Automated needle flow tester for syringes
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The instrument is designed to meet the following standards:
(1) YY 0321.1: Disposable anesthetic puncture kit (China)
(2) YYB: Stainless steel injection needle for pre-filled and sealed syringes (China)
Optional compliance with international standards:
(3) ISO 7886-1: Sterile hypodermic syringes – Part 1
(4) ISO 8537: Single-use sterile needles and syringes
Features
(1) Measures syringe flow rate accurately over a fixed time interval.
(2) Integrated control system for precise pressure and flow management.
(3) Supports different syringe sizes with customizable fixing fixtures.
(4) Real-time monitoring of test parameters, ensuring reliable results.
(5) Easy-to-use interface for rapid setup and operation.
(6) Compact design suitable for laboratory or production environments.
Parameters
| Parameter | Specification |
|---|---|
| Test pressure | 0 – 400 kPa |
| Pressure resolution | 0.1 kPa |
| Pressure accuracy | ±1% FS |
| Air source interface | 8 mm polyurethane pipe |
| Test interface | 6% Ruhr connector (customizable) |
| Host dimensions (L × W × H) | 325 × 420 × 170 mm |
Accessories
(1) Test body – 1 set
(2) Syringe fixing device – 1 set
(3) Flow meter – 1 set
(4) Control system – integrated
(5) Operation manual – 1 set
(6) Optional connectors and adapters for different syringe types
Test Procedures
(1) Secure the syringe in the fixing device according to its nominal size.
(2) Connect the air source and test interface.
(3) Set the test pressure and duration on the control system.
(4) Start the test; the device will measure the volume of liquid discharged over the specified time.
(5) Record the flow rate data and analyze results for compliance.
Maintenance Information
(1) Keep the device clean and free from liquid or dust ingress.
(2) Inspect the flow meter and syringe fixtures regularly to ensure accurate readings.
(3) Avoid exposing the device to excessive humidity, vibration, or temperature fluctuations.
(4) Calibrate pressure sensors periodically to maintain accuracy.
(5) Follow standard electrical safety guidelines during operation.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Battery Isothermal Calorimeter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Battery Isothermal Calorimeter is a professional instrument developed based on the principl... | |
| Photovoltaic Module Outdoor Exposure Test System for Solar Reliability Testing | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Photovoltaic Module Outdoor Exposure Test System is designed to simulate ... | |
| Coatings Shear Fatigue Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0988.13-2017 (China): Specifies shear fatigue testing protocols for coatings... | |
| Plastic ball indentation hardness tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Plastic Ball Indentation Hardness Tester is a precision instrument designed to evaluate... | |
| Walk-In Drug Stability Chamber | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards WHO Principles Require: Long-term stability testing conditions at 25°C/60% ... |