Gas Oxidizing Properties Tester | ISO 10156 Oxidizing Gas Analyzer
Get a quote:
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Scope of Application and Standards
The tester is used to determine whether a gas or gas mixture has an oxidizing property higher than air, applicable to toxic and corrosive gases or their mixtures.
Complies with the following standards:
ISO 10156-2010. Part 4: Selection of Cylinder Valve Outlet by Determination of Combustion Potential and Oxidizing Ability of Gases and Gas Mixtures
GB/T 27862-2011. Part 4: Test Methods for Classification of Chemical Hazards — Combustion Potential and Oxidizing Ability of Gases and Gas Mixtures
Technical Specifications
| Parameter | Specification |
|---|---|
| Working Environment | 5℃–40℃, <85% RH |
| Test Chamber Material | Stainless Steel 316L |
| Test Chamber Pressure Resistance | ≥5000 kPa |
| Test Chamber Volume | ≥0.005 m³ |
| Accessories | Magnetic stirrer with adjustable speed |
| Ignition System Metal Rod Diameter | ≥3 mm |
| Nichrome Wire Diameter | 0.05–0.2 mm |
| Ignition Energy | 10–20 J |
| Explosion Pressure Sensor Range | 0–4000 kPa |
| Explosion Sensor Pressure Resistance | ≥6000 kPa |
| Explosion Sensor Accuracy | ±0.1% FS |
| Explosion Sensor Response Time | ≤1 ms |
| Gas Distribution Sensor Range | 0–200 kPa |
| Gas Distribution Sensor Accuracy | ±0.1% FS |
| Temperature Measurement Range | 0–150℃ |
| Temperature Measurement Accuracy | ±1℃ |
| Main Unit Dimensions | 830 mm × 650 mm × 1200 mm |
| Electrical Control Box Dimensions | 320 mm × 500 mm × 500 mm |
| Weight | 280 kg |
| Main Unit Power Supply | AC 220V ±10%, 50 Hz ±3 Hz |
| Rated Power | 400 W |
| Ignition Power | 3000 W |
Functional Features
Separate design for the testing apparatus and control system, ensuring safety of both the operator and instrument.
Comes with PC operation software to reduce operational difficulty.
Supports remote ignition for operator safety.
Equipped with high-precision pressure sensors to accurately measure explosion pressure and control gas partial pressure and mixture concentration.
Magnetic stirrer for uniform gas mixing.
Electronic microprocessor control of ignition timing.
Automatic cleaning function.
Automatic pressure relief option after testing, releasing pressure in the test chamber.
Multiple explosion-proof features to protect personnel.
Automatic fault detection and audiovisual alarm system.
Fully automated test process, supporting parameter setting, data storage, and report generation without manual intervention.
Accessories
(1)Gas Sample Chamber: Securely holds the gas for testing its oxidizing properties.
(2)Oxidation Sensors/Detectors: Measures oxygen content and reactivity.
(3)Control Unit: Regulates temperature, pressure, and gas flow during tests.
(4)Data Interface: Connects to a computer or display for monitoring and recording results.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Safety Helmets Acceleration Impact & Penetration Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Safety Helmets Acceleration Impact & Penetration Tester is a dual-mode testing inst... | |
| Mechanical Scratch Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Primary test purposes Coating scratch resistance evaluation Coating hardness... | |
| Safety Footwear Compression Puncture meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Test purpose Measure the compression resistance of the shoe head cover and the puncture re... | |
| Walk In Pharmaceutical Stability Chamber.Pharmaceutical stability test chamber | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Pharmaceutical Stability Testing Laboratory is a high-precision environmental simulatio... | |
| Industrial chillers | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ASHRAE 15(Safety Standard for Refrigeration Systems) ISO 5149(Refrigerating S... |