Oil Thread Leak Test System

Get a quote:
Whats/App: +86 181 0160 0864
Web: gellowen.co.uk
Test System for oil thread leak test of HEPA filter according to ISO 29463-4 (EN 1822-4). The oil thread test mentioned in ISO 29463-4 (1822-4), annex A is used for visual proof of the absence of leakage of HEPA filters. This simple quality test procedure may be an alternative to the scan method.
The oil thread test (DOP Test) is also applicable if the scan method is not fully applicable because of the filter design (e.g. V-shaped arrangements).
Applications
integral visual leak testing of HEPA/ULPA filters
Standards
This system complies with or supports testing according to the following international standards:
(1) ISO 16890-2: Air filters for general ventilation – Part 2
(2) ISO 16890-3: Air filters for general ventilation – Part 3: Efficiency testing
(3) ISO 16890-4: Air filters for general ventilation – Part 4
(4) ASHRAE 52.2: Method of Testing General Ventilation Air-Cleaning Devices
(5) EN 779: Particulate air filters for general ventilation
(6) ISO 10121-3: Cleanroom air filter testing
Benefits
cost-effective solution for leak testing of filters ranging up to group H (filter class H13 and H14)
easy manual operation
constant test conditions due to pre-adjustable test parameters
effective testing of larger numbers with pre-adjusted test parameters
flexible and easily configurable filter holder for different filter dimensions
Principle
The filter element (sample) is mounted horizontally with the upstream side down on the test bench and secured with a pneumatically operated fixing device. For testing, the filter is then exposed to a flow of a polydisperse aerosol of oil droplets (DEHS).
An existing leak is made visible on a rising oil thread on the downstream side (upper side) of the filter.
Technical data
| Parameter | Unit | Value |
|---|---|---|
| Test flow rate | L/min | 50 … 600 |
| Dimensions of test specimen (W × H × D) | mm | min. 305 × 305 × 30 / max. 915 × 800 × 300 |
| Operating medium (aerosol substance) | – | DEHS |
| Compressed air supply (operating gas/air) | bar | 6 |
| Power supply | – | 230 VAC, 1 A |
| System dimensions (W × H × D) | mm | 1200 × 610 × 910 |
| Weight | kg | approx. 100 |
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| SO2 H2S HCL Corrosion Test Equipment | Get a quote: Whats/App: +86 181 0160 0864 TheSO₂ H₂S HCl Corrosion Test Equipmentis designed to simulate corrosive environments using noxious gas... | |
| Suitcases Vibration Impact Tester | Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk This machine is used to test the shock resistance of various luggage handles, car ... | |
| Surface Scratch Resistance Testing Machine | Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Surface Scratch Resistance Testing Machine is a PLC-controlled instrument desi... | |
| PV Hail Impact Tester | Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The PV Hail Impact Tester is a precision instrument designed to evaluate the mecha... | |
| 45-degree dry cleaning tester/ 45 Degree Dry Cleaning Testing Machine | Product Introduction: The instrument is composed of a bottom plate, a control panel and a drum. The clothes to be washed and the dry cleaning agen... |
Похожие товары
| MMA-120 Handheld Arc Welder | Продавец: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... | |
| SHX Series Circulating Fluidized Bed Steam Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi... | |
| YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ... | |
| DHL Series Coal-Fired Steam Boilers | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b... | |
| SZL Series Biomass Steam Boiler | Продавец: Hebei Genjux Boiler Co.,Ltd | Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi... |