IV Set Pressure Leakage Testing Equipmen
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The Infusion Set Positive Pressure Leakage Tester complies with the following medical device standard:
GB 8368-2018 (Disposable infusion sets for single use, gravity feed)
Parameters
| Parameter | Specification |
|---|---|
| Positive Pressure Output Range | 120 kPa–160 kPa (adjustable), LED digital display, accuracy within ±5% |
| Positive Pressure Holding Performance | Pressure drop shall not exceed 1 kPa within 3 minutes |
| Test Duration | 1 s–60 s (adjustable), LED digital display, accuracy within ±1 s |
| Pressure Range | 0–69 kPa |
| Time Setting Range | 0–999 s |
| Power Consumption | 200 VA |
| Display Type | 320 × 240 monochrome LCD, full Chinese window interface |
| Control Method | Keyboard control for menu navigation and on-screen keypad operation |
| Operating Temperature | 5–40 °C |
| Relative Humidity | 45–80% |
| Power Supply Voltage | 220 V ±22 V |
| Power Frequency | 50 Hz ±1 Hz |
| Equipment Classification | Class I, Type B |
Features
Standard-oriented design
Developed specifically to meet the positive pressure leakage test requirements defined in national medical device standards.
Integrated intelligent control system
The built-in CPU, mainboard, LCD display, and keyboard enable stable control and precise parameter setting.
Clear and intuitive operation
Full-window Chinese interface with menu-driven operation allows straightforward use without specialized training.
Stable pressure control and monitoring
Adjustable positive pressure output with digital display ensures consistent test conditions and reliable leakage detection.
Compact and functional structure
Combines control, measurement, and data output functions in a single, practical testing unit.
Accessories
Power cable
Built-in printer
Drain outlet connector
Operation manual
Test Procedures
Place the instrument horizontally on a stable testing bench.
Install the test specimen securely onto the drain outlet connector.
Connect the power plug to the distribution board, ensuring reliable grounding of the power socket.
Power on the instrument and set the required positive pressure and test duration via the control panel.
Start the test and observe pressure stability during the test period.
Record or print the test results as required after test completion.
Maintenance Information
Regularly inspect pressure connections and sealing components to prevent leakage.
Keep the display screen and control panel clean and free from moisture.
Verify pressure accuracy periodically to maintain compliance with test requirements.
Ensure the power supply grounding is reliable before each operation.
Store the instrument in a clean, dry environment when not in use.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Custom low-load wear testing instrument | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The Low-Load Wear Testing Machine is widely used in applications where ultra-l... | |
| Femoral Condyles Components Fatigue Performance Tester | Standard ASTM WK51649: Specifies fatigue testing protocols for metallic femoral components in total knee arthroplasty, including F-N curve generat... | |
| In-situ Gas Analysis System for Lithium Battery Thermal Runaway,thermal runaway gas monitoring system | Whats/App: +86 156 1874 6768 Web: standard-groups.com The In-situ Gas Analysis System for Lithium Battery Thermal Runaway is an advanced testing ... | |
| ASTM D1238 melt flow indexer | Introduction Melt flow indexer(MFI) are used for military uniforms, canvas, canvas, labor insurance products. The fabric fusing tester is designed... | |
| Intramedullary Nails Static Four-point Bending Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0591 (China): Standard for static four-point bending tests of lockable intra... |
Похожие товары
| Bluetooth catheter flow meter | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 15812.1 - 2005 Non - Vasc... | |
| Automatic Catheter Hydraulic Leakage Tester device | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 15812.1: Disposable steri... | |
| Balloon Catheter Leak and Compliance Tester | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 13988-1:2017: Intravascula... | |
| Advanced catheter residual vacuum tester instrument | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards YY/T 0339 - 2019: Suction Cath... | |
| Bend catheter flexibility equipment | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 10555 -Intravascular cathe... |