Automatic Reaction Calorimeter

This instrument is a professional testing instrument that simulates the specific process of chemical reactions on a liter scale, and measures and controls important process variables. It can simulate the real situation of batch or semi-batch reactions in a factory under laboratory conditions.

Applications

This instrument simulates real-world intermittent or semi-intermittent reactions in a factory, measuring the exothermic heat flow of samples within the reactor in real time. It obtains information on the total exothermic heat, specific exothermic heat, real-time conversion rate, and material accumulation, automatically calibrating the specific heat capacity and heat transfer coefficient of the material system. It then calculates the adiabatic temperature rise of the target reaction and the maximum temperature the system can reach after runaway, thereby assessing the hazard level of the reaction process and providing a basis for process optimization and scale-up. It is widely used in reaction process design, process optimization and scale-up, process safety assessment, and scientific research on reaction thermodynamics in fields such as fine chemicals, pharmaceuticals, agricultural fertilizers, and third-party safety assessment.

Standards

GB 51283-2020: Fire Protection Standard for Engineering Design of Fine Chemical Enterprises

GB/T 22232-2008: Determination of Thermal Stability of Chemical Substances - Differential Scanning Calorimetry

GB/T 17802-2011: Thermal Analysis Test Method for Kinetic Constants of Thermally Unstable Substances

SN/T 3078.1-2012: Guidelines for the Evaluation of Thermal Stability of Chemicals - Part 1: Accelerated Calorimetry

Features

1. It offers three calorimetry methods: heat - flow, power - compensation, and reflux, allowing users to choose according to their needs.

2. It supports multiple operation modes such as isothermal, constant - temperature, and scanning.

3. Optional glass normal - pressure reaction vessels, glass medium - pressure reaction vessels, and metal high - pressure reaction vessels are available.

4. It has an external database, and one system supports the interchange of multiple sets of reaction vessels.

5. With high - power heating and cooling units, it features fast temperature - control response and high - precision exothermic measurement.

6. Automatic feeding control can be achieved by mass or volume measurement.

7. It can accurately measure and obtain process - safety - related data such as reaction heat - flow, reaction enthalpy, conversion rate, specific heat capacity of samples, adiabatic temperature rise, and the maximum temperature that an out - of - control system can reach.

8. The Chinese - language software platform enables flexible programming of experimental procedures, real - time monitoring of key reaction - process data, and online modification of experimental procedures and parameters.

9. It can display the temperature inside the vessel, the temperature of the jacket, the mass of the added materials, and other test status information in real - time.

10. The safety thresholds for key parameters and states can be set. In case of reaction out - of - control, it can be quickly cooled with one click, and it will issue alarms and automatically stop in case of abnormal states, effectively ensuring safety.

11. It can automatically generate and save charts and data, and export experimental analysis reports.

Technical Parameters

Calorimetry Method

Heat Flow, Power Compensation (Option), Reflow (Option)

Temperature Control

Kettle Temperature Range

(-25 to 200)°C

Oil Bath Temperature Range

(-45 to 250)°C

Control Method

Isothermal, Constant Temperature, Scanning

Temperature Resolution

1.0mK

Temperature Control Precision

±0.1K

Silicone Oil Circulation Speed

(35 to 76)L/min

Control

Driver Voltage Range

(0 to 50)VDC

Driver Maximum Current

3.0A

Heater Maximum Power

120W

Atmospheric Pressure Glass Kettle (Option)

1000mL, other volumes available

Reactor Volume

Atmospheric Pressure

Working Pressure

Glass

Reactor Material

Medium-Pressure Glass Kettle (Option)

Medium-Pressure Glass Kettle (Option)

Reactor Volume

0.6MPa or 1.2MPa

Working Pressure

Glass

Kettle Material

316L Stainless Steel or Hastelloy

Kettle Lid Material

1000mL, other volumes customizable

High-Pressure Metal Reactor ( (Optional)

10MPa

Reactor volume

316L Stainless Steel or Hastelloy

Working pressure

Stirrer

Reactor material

2000r/min, 3000r/min optional

Stirrer

Maximum speed

Anchor or Paddle

Maximum torque

PTFE, 316L, or Hastelloy

Stirrer paddle form

Sampling system

Stirrer paddle material

Range 3100g, Accuracy 0.01g

Sampling system

Precision balance

Medium-pressure pump: electromagnetic diaphragm pump, the maximum flow rate of 2L / h; high-pressure pump: precision piston pump, the maximum flow rate of 2.4L / h

Sampling channel

3*400V/50Hz(±10%)/20A

Sampling pump

7000VA

Other parameters

1200mm*600mm*1850mm

Power supply

600mm*700mm*1300mm

Power

210kg

Test area size

300kg

Oil bath size

Heat Flow, Power Compensation (Option), Reflow (Option)

Oil bath weight

Temperature Control

Total machine weight

(-25 to 200)°C

Accessoriess

Optional reactor types: Glass atmospheric pressure reactor, glass medium pressure reactor, metal high pressure reactor

Functional components: High-power heating/cooling unit, automatic feeding control unit (supports mass/volume measurement)

Auxiliary modules: External database module, Chinese experimental software platform

Test Procedures

1. Pre-Experiment Preparation

Select a suitable reactor (glass atmospheric/medium pressure or metal high pressure)

Install the stirrer, temperature sensor, and condenser

Connect the heating/cooling circulation system

Initialize the software and set experimental parameters (temperature, stirring rate, injection strategy)

2. Calibration Procedure

Heat Flow Method: Baseline measurement → Power calibration → Baseline recovery → Recalibration (obtain UA value)

Power Compensation Method: Set jacket temperature → Built-in heater compensates to maintain zero temperature difference → Directly measure compensated power

3. Reaction Testing

Add the base material and start stirring for preheating

Pre-reaction calibration (establish heat loss baseline)

Start the reaction (manual/automatic feeding), monitor temperature, pressure, and heat flow in real time

Post-reaction calibration (calculate total heat release)

Data Recording: Enthalpy change, adiabatic temperature rise, maximum reaction rate time (MTR)

4. Safety Precautions

High-pressure experiments must use a dedicated reactor (such as a Calvet reactor). (Type) Strongly exothermic reactions require safety interlocks and emergency cooling. Wear protective equipment; ensure good ventilation in the experimental area. Regularly check the oxygen bomb's seal (it is recommended to replace the sealing ring every six months).


Отправить запрос, связаться с поставщиком

Кому: Standard International Group (HK) Limited
Ваш E-mail:
Текст письма:


Send to other suppliers

Другие товары поставщика

Gakushin Tester
Gakushin Tester ThisGakushin Type Rubbing Testeris a standard testing machine of the Type II rubbing tester (GAKUSIN TYPE) specified in JIS L0849Test Method of Rub...
Spectrodensitometer
Spectrodensitometer Standard The SpectroDensitometer YD5000 complies with international standards includingISO 5-4.CIE No.15. andISO 13655. ensuring reliable and stan...
Shoe heat insulation testing machine -- sand bath
Shoe heat insulation testing machine -- sand bath The sand bath component of the shoe heat insulation testing machine is primarily used to test the heat insulation or thermal insulation performance...
Taber Linear Abraser
Taber Linear Abraser TheTABER Linear Abraser (Abrader) - Model 5750has been engineered to evaluate product’s abrasion resistance, scratch resistance (single or mu...
Aluminum Ointment Tube Toughness Tester
Aluminum Ointment Tube Toughness Tester The ductility tester for aluminum tubes is a device specifically used to detect the ductility indicators of packaging materials such as aluminum so...
Все товары поставщика

Похожие товары

Air Permeability Tester STD- F8
Air Permeability Tester STD- F8 Продавец: Standard International Group (HK) Limited The Air Permeability Tester is designed to quickly and accurately measure the breathability and p...
Surface Performance Tester STD-FB4 - A
Surface Performance Tester STD-FB4 - A Продавец: Standard International Group (HK) Limited Introduction TheSurface Tester analyzes finger movements-particularly, strokes-performed by arti...
Compression Tester STD - FB3 - A
Compression Tester STD - FB3 - A Продавец: Standard International Group (HK) Limited TheCompression Tester analyzes hand movements-particularly, pushing with a finger-performed by ar...
Single - Strand Bending Tester STD - FB2 - SH
Single - Strand Bending Tester STD - FB2 - SH Продавец: Standard International Group (HK) Limited Standards ASTM A370. ASTM A416. ISO6934.4 The Single Fiber Bending Tester makes it possible to ...
Bending Tester STD- FB2 - S
Bending Tester STD- FB2 - S Продавец: Standard International Group (HK) Limited Compliance Standards ASTM D790 (Plastics Flexural Properties) ISO 178 (Flexural Strength of Rig...