Winner2008 wet laser particle size analyzer
Winner2008A full automatic wet laser particle size analyzer adopt MIE scattering principle, measure size is from 0.01 to 2000 micron, Which offerreliable and repeatable particle size analysis for a diverse range of applications.It use dual-beam& multiple spectral detection systems and side light scatter test technology to significantly improve precision and performance of test, on behalf of the domestic advanced level in the field.
Main Specifications:
Model Name Winner2008A Winner2008D
Standard ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009
Principle MIE scattering principle
Measuring Range 0.01-2000μm 0.01-1200μm
Channels Number 127 127
Accuracy <1% (Deviation of D50 on national standard sample)
Repeatability <1% (Deviation of D50 on national standard sample )
Light source High performance He-Ne Laser (λ= 632.8nm, P>2MW)
Lifetime>25000hours
Dispersion
Method Ultrasonic Frequency:40KHz Power:35W, Time: ≥1S
Stir Revolutions Speed: 0-300RPM (Adjustable)
Circulate Rated Flow:8L/min Rated Power:10W
Sample Pool Volume:350mL
Micro-
Sample Pool Volume: 10mL ( Available)
Operation Mode Full automatic/ manual control, freely choose
Optical bench alignment system Full automatic,precision is up to 0.1um
Software function Analysis mode Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc.
Statistic Method Volume Distribution, Quantity Distribution
Statistic Comparison Several Testing Results of samples
Different batches of samples testing result,
Samples before and after processing,
Test result of samples in different time.
User-defined Analysis Figure out percentage according to the particle size
Figure out particle size according to the percentage
Figure out percentage according to the particle size range
Meet demands of representation of particle test in different industries
Test Report Word, Excel,Photo( Bmp), Text etc
Multiple-language Support Multiple language Support
Intelligent operation Automatically control water inflow, dispersion,test and analysis.Better Repeatability after remove human-factor
Testing speed <2min/time
Outer dimension L66cm*W32cm*H40cm
Net weight 65Kg
Main Features:
1)Winner Patent -Optical path design
Converging light Fourier transform light patented technology, make scattering light not be restricted to lens aperture. Double spectrum design make its precision achieves from 0.01um to 2000μm,and Dual-laser orthogonal light make use of the semiconductor auxiliary laser extend the test angle from 45 degree to 135 degree, ensure receive all the angles of signals.
2) Expand Size Range
Winner2008A measurement range significantly expanded to 0.01um, particularly suitable for small samples of the user, the sub micron particle test results reached the international level
3)Full built-in dispersion system
The ultrasonic dispersion, mechanical stirring and circulation channel reasonable integrated in the interior of the instruments, ensure the uniform dispersion and distribution of particles in the testing process, effectively avoid uneven distribution, large particle deposition phenomenon caused by long pipe line of outside dispersion system , to ensure the accuracy of the test results.
4)Intelligent Operation mode
With intelligent automatic mode of operation, to achieve a key test, as long as according to the prompt addition of sample, click the "test", all processes including water-supply, dispersion, circulation, testing, cleaning, data record, data analysis, save and print are automatically completed, not only reduce the testing workload, but also eliminate the interference of human factors, to further improve the accuracy and authenticity of testing results.
5)Automatic Optical path alignment system
The precision of four hybrid stepping motor in the automatic system of optical components, micro precision of 0.1um, the instrument of light path is always at its best to eliminate manually on the light path and the troubles and difficulties but also enhance the accuracy and stability of test results.
6)Authoritative calibration method
In the entire test range using the national standard materials for calibration, only a year calibration once the instrument can. Calibration method is simple and quick.
7) Unconstrained free fitting technique
Winner original unconstrained free fitting technology, particle size analysis is not affected by any functional constraints, can be a true reflection of the distribution of particles.
Adopt Patents Technology:
1.Optical bench design is protected by patent No.- ZL 2014 2 0378380.8,
2.Optical bench alignment system is protected by patent No.- ZL 2013 2 0835882.4.
3.Mie scattering principle application patent No.- ZL 2013 2 0812021.4.
4.Dual laser beam orthogonal application is protected by patent No.-ZL 2007 2 0025702.0
Application:
Winner2008 is widely used in mineral materials, cement, ceramics, chemicals, paint, emulsion, dyes, pigments, fillers, chemical products, catalysts, drilling mud, abrasive, lubricant, coal, sand, dust, bacteria, cells, food, additives, pesticides, explosives, graphite, photosensitive materials, fuel, and ink metal and non metal powder, calcium carbonate, kaolin, coal water slurry and other powder material.