Catheter Flow Meter
The Catheter Flow Meter is specifically designed for measuring the water flow rate of medical catheters, including non-vascular catheters, intravascular catheters, and single-use sterile catheters. It is also known as a catheter flow meter, catheter flow detector, non-vascular catheter flow device, infusion set flow tester, or medical device flow tester.
Applications
The catheter flow tester is mainly used to measure and evaluate the flow performance of medical catheters, such as intravascular catheters, non - intravascular catheters, and infusion set catheters. In the medical field, this tester has a wide range of applications, including but not limited to the following aspects: Product R & D: During the design and R & D stage of catheter products, the tester can help engineers accurately assess the flow performance of different design schemes, providing powerful data support for product iteration. Quality Control: Through batch testing, the tester can ensure that each batch of catheters meets the established flow standards and performance requirements, thus improving the overall quality of products. Clinical Application: The tester can be used to monitor the flow changes of catheters in patients, providing real - time physiological parameter feedback to doctors, which is helpful for adjusting treatment plans, assessing the progress of the disease, and preventing complications.
Standards
GB/T 15812.1 - 2005 Non - Vascular Intravascular Catheters - Part 1: General Performance Test Methods
YY 0285.1 - 2017 Disposable Sterile Intravascular Catheters
Features
• Independently developed hydrostatic pressure testing device, capable of maintaining a liquid level of 1000 ± 10 mm.
• Flow rate of 525 ml/min, meeting standard flow rate requirements.
• Equipped with a constant temperature water bath to ensure the test water temperature is 23℃.
• Equipped with a timer for timed testing.
• Test fixture designed strictly according to standard requirements, easy to operate and highly efficient.
• Uniquely designed lifting and lowering structure allows for height adjustment, facilitating testing at different water pressures.
• Equipped with a balance station for convenient placement of balances or measuring cylinders.
• Equipped with an adjustable precision water pump for easy flow rate adjustment.
Technical Parameters
| Parameter | Specification |
|---|---|
| Testing Range | Usually within a weight/volume range (e.g., 1.00 - 600.00g, varying by catheter specifications) or a specific flow range |
| Accuracy | High testing accuracy (e.g., ±0.01%FS, one-ten-thousandth of full scale) or higher requirements |
| Static Water Level Height | e.g., 200±5mm, 1000±5mm (meets different testing needs); water pump has automatic circulation function; static water level is adjustable |
| Time Controller | Usually 0 - 60min, with an error ≤ ±1S; used to control testing time |
| Fixtures | Adapt to various catheter specifications/types, ensuring test versatility and flexibility |
| Flow Calculation | Generally uses the weighing method (imported flowmeters optional) to calculate flow, ensuring result accuracy |
| Flow Rate | e.g., 525ml/min (arbitrary selection) ±25ml/min, etc.; adapts to different testing scenarios |
| Intelligent Temperature Control | Equipped with a micro-computer intelligent temperature-control system, ensuring constancy and accuracy of test temperature |
| Special Accessories | Includes stainless-steel thermostatic water baths, special connectors, etc.; meets testing needs of different catheter specifications/types |
Accessoriess
Static pressure head, height adjustment device, constant temperature water bath, PU tubing
Test Procedures
1. Before testing, appropriate test conditions, such as water flow rate and pressure, should be set according to the pipe specifications and testing requirements. This helps to ensure the accuracy and reliability of the measurement results.
2. When connecting the pipes, ensure a tight, leak-free connection. Any leaks will affect the accuracy of the measurement results.
3. During the testing process, keep the testing instrument and pipes clean to avoid contamination affecting the test results. Regularly clean and maintain the instrument to ensure its proper functioning and extend its service life.
4. After the test is completed, turn off the testing instrument, disconnect the pipes, and clean and dry them for future use. This helps prevent corrosion and damage to the instrument and pipes due to prolonged disuse.
在线联系供应商
Other supplier products
| Minimum Ignition Temperature (MIT) – Dust Cloud Tester | Dust Cloud Minimum Ignition Temperature Tester is mainly used to assess the explosiveness of dust in the production and processing places such as a... | |
| Hot Tack Tester / Heat Seal Tester | Main Uses of the Hot - Tack Tensile Tester The hot - tack tensile tester is an instrument specifically designed to test the hot - tacking and heat... | |
| Bally Type Flexometer丨Bally Type Flexometer Meter | Standards EN ISO 20344:6.6. I.U.P.10+20. I.U.P.20. UNI 4818. DIN 53.351. UNE/I.U.P.20 Applications The Bally flexometer is the internationally-a... | |
| Spring torsion tester | The Spring Torsion Tester is designed to precisely measure torsional properties, including torque and angle, for a wide range of torsion springs an... | |
| Aerosol Foams Flammability Tester | Product Introduction This product complies with "Recommendations on the Transport of Dangerous Goods - Manual of Tests and Criteria" 31.6 Flammabi... |
Same products
| Gas Oxidizing Property Tester | 卖方: Standard International Group (HK) Limited | Product Description Conforms to ISO 10156-2010 “Gases and Gas Mixtures Selection of Cylind... | |
| Battery Powder Compaction Density Measurement System | 卖方: Standard International Group (HK) Limited | The Battery Powder Compaction Density Measurement System is a precision testing system designed... | |
| Battery Powder resistance tester | 卖方: Standard International Group (HK) Limited | The Powder Resistivity Meter is a precision testing system designed to measure electrical resisti... | |
| KES-SESRU surface roughness/friction tester | 卖方: Standard International Group (HK) Limited | TheRoughness/Friction Tester analyzes hand movements-particularly, strokes-performed by artisans ... | |
| Air Permeability Tester STD- F8 | 卖方: Standard International Group (HK) Limited | The Air Permeability Tester is designed to quickly and accurately measure the breathability and p... |