Electric Cables Halogen Acid Gas Release Tester,Cable Halogen Acid Gas Release Test Machine
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Application
This apparatus is widely used in laboratories and quality control departments involved in fire safety and material performance evaluation. Typical application scenarios include:
Testing halogen acid gas release characteristics of power cables, control cables, and communication cables
Evaluation of low-smoke halogen-free (LSHF) cable materials
Fire performance assessment of insulation and sheath compounds
Research and compliance testing for cable manufacturing enterprises
Type testing and routine inspection in third-party testing laboratories
Standards
The test apparatus is designed in accordance with the following international and national standards:
(1) GB/T 17650 – Determination of the Total Amount of Halogen Acid Gas Released During Combustion of Electric Cables
(2) IEC 60754-1 – Test on Gases Evolved During Combustion of Materials from Cables – Part 1: Determination of the Amount of Halogen Acid Gas
(3) IEC 60754-2 – Test on Gases Evolved During Combustion of Materials from Cables – Part 2: Determination of Acidity (by pH Measurement) and Conductivity
Parameters
| Item | Specification |
|---|---|
| Tube Furnace Heating Length | 600 mm |
| Maximum Heating Temperature | Up to 950 °C with thermal equilibrium |
| Temperature Measurement Range | 0–1100 °C |
| Temperature Sensor | High-temperature K-type thermocouple |
| Temperature Accuracy | ±0.1 °C |
| Quartz Tube | Heat-resistant quartz, thickness ≥ 2 mm |
| Carrier Gas | Compressed air |
| Equipment Dimensions | 1500 (W) × 450 (D) × 400 (H) mm |
| Control Cabinet Dimensions | 500 (W) × 400 (D) × 250 (H) mm |
| Power Supply | AC 220 V, 16 A |
| Weight | Approx. 80 kg |
Features
Integrated structural design places the specimen guide tube and tube furnace on the same plane, minimizing positional deviation and improving glassware sealing performance.
Independent control cabinet enhances operational convenience and flexible equipment layout.
Tube furnace heating system adopts a temperature control module combined with a power regulator and PID control to ensure uniform and stable temperature rise.
Protective enclosure with perforated structure improves heat dissipation while protecting the furnace body.
Heat-resistant quartz tube is used for both gas transmission and specimen heating, with ground joints to enhance sealing performance.
Ceramic combustion boat is driven by a strong magnetic mechanism and transmission rod, ensuring stable and precise movement of specimens into the furnace core zone.
Lead screw transmission mechanism enables accurate positioning and smooth operation of the traction tube, reducing the risk of quartz tube damage caused by manual operation.
Imported precision rotameter controls carrier gas flow, combined with branded pressure gauge and pressure regulator to ensure stable airflow.
Dual filter columns filled with activated carbon and color-changing silica gel are used to purify and dry the carrier gas.
PLC and touchscreen control system enables intelligent temperature regulation and improved furnace temperature accuracy.
Built-in test programs for total halogen acid gas, conductivity, pH value, and quartz tube cleaning, with stored standard reference values.
Supplied with complete testing accessories including washing bottles, beakers, graduated cylinders, and magnetic stirrers.
High-quality pH meter and conductivity meter are included for accurate measurement of solution acidity and conductivity.
Accessories
(1) Heat-resistant quartz tube
(2) Ceramic combustion boat
(3) Precision rotameter
(4) Activated carbon filter column
(5) Color-changing silica gel filter column
(6) Glass washing bottles
(7) Beakers and graduated cylinders
(8) Magnetic stirrer
(9) pH meter
(10) Conductivity meter
Test Procedures
Prepare the cable specimen according to the relevant standard requirements and place it in the ceramic combustion boat.
Insert the combustion boat into the quartz tube and position it outside the furnace heating zone.
Set the target furnace temperature according to the selected standard (800 °C for IEC 60754-1 or 950 °C for IEC 60754-2).
Start the carrier gas flow and ensure stable airflow through the system.
Move the combustion boat into the furnace core zone using the magnetic transmission mechanism.
Collect the combustion gases through the absorption system.
Measure the total halogen acid gas volume, pH value, and electrical conductivity according to the selected test method.
Complete the test and perform the quartz tube cleaning procedure using the built-in program.
Maintenance Information
Regular inspection of the quartz tube, sealing joints, and gas pipelines is recommended to ensure reliable test results. The filter materials in the purification columns should be replaced periodically based on usage conditions. Temperature calibration of the thermocouple and verification of pH and conductivity meters should be conducted according to laboratory quality management requirements. After each test, the quartz tube and glassware should be thoroughly cleaned to avoid cross-contamination.
Conclusion
The GB/T 17650 Halogen Acid Gas Release Test Apparatus for Electric Cables provides a standardized and reliable solution for evaluating halogen acid gas emissions from cable materials under high-temperature conditions. With precise temperature control, stable gas flow regulation, and integrated measurement of halogen acid content, pH value, and conductivity, the system fully meets the requirements of GB/T 17650 and IEC 60754 standards, making it suitable for both research and routine compliance testing.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Smoke Density Chamber, ISO 5659,Smoke density testing equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction The main principle is to measure smoke density in a particular space, cable o... | |
| Geotextile-Bentonite Cushion Permeameter | Whats/App: +86 156 1874 6768 Web: standard-groups.com The geotextile-bentonite cushion is a finished waterproof soil mat with excellent water-ret... | |
| Flame retardant testing machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Feature highlights: Flame Retardant Tester is designed for testing plastic runway surface ... | |
| Best Leather Rubbing Color Fastness Tester | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com In daily life, leather products, such as leather clothing, purses, leather sh... | |
| Precision Fiber Slicer | Whats/App: +86 156 1874 6768 Web: standard-groups.com introduction G214Precision Fiber Slicer is used to make the accurate fiber cross-sectional... |
Похожие товары
| ER4047 Aluminum Mig Welding Wire | Продавец: 868776 | Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing reliable welding material... | |
| ER4043 Silicon Aluminum Welding Wire | Продавец: 868776 | Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing high-quality welding mate... | |
| Choosing the Right Equipment for Heavy-Duty Cleaning | Продавец: Taizhou Feike Mechanical Technology Co., Ltd. | FEIKE’s High Pressure Washer is designed for industrial cleaning, equipment maintenance, ve... | |
| ER4043 Silicon Aluminum Welding Wire | Продавец: 868776 | Hangzhou Kunli Welding Materials Co., Ltd. offers ER4043 Silicon Aluminum Welding Wire for alumin... | |
| Высокоскоростная экструзионная линия для производства гофрированных труб с двойными стенками из ПЭВП с водяным охлаждением | Продавец: Shandong HUASU Plastic Technology Co.,Ltd. | Формы имеют структуру с двойным каналом потока и фильерную головку, обработанную азотированием и ... |