HEPA Filter Element Quality Control Test SystemEN 1822-3 HEPA and ULPA filter testing System

The HEPA Filter Element Quality Control Test System is designed for quality control of small HEPA filter elements in production. It complies with EN 1822-4 and ISO 29463-4. Appendix E (“Efficiency and leakage test for particle size 0.3–0.5 µm”). The system is suitable for filter elements that cannot be scanned due to geometric design, ensuring overall filter efficiency is sufficient for production QC.

Application

(1) Small HEPA Filter Element Efficiency Testing: Verification of overall efficiency for small HEPA elements.

(2) HEPA Filter Grade Confirmation: Determine HEPA classification according to standards.

(3) Integrated Production Control: Can be integrated into the production line for real-time quality assurance.

Standards

The system supports and complies with:

(1) GB/T 13554-2020 – High-efficiency air filters.

(2) GB/T 6165-2021 – HEPA filter performance: efficiency and resistance.

(3) EN 1822-3 – HEPA and ULPA filter testing methods.

(4) EN 1822-4 – Efficiency and leakage test for HEPA filters (0.3–0.5 µm).

(5) ISO 29463-4 – HEPA/ULPA filter performance standard.

(6) ISO 29463-3 – HEPA/ULPA filter design, performance, and acceptance testing.

Parameters

ParameterUnitValue
Test Flow Rate m³/h 25 … 130
Test Cycle s 30
Differential Pressure Range Pa 0 … 3000 (±1% final value)
Test Sample Dimensions mm Max 300 × 300 × 300
Power Supply 3 × 400 VAC, 50/60 Hz, 16 A
Operating Gas/Air Custom (DEHS, paraffin oil, NaCl, etc.)
Particle Measurement Particle counter (4 or 8 channels) with vacuum pump
Other Sensors Temperature, relative humidity, barometric pressure
System Dimensions (W × H × D) mm 2000 × 1500 × 1000
Weight kg approx. 600

Features

Configurable Target Efficiency and Pressure Drop: Freely set for each filter element according to production requirements.

Custom Filter Adapters: Adaptable to different small HEPA elements, quick change (~5 minutes).

Compact Design: Small footprint suitable for production environments.

High Automation: Minimal operator intervention, fast test completion.

24/7 Production Support: Suitable for continuous manufacturing operations.

Fast Test Cycle: 30 seconds per filter element ensures high throughput.

Safety Features: Complete protection for production environments.

Traceable Data Management: Test results are recorded and reports generated for quality tracking.

Accessories

AccessoryQuantity
Aerosol Generator 1
Particle Counter (4 or 8 channels) 1
Self-Regulating Dilution System 1
Test Filter Adapter Customized per filter element
Temperature & Humidity Sensor 1

Test Principle

Overall Efficiency Testing:

DEHS aerosol is generated upstream; particle concentrations are measured upstream and downstream.

Upstream concentration is measured with an optical particle counter and adjusted via the dilution system.

Downstream concentration is measured similarly; overall efficiency is calculated and compared to nominal target.

Efficiency Comparison:

Overall measured efficiency is compared with target efficiency to quickly classify “pass” or “fail.”

Differential Pressure Measurement:

Measures pressure drop across the filter element, generating pressure curves and dust loading capacity data to ensure stable performance at different flow rates.

Operation Instructions

System Startup:

- Check connections (air, power, aerosol generator).

- Set target efficiency, pressure drop, and other test parameters.

- Start the test program.

Testing Process:

- Install small HEPA filter elements into custom adapters.

- System automatically conducts efficiency and differential pressure tests.

Data Recording and Reporting:

- Upon test completion, results are stored in the database.

- Generate detailed reports supporting quality traceability.

Experimental Procedure

Preparation:

- Configure test parameters (target efficiency, flow rate).

- Install filter elements in adapters.

- Verify aerosol generator and particle counters are operational.

Testing:

- System automatically measures differential pressure, efficiency, and dust loading.

- Upstream and downstream particle concentrations are recorded.

Result Processing:

- System provides “pass/fail” results and generates detailed reports.

Maintenance Information

Regularly check aerosol generators and particle counters.

Calibrate sensors periodically.

Keep the system clean to maintain accurate measurements.


Отправить запрос, связаться с поставщиком

Кому: Standard International Group (HK) Limited
Ваш E-mail:
Текст письма:


Send to other suppliers

Другие товары поставщика

Mask Synthetic Blood Penetration Tester
Mask Synthetic Blood Penetration Tester Medical Face Mask Synthetic Blood Penetration Resistance Tester Medical masks synthetic blood penetration tester is used to determine the resistan...
Helmet Impact Acceleration Measurement Device
Helmet Impact Acceleration Measurement Device Purpose This instrument is suitable for impact - collision tests of helmets. It is a testing instrument used to evaluate the impact - resistance p...
Minimum Ignition Energy (MIE) – Dust Cloud Tester
Minimum Ignition Energy (MIE) – Dust Cloud Tester Description: Dust Cloud Minimum Ignition Energy Tester disperses a certain mass of dust specimen to be tested into the Hartmann tube device throug...
Zipper Fatigue Tester
Zipper Fatigue Tester Product Introduction: The functional life of a zipper is often determined by its ability to withstand numerous opening and closing cycles without ...
Appliance Stability Testing Apparatus (for Appliance Stability Tester)
Appliance Stability Testing Apparatus (for Appliance Stability Tester) Product Introduction The device conforms to GB4706-1 Article 20.1. IEC60335-1 Article 20.1 and other standards. The device is composed of a workbe...
Все товары поставщика

Похожие товары

Dynamic Shear Rheometer
Dynamic Shear Rheometer Продавец: Standard Groups Dynamic Shear Rheometers are configured to meet the demands of everyday quality control testing o...
Noxious Gas Test Chamber
Noxious Gas Test Chamber Продавец: Standard Groups Noxious gas tеst chambеrs play a crucial role in thе industrial sector by еnsuring thе longеvity ...
Chest Drainage Tester
Chest Drainage Tester Продавец: Standard Groups The Chest Drainage Tester is a precision instrument designed to evaluate the performance of dispo...
Plastic Smoke Density Tester
Plastic Smoke Density Tester Продавец: Standard Groups You can use the F313 Plastic Smoke Density Tester in the following areas: Plastics industry: Eva...
Washing Fastness testing machine
Washing Fastness testing machine Продавец: Standard Groups G228B washing fastness tester is used for washing, dry-clean color fastness testing. Electric tou...