Flooring Radiant Panel Test Apparatus EN ISO 9239-1 | Fire Performance Flooring Test Equipment Manufacturer
Get a quote:
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Technical Parameters
| Parameter | Specification |
|---|---|
| Maximum power | 5 kW |
| Timing accuracy | <1 s/h |
| Heat flux sensor range | 0–15 kW/m² |
| Heat flux sensor accuracy | ±0.2 kW/m² |
| Heat flux sensor precision | <±3% |
| Radiant high-temperature sensor range | 400–600°C, ±0.5°C |
| Black body temperature error | ±5°C |
| Exhaust air velocity | (2.5 ± 0.2) m/s |
| Exhaust capacity | 39–85 m³/min |
| Porous refractory radiation panel size | 300 × 450 mm, withstands ~900°C |
| Thermocouple specification | Ø3.2 mm, K-type stainless steel armored |
| Burner specification | Outer Ø10 mm, inner Ø6 mm, nozzle Ø0.7 mm |
| Flame height adjustment | 60–120 mm, adjustable |
| Ignition system | High-voltage electronic ignition |
| Exhaust system | Forced ventilation, fumes discharged outdoors |
Accessories
| Accessory | Quantity | Description |
|---|---|---|
| 304 stainless steel flame height gauge | 1 | Measures flame height |
| Gas pipe clamps | 2 | Secures gas tubes |
| User manual | 1 | Instructions for operation and maintenance |
| Certificate of conformity | 1 | Confirms product quality inspection |
Features
The main components are imported, which is safe and reliable, long service life;
Gas pipeline layout is reasonable, equipped with gas leakage and tempering device. The instrument automatically closes at the inlet of gas leakage, and alarm;
Radiant plate temperature can be automatically adjusted.
Propane gas supply is equipped with imported mass flow controller. The air output is equipped with a separate blower. Automatic adjustment of the frequency without manual;
Normal test, the flame spread distance can be recorded through the foot switch, without the need to manually click the mouse to operate the computer;
Instrument with industrial one machine, is easy to operate and reliable;
The instrument adopts advanced control software developed by VB language. Centralized control acquisition and processing data analysis;
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Mobile Phone Air Tightness Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Mobile Phone Air Tightness Tester is designed to evaluate the sealing performance of sm... | |
| Damp heat test chamber for solar panels IEC standard | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Photovoltaic Module Damp Heat Test Chamber is designed to simulate long-term tropical a... | |
| Building Materials Single Burning Item Flammability Testing Device | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB 8624-2012: building materials and product combustion performance rating GB/T... | |
| Cut Susceptibility Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard IEC 61730-2: Photovoltaic (PV) Module Safety Qualification – Requirements f... | |
| Manikin Breathing System | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application HVAC education and research, including indoor air quality evaluation, human br... |