Automatic luer conical joint multi-function device
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The tester complies with the following domestic and international standards for medical Luer connections:
(1) GB/T 1962.1-2015: Syringes – Luer lock and Luer slip connections (China)
(2) GB/T 1962.2-2001: Medical syringes – Performance testing (China)
(3) ISO 80369: Small-bore connectors for liquids and gases in healthcare applications (International)
(4) ASTM F2394: Standard test method for Luer connectors (International)
Features
(1) Multi-function testing of leakage, gas tightness, separation force, torque, assembly ease, slip resistance, and stress cracking.
(2) Supports both locked and non-locked Luer connections.
(3) Adjustable parameters for axial force, separation force, internal water pressure, torque, and hold time.
(4) High-precision measurement with minimal error (≤1%).
(5) Compact and portable design for laboratory or production use.
(6) User-friendly interface with easy adjustment of test parameters.
Parameters
| Parameter | Specification |
|---|---|
| Nominal capacity | 1 mL – 60 mL (optional) |
| Axial force | 0 – 100 N, error ≤ 0.5% |
| Separation force | 20 – 40 N, error ≤ 1% |
| Internal water pressure | 200 – 400 kPa, error ≤ 0.5% |
| Torque | 0.01 – 0.5 N·m, error ≤ 0.5% |
| Hold time | 1 – 60 s, error ≤ 1 s |
| Dimensions (L × W × H) | 450 × 350 × 400 mm |
| Net weight | 28 kg |
| Standards | GB/T 1962.1-2015. GB/T 1962.2-2001. ISO 80369. ASTM F2394 |
Accessories
(1) Main unit – 1 set
(2) Fixture adapters for different Luer sizes – 1 set
(3) Pressure calibration kit – 1 set
(4) Operation manual – 1 set
(5) Power cable – 1 set
Test Procedures
(1) Select the appropriate Luer adapter for the device under test.
(2) Mount the syringe, needle, or catheter securely in the fixture.
(3) Set the required test parameters: axial force, separation force, internal pressure, torque, and hold time.
(4) Start the test cycle. The instrument measures all performance indicators in real time.
(5) Record and print the results, including any failures or deviations.
Maintenance Information
(1) Keep the tester clean and free from dust or fluid contamination.
(2) Inspect adapters and fixtures regularly for wear or damage.
(3) Calibrate pressure and torque sensors periodically to maintain accuracy.
(4) Store in a stable, dry environment.
(5) Follow electrical safety regulations during operation.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Vertical low temperature cold resistance testing machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Luggage rod reciprocating testing machine for luggage rod reciprocating fatigue test. In th... | |
| Alternative to Bearing Friction Torque Measuring Instrument | Whats/App: +86 156 1874 6768 Web: standard-groups.com Bearing friction torque is an important factor affecting bearing noise, vibration, and life... | |
| Femoral Stem Neck Prostheses Fatigue Testing Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0809.6 (China): Specifies neck fatigue testing for femoral stems in partial/... | |
| Finger Tip Abrasion Tester,IEC 60068,Fingertip Wear Testing Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Finger Tip Abrasion Tester is designed to simulate human fingertip contact on a test sp... | |
| Toy Flame Suppression Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards: EN 71-2 ISO 8124-2 GB 6675.3 EN 16890-2017 Application: Toys Flammability ... |