resistance of plastic film, thin film and composite film to pendulum impact. Such as food and drug packaging bags with PE/PP composite film, aluminum plating film, aluminum plastic composite film, nylon film, etc
Paper, Sheet Testing:It is suitable for testing the impact resistance of paper and paperboard, such as aluminized paper for cigarette packaging and aluminum-plastic paper composite material for Tetra Pak packaging.
Standards
ASTM D3420:Standard Test Method for Flexural Properties of Plastics
ASTM D256:Standard Test Method for Izod Impact Strength of Plastics and Reinforced Plastics
ASTM D4812:Standard Test Method for Tensile Properties of Thermoplastic Films and Sheeting
ASTM D6110:Standard Test Method for Tensile Properties of Rubber and Elastomers (Dumbbell Shapes)
ISO 180:Plastics — Determination of Izod impact strength
ISO 179:Plastics — Determination of Charpy impact strength
ISO 8256:Plastics — Cellular plastics — Determination of tensile properties
ISO 7628:Fibre-reinforced plastics — Determination of flexural properties
ISO 9854:Textiles — Determination of tensile properties — Part 1: Breaking force and elongation at break
DIN 53453:Kunststoffe — Bestimmung der Schlagzähigkeit (Charpy-Verfahren)
DIN 53753:Kunststoffe — Bestimmung der Biegeeigenschaften
Technical Parameters
| Technical Parameter | Specification |
|---|---|
| Applications | Impact Testing, Sample Testing, Strength Testing |
| Materials | Aluminum, Composites |
| Weight | 110 kg |
| Dimensions | 630×500×760 mm (W×D×H) |
| Power | 110-240VAC, 50-60hz |
| Display | 160 x 128 dots graphic LCD module with alphanumeric keypad |
Accessoriess
Izod Vice
Charpy Rest
Tension Impact Rest (Specimen in bed method)
Puncture Impact Fixture
Pipe Rest
Component fixture to customer requirements
Variable weight impact hammers up to 50 Joules
Thermal printer
Full safety enclosure
Low-temperature chamber
Test Procedures
1. Environmental Condition Control
Samples must be pretreated for at least 4 hours in an environment with a temperature of 23±2℃ and humidity of 45%~55% to avoid the influence of temperature and humidity on material properties (refer to GB/T 8809 and ISO standards).
2. Sample Preparation and Installation
Samples should be uniformly 100mm×100mm in size, with no wrinkles, bubbles, or damage on the surface.
Select the punch size according to the material characteristics (e.g., Φ12.7mm for thin films, Φ25.4mm for thick composite films) and match the corresponding fixture.
3. Equipment Parameter Calibration
Before testing, the initial angle of the pendulum (90°), impact velocity (2.5m/s), and energy zero point must be calibrated to ensure compliance with standards such as ASTM D3420.
Verify the equipment accuracy using standard samples to avoid errors caused by mechanical wear or sensor drift.
4. Data Acquisition and Analysis
Record the energy consumed by the punch during the impact process and analyze the material deformation characteristics in conjunction with the stress-strain curve.
Compare the material to standard thresholds (e.g., the impact resistance of food packaging composite films should be ≥2J) to assess whether the material meets the application requirements.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Safety Helmet High Temperature Pre-Treatment Chamber | The safety helmet high-temperature pre-treatment chamber is manufactured according to the GB/T2812-2019 standard for safety helmets and is primaril... | |
| Catheter Sliding Performance Tester | The Catheter Sliding Performance Tester is professionally used for measuring the surface friction and sliding performance of interventional cathete... | |
| Xenon Test Chamber | Features and Benefits 1.The use of xenon arc lamp as a light source, can simulate the full sunlight spectrum, including ultraviolet, visible light... | |
| Touch Screen Tap-and-Scratch Durability Tester | Application Primary applications Durability testing of mobile touch screens Repeated tapping life test of capacitive displays Scratch and slid... | |
| Fully Automated Continuous Flow Reaction Heat Meter | Standards SN/T 3078.1. NY/T 3784. GB 51283 Product Features: 1. Using the coefficient calibration method and flow - regulation calorimetry, it c... |
Похожие товары
| Tabletop Xenon Arc Weathering Tester (SunMaster) | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Water-Cooled Xenon Lamp Weat... | |
| Blow-By Test Rig | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test system for the characterisa... | |
| Air filter System(ISO 16890/EN 779) | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Filter test system for general a... | |
| Oil Thread Leak Test System | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test System for oil thread leak ... | |
| Manual HEPA/ULPA Filter Scanning Test System | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Manual HEPA/ULPA Filter Scan... |