Hypodermic needle toughness tester instrument
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The device meets and supports testing according to the following international and national standards:
(1) GB15811 – Sterile disposable injection needle
(2) GB18457 – Stainless steel needle for medical equipment manufacturing
(3) ISO 9626 – Stainless steel needle for medical use
Features
(1) High-precision touch LCD interface for intuitive operation.
(2) Adjustable bending angles for normal, thin, and ultra-thin needle walls.
(3) Automatic bi-directional bending for standardized toughness testing.
(4) High-brightness 4-digit LED display for accurate parameter monitoring.
(5) Supports multiple syringe diameters (0.2–3.4 mm).
(6) Compact and robust design with net weight of 10 kg.
(7) Free perpetual upgrades according to national standards.
(8) Complete accessory kit included for immediate testing.
Test Procedure
(1) Fix one end of the needle tube firmly in the jig.
(2) Adjust the span according to needle wall type:
Normal wall: 25° bending
Thin wall: 20° bending
Ultra-thin wall: 15° bending
(3) Apply force at the specified span position, bi-directionally, 20 times at a frequency of 0.5 Hz.
(4) Observe the needle for breakage, deformation, or failure.
Principle: One end of the needle is fixed. A force is applied from the fixed point to the prescribed span, bending in one direction and then in the opposite direction, repeated for the defined number of cycles.
Technical Parameters
| Parameter | Specification |
|---|---|
| Syringe nominal specifications | 0.2 – 3.4 mm |
| Test frequency | 0.5 Hz |
| Rocker arm bending angle | 15°, 20°, 25° (switch selectable, displayed on high-brightness 4-digit LED) |
| Bending distance | 6 – 95 mm, accuracy ±0.1 mm (10 selectable spans) |
| Number of experiments | Bi-directional force 20 times |
| Net weight | 10 kg |
| Overall dimensions (L × W × H) | 300 × 240 × 240 mm |
| Standards | GB15811. GB18457. ISO 9626 |
| Control & Display | 5.7-inch LCD touch menu (120 × 90 mm) |
| Upgrade | Free perpetual upgrade according to national standards |
| Power supply | AC 220 V, 50 Hz |
| Part List | Host (1 set), Printing paper (1 roll), Fixture (1 set), Weight (1 set), Operating Instructions |
Accessories
(1) Printing paper (1 roll)
(2) Needle fixture (1 set)
(3) Weight set for bending calibration
(4) Operating instructions in English/Chinese
Maintenance Information
(1) Keep the device clean and dry; wipe with a soft cloth after use.
(2) Check the calibration of bending angles and LED display periodically.
(3) Avoid impact or vibration to ensure precise mechanical performance.
(4) Follow local electrical safety standards for power supply and grounding.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Steel Structure Fireproof Coating Small Sample Test Furnace | Whats/App: +86 156 1874 6768 Web: standard-groups.com TheSmall Sample Test Furnace for Steel Structure Fireproof Coatingis a specialized thermal ... | |
| Catheter Guidewire Friction meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Test standard YY/T1536-2017.ISO 11070 Test principle The sample is clamped between the t... | |
| Factory Supply RPP Testing Machine for Fire Resistant Protective Clothing | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Outdoor Sports Gear: Evaluating the waterproof breathability and... | |
| Hair Dryer Temperature Test System | Whats/App: +86 156 1874 6768 Web: standard-groups.com TheHair Dryer Temperature Testing Deviceis a dedicated test apparatus designed for evaluati... | |
| Automotive Surface Durability Testing Machine for R&D and OEM Use | Factory Direct Manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Features Full stainless steel test chamber with transparent doors for observ... |