Fully Automatic Closed Mouth Flash Point Tester
Meet the standard:
ASTM D93
ISO 2719
IP 34
ASTM D7236
Product Introduction:
The flash point is a critical parameter for classifying flammable and combustible liquids, essential for safety regulations related to transportation, storage, and handling. The closed-cup method, such as the Pensky-Martens method, is widely used for fuels (like diesel, kerosene), lubricating oils, solvents, and other liquids with varying viscosities.
In a fully automatic closed-cup flash point tester, a sample of the liquid is placed into a test cup, which is sealed with a lid except for a shutter that can be opened to introduce the ignition source. The instrument's heating system raises the sample temperature at a precise, controlled rate as specified by the relevant standard.
A stirring mechanism may operate throughout the test to ensure uniform temperature distribution. At predetermined temperature intervals, the automatic system opens the shutter, introduces a small flame or electric igniter into the cup, and then closes the shutter. A sensor (thermal, ionization, or optical) monitors the vapor space for a flash (a brief ignition). The temperature at which a flash is detected is recorded as the flash point. The test is complete when a flash is detected or the maximum test temperature is reached. Fully automatic testers perform all these steps autonomously and can often cool down quickly between tests for high throughput.
The instrument is used in laboratories for:
Safety classification of flammable and combustible liquids.
Quality control of petroleum products, solvents, and chemicals to ensure they meet flash point specifications.
Receiving and shipping inspection of bulk liquids.
Research and development of new liquid formulations.
Compliance testing with national and international safety standards.
Product Advantages:
Fully automatic operation reduces operator labor, minimizes human error, and improves test repeatability.
Complies with key international standards for closed-cup flash point testing (ASTM D93. ISO 2719).
Provides accurate and reliable determination of flash point, a critical safety parameter.
Enhanced safety features protect the operator when testing flammable materials.
Increased sample throughput compared to manual or semi-automatic testers.
Digital display and data management capabilities for easy recording and reporting of results.
在线联系供应商
Other supplier products
| Hydraulic Oil Filtration and Dewatering Tester | Applications: Hydraulic and lubrication system oil purification and filtration; Hydraulic and lubrication system oil degassing; Hydraulic and lu... | |
| Spontaneously Igniting Solids Tester | Product Information: Product Features 1. It adopts an embedded processor and the Windows CE operating system. 2. With an 8 - inch liquid - crysta... | |
| Safety Helmet Impact Acceleration Tester | Purpose This instrument is suitable for impact - collision tests of helmets. It is a testing instrument used to evaluate the impact - resistance p... | |
| KES - FB2 - SH Single - Strand Bending Tester | Content The Single Fiber Bending Tester makes it possible to easily measure and quantify softness, elasticity, and diffusion differences in a si... | |
| Rapid screening Calorimeter | Product Introduction: The rapid - screening calorimeter is a professional calorimetric instrument for the quick assessment of reaction thermal haz... |
Same products
| Large Naïve VHH Library | 卖方: AlpVHHs | Large Naïve VHH Library VHHare single-domain antibodies derived from the variable regions o... | |
| PH-2398 pH/EC/ORP/温度 多参数测试仪 | 卖方: 科立龙电子股份有限公司 | 大显示,可同时进行水质PH、氧化还原电位、电导率、温度的测量; 带背光LCD显示清晰,功能齐全,精度高; 仪表采用微电脑测量与控制技术,具有自动校准功能,方便实用。 内置蓝牙,可以与电脑或... | |
| ORP-2026S 防水型 土壤ORP&温度 二合一测试仪 | 卖方: 科立龙电子股份有限公司 | 范围 ORP -999~999mV Temp 0~50℃ 分辨率 ORP 1mV... | |
| 防水型 EC /TDS/温度三合一参数测试仪 | 卖方: 科立龙电子股份有限公司 | 范围 EC ~19.9mS TDS Temp 0~50℃ 分辨率 ... | |
| EC-98330 电导率营养测试棒 | 卖方: 科立龙电子股份有限公司 | 本营养测试仪适用于水培和农业应用,能够准确测量水中的营养液电导率(EC)单位为mS/cm,以及总溶解固体(TDS)转换为百万分之一(ppm)。 适合需要较高EC值的植物,该测试仪具有高达6.0... |