Catheter Leakage and Compliance Tester
The Catheter Leakage and Compliance Tester is a professional testing device used to evaluate the expansion capacity (compliance) and leak resistance of medical tubing (such as infusion lines, vascular catheters, and respiratory tubing) under varying pressure conditions.
Applications
The catheter leakage and compliance tester is mainly used to detect the leakage and compliance of medical catheters (such as the breathing circuits of ventilators and anesthesia machines). Specifically, it can: Leakage Test: By applying a certain pressure inside the catheter and monitoring the rate or extent of pressure drop, it determines whether there is leakage in the catheter. This is crucial for ensuring the sealing and safety of the catheter during medical procedures. Compliance Test: Compliance refers to the ability of the catheter to deform when subjected to external forces. The tester applies a certain pressure inside the catheter and measures the volume change of the catheter to evaluate its compliance. This is of great significance for understanding the flexibility and elasticity of the catheter, as well as its applicability in medical applications.
Standards
ISO 13988-1:2017: Intravascular catheters - Part 1
ISO 8536-4:2016: Medical infusion equipment
YY/T 0285.4-2017: Intravascular catheters - Part 4
YY/T 1551.1-2017: Non-intravascular catheters
Features
Color LCD large screen display: Real-time display of test data and statistical values, intuitive and clear.
Intelligent touch control operation: Simple human-machine interface for quick experiment switching.
High-precision electronic flow meter: Achieves precise flow control, avoiding testing errors.
Top-tier sensor system: High-sensitivity data acquisition ensures stable and consistent test results.
Data security management: Four-level access control and encrypted storage prevent unauthorized operations and ensure the integrity and traceability of results.
Convenient result output: Equipped with a micro printer, supporting instant printing of single test results and statistical data.
Technical Parameters
| Parameter | Specification |
|---|---|
| Pressure Range | Determined by test object & application scenarios:- e.g., 0 - 300kPa for medical breathing circuits- Narrower range (e.g., 0 - 50kPa) for routine compliance testing |
| Pressure Accuracy | Key for test accuracy; generally requires ±0.1% or higher (e.g., 0 - 50kPa pressure sensor can reach one-thousandth accuracy) |
| Flow Range | Covers various flow situations of tested catheters; common flowmeter range: 0 - 100L/min |
| Flow Accuracy | Important indicator; generally requires ±1% or higher (e.g., ±1%FS for 2L/min mass flow controller) |
| Volume Measurement Range and Accuracy | - Range: From a few milliliters to several liters (adapts to different catheter volumes)- Accuracy: Generally needs to reach ±0.1ml or higher (to accurately calculate catheter compliance) |
| Test Time Range | - Leakage tests: Flexible setting (a few seconds to several hours)- Compliance tests: Controls pressure change rate & time interval to capture catheter volume changes at different pressure stages |
Accessoriess
Standard accessories include: specialized connectors for different sizes of medical tubing (such as ventilator/anesthesia machine breathing circuits), pressure/flow test tubing, tubing fixing clamps (to ensure sample stability during testing), standard test samples for equipment calibration (including standard leakage and compliance components), operating manual, and calibration certificate.
Test Procedures
Sample Pretreatment
Tubing Aging Treatment: Immerse in physiological saline at 37°C for 24 hours to relieve material stress;
Connection Sealing Inspection: Apply soapy water to the interfaces to confirm no initial leakage.
Typical Test Modes
Static Compliance Test:
Pressurize to 100kPa, maintain pressure for 30 seconds, and record the volume change ΔV1;
Release pressure to 0kPa, and record the volume recovery ΔV2;
Compliance calculation: C = (ΔV1 - ΔV2)/ΔP (unit: mL/kPa).
Dynamic Leakage Test:
Apply a sinusoidal pressure wave with a frequency of 1Hz and an amplitude of ±50kPa for 10 minutes;
Collect the leaked fluid and calculate the leakage rate using the weighing method (≤0.5mL/10min is considered acceptable).
Result Evaluation Standards (Refer to YY/T 0285.4-2017)
Compliance Deviation: ≤±15% compared to the nominal value;
Leakage: ≤1mL/minute under positive pressure of 100kPa, and no visible bubbles under negative pressure of -50kPa.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Metal Bone Screws Self - tapping Tester | Standard ISO 6475: Implants for surgery – Wear of total hip-joint prostheses ASTM F543: Standard Specification and Test Methods for Metalli... | |
| Civil Aviation Aircraft Interior Non-metallic Materials Flame Retardant Property Tester | Compliance Standards FAR25.853 (FAA Airworthiness Standards for Transport Category Aircraft) CCAR-25-R4.853 (Chinese Civil Aviation Regulations App... | |
| Safety Helmets Acceleration Impact & Penetration Tester | Standard GB/T 2812: Safety Helmets (China National Standard) Equivalent Compliance: ISO 6220-1 (Industrial Safety Helmets – Impact Absorpti... | |
| Electrostatic Decay Tester | Production: Electrostatic Decay Testeris professional designed to determine the attenuation of static electricity, corona discharge time, static v... | |
| Carpet Dimensional Stability Tester | Standard ASTM D5417(Standard Test Method for Assessing Carpet Dimensional Stability) ISO 2551(Textile Floor Coverings - Determination of Dimens... |
Похожие товары
| Full-Body Garment Stress Tester,FZ/T 70015 | Продавец: Standard International Group (HK) Limited | TheFull-Body Garment Stress Testeris a precision instrument designed to quantify the pressure exe... | |
| Industrial Dry Cleaning Machine,industrial garment steamer | Продавец: Standard International Group (HK) Limited | TheIndustrial Dry Cleaning and Energy-Saving Dryer Machinesare designed for high-performance, rel... | |
| MST Medical Compression Stocking Tester,compression sock tester | Продавец: Standard International Group (HK) Limited | The MST Medical Compression Stocking Tester is a professional instrument designed to evaluate the... | |
| MMT Liquid Water Separation Tester,MMT Liquid Water Analysis Tool | Продавец: Standard International Group (HK) Limited | Applications Textile Manufacturing:Testing waterproof fabrics for outdoor gear R&D:Developi... | |
| Dental material color stability tester | Продавец: Standard International Group (HK) Limited | Application The applications of color stability testers for dental materials are mainly reflecte... |