Heat Flow Method for Thermal Conductivity Measurement

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The instrument complies with and references the following international and national standards:

ASTM D5470-2017

Standard Test Method for Thermal Transmission Properties of Thin Thermally Conductive Solid Electrical Insulation Materials

MIL-I-49456A

Insulation Sheet Materials, Thermally Conductive, Resin-Based and Glass Fiber Reinforced

GB/T 5598-2015

Method for Determination of Thermal Conductivity of Beryllium Oxide Ceramics

GB/T 29313-2017

Test Method for Thermal Conductive Properties of Electrical Insulating Materials

Parameters

ItemSpecification
Thermal conductivity range 0.01–50 W/m·K; 5–500 W/m·K (high-conductivity mode, automatic software switching)
Thermal resistance range 0.05–500 cm²·K/W
Sample size (standard) Φ30 mm or 20 × 20 mm
Optional sample sizes Φ15 mm, Φ50 mm, 25.4 × 25.4 mm (customized by contract)
Sample thickness range 0.001–50 mm (typical: 0.02–20 mm)
Hot plate temperature range Room temperature to 99.99 °C (standard)
Optional hot plate range Room temperature to 299.9 °C / 500 °C
Cold plate temperature range 0–99.99 °C
Cold plate control accuracy 0.01 °C
Pressure range 0–1000 N
Pressure control Servo motor control, accuracy 0.1 N
Thickness measurement range 0–50.00 mm
Thickness resolution 0.001 mm, automatic measurement
Number of samples 1 piece (thin films or stacked layers supported)
Measurement error ≤3% (thermal conductivity & resistance)
Contact thermal resistance error ≤5%
Power supply AC 220 V / 50 Hz
Power consumption 1000 W

Features

Designed and manufactured fully in accordance with ASTM D5470-2017

Automated servo motor pressure control with precise force setting and holding

Automatic thickness measurement with high resolution

Six-point temperature gradient detection for improved measurement accuracy

Thermal protection structure around the test rod to reduce environmental influence

Capable of measuring thermal resistance curves under different pressures and temperatures

Optimized mathematical models for thermal conductivity and contact thermal resistance

Automatic cold-end temperature compensation without ice-water compensation

Fully computer-controlled operation with data acquisition, analysis, and reporting

Available in standard floor-standing and compact benchtop configurations

Accessories

Main testing unit

Thermal analysis software (Chinese and English versions)

Precision low-temperature constant-temperature water bath (0.01 °C accuracy)

Computer (user-selectable)

Calibration reference samples (2 pieces)

Sample fixtures and accessories (solid sample molds, powder, paste, and grease sample cells)

Test Procedures

Select the appropriate test mode according to material type and test objective.

Place the solid sample between the hot and cold plates; stack thin samples if required.

Set pressure, temperature, and test parameters via the control software.

Apply pressure automatically using the servo motor system.

Measure sample thickness automatically and initiate the test.

Record temperature gradients and heat flow data during steady-state measurement.

Calculate thermal conductivity, thermal resistance, and contact thermal resistance.

Output test curves and reports through the software.

Maintenance Information

Keep hot and cold plates clean and free from surface damage

Regularly inspect pressure and thickness measurement systems

Verify system accuracy using reference samples when required

Maintain stable water bath operation for cold-end temperature control

Store accessories and sample fixtures properly when not in use


在线联系供应商

至: Standard International Group (HK) Limited
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

ISO 11155 Filter Test System
ISO 11155 Filter Test System Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system complies with: (1) ISO 11155-1 – Test method for particulate f...
Professional Aviation Fuel Testing Laboratory Equipment Supplier
Professional Aviation Fuel Testing Laboratory Equipment Supplier Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com E-mail:info@standard-groups。com Parameters Parameter CategoryDetails ...
Athletic Footwear Shock Attenuating Tester
Athletic Footwear Shock Attenuating Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard SATRA TM142 ASTM F1614 ASTM F1976 Product Advantages 1. Uncompromising Precis...
Contact Lens Oxygen Permeability Tester
Contact Lens Oxygen Permeability Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Contact Lens Oxygen Permeability Tester (Polarography Method) is a precision instrument...
Textile far-infrared radiation temperature rise test instrument
Textile far-infrared radiation temperature rise test instrument Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Textile Far Infrared Radiation Temperature Rise Tester is used for various te...
供应产品

Same products

ER4047 Aluminum Mig Welding Wire
ER4047 Aluminum Mig Welding Wire 卖方: 868776 Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing reliable welding material...
ER4043 Silicon Aluminum Welding Wire
ER4043 Silicon Aluminum Welding Wire 卖方: 868776 Hangzhou Kunli Welding Materials Co., Ltd. specializes in manufacturing high-quality welding mate...
Choosing the Right Equipment for Heavy-Duty Cleaning
Choosing the Right Equipment for Heavy-Duty Cleaning 卖方: Taizhou Feike Mechanical Technology Co., Ltd. FEIKE’s High Pressure Washer is designed for industrial cleaning, equipment maintenance, ve...
ER4043 Silicon Aluminum Welding Wire
ER4043 Silicon Aluminum Welding Wire 卖方: 868776 Hangzhou Kunli Welding Materials Co., Ltd. offers ER4043 Silicon Aluminum Welding Wire for alumin...
HDPE高速水冷双壁波纹管生产线
HDPE高速水冷双壁波纹管生产线 卖方: Shandong HUASU Plastic Technology Co.,Ltd. 机头模具采用双流道结构、氮化磨光处理机头,偏壁调整方便快捷,保证内外层壁厚均匀和稳定成型。 最新设计密闭压力水冷冷却结构,冷却充分,多点分布真空系统,效率更高,生产速度更快。 双层在线扩口功...