EM-5 Flue Gas Analyzer

Overview

EM-5 series flue online gas analyzeris a series of self-developed gas analysis products for online gas analysisof domestic as well as foreign environmental protection and industrial control. Based on UV Differential Optical Absorption Spectroscopy (DOAS) technology and chemometrics algorithm (PLS), it is able to measure the gas concentration of SO2, NO, NO2, O2, NH3, Cl2, O3, H2S, Ch3 and etc.

Features Of EM-5 Flue Gas Analyzer

High measurement accuracy

Minimum detection limit of SO2, NO, NO2 can reach 1mg/m3 (15m optical path)

Ultraviolet has no moisture absorption, undisturbed by moisture and dust

No cross interference between measured gases (refer to table 1)

Measure NO and NO2 simultaneously without converter

Low detection limit

Measuring gas

Interference

SO2

NO

NO2

O2

SO2 (500ppm)

/

<1ppm

no

no

NO (500ppm)

no

/

no

no

NO2 (500ppm)

no

<1ppm

/

no

H2O (no dew)

no

no

no

no

CO (1000ppm)

no

no

no

no

CO2 (20%)

no

no

no

no

O2 (21%)

no

no

no

/

High reliability

Small zero drift and span drift

Modular design

No optical moving parts, without vibration influence

Strong gas chamber, low cost

Automatic adjustment spectrum technology, long maintenance-free period

Adopt pulse light source with 10-year service life

Technical Parameter Of EM-5 Flue Gas Analyzer

Measurement Principle

UV-DOAS+PLS

Versions

Regular range

Low range

Ultra-low range

Ranges

Min range

Max range

Min range

Max range

Min range

Max range

Component

SO2

0-300ppm

0-3000ppm

0-100ppm

0-300ppm

0-50ppm

0-100ppm

NO

0-300ppm

0-3000ppm

0-100ppm

0-300ppm

0-50ppm

0-100ppm

NO2

0-500ppm

0-4000pp

0-300ppm

0-1000ppm

0-100ppm

0-200ppm

O2 (ECD or ZrO2)

0-5%

0-25%

0-5%

0-25%

0-5%

0-25%

Linearity

≤±3%F.S.

Repeatability

≤2%

Zero Drift

≤±2%F.S./7d

Span Drift

≤±2%F.S./7d

Response Time(T90)

≤50s

Working Temperature

-20℃~+50℃

Preheating Time

No preheating

20min

40min

Sample Gas Interface

Φ6 Bi-Lok

Sample Gas Flow

Range: (0.5~2)L/min, fluctuation<25%

Sample Gas Pressure

Current ambient pressure ±0.1Bar

SampleGas Temp.

0℃~+50℃

Sample Gas Humidity

No condensation (under sample gas temperature)

4-20mA Input

3 , configurable, 100Ω load

4-20mA Output

5, output content can be configured, maximum load capacity ≤800Ω

Switching valueInput

6 , configurable

Relay Output

14, output content can be configured. DC30V2A

Digital I/O

1×Rs232, 1×Rs485 (Support Modbus protocol)

Installation Method

19’panel

Supply Voltage

220VAC±10%

Rated Power

About 100W

About 120W

Application Of EM-5 Flue Gas Analyzer

Coal-fired power plant

Cement plant

Industrial furnace

DeSOX process monitoring

DeNOX process monitoring

Waste incineration plant

PVC process in chlor-alkali plant

Titanium dioxide production process

Sulfur recovery process

Natural gas purification process

Methyl iodide analysis for the coal chemical industry

On-line air monitoring

With features of high measurement accuracy, high reliability, fast response time, wide application scope, it can be widely applied in occasions of environment online monitoring, industrial process control, safety monitoring and etc.

Depended on years of flue gas analyzer research experience and combined with practical experience of thousands of successful application cases, the analyzer has derived multiple types (standard type, low emission type, ultra-low emission type, and customized type) to meet different index requirements in different working condition.

UV DOAS technology. The optical technology platform consists of a light source, gas chamber, optical fiber and spectrograph (including the diaphragm, a holographic grating, linear array detector) and other optical components.

Ultraviolet light emitted by light source enters the gas chamber by an optical window and absorbed by measured sample gas flowed through the gas chamber. The light carrying sample absorption information will be gathered by the lens and coupled into an optical fiber and then transmits into spectrometer through fiber. After light splitting and photo-voltaic conversion, the gas absorption spectrum is obtained.

The corresponding component concentration in the gas can be calculated by analyzing the spectrum.

Note: the difference among standard analyzer, low emission analyzer, and ultra-low analyzer is the different optical path length of gas chamber; for standard analyzer, it adopts dual lens collimation system with 0.25m optical path; for low emission analyzer, it adopts multiple return structure with 1m optical path; for ultra-low emission analyzer, it uses white cell structure.

Zetian Group has high-quality continuous flue gas analyzerfor sale, flue gas analyser for saleand gas analyzer for sale,contact us now.

Our company is good at flue gas analyser working principleand we also produce air quality monitor,online flue gas analyzer,online gas analyzerand online gas analyser.


在线联系供应商

至: HANGZHOU ZETIAN TECHNOLOGY CO., LTD
Your E-mail:
消息正文:


Send to other suppliers

Other supplier products

Continuous Emission Monitoring & Process Control
Continuous Emission Monitoring & Process Control What is the continuous emission monitoring system working principle? The instruments and analyzers produced by Zetian now are widely used at home a...
EM-5 Flue Gas Analyzer
EM-5 Flue Gas Analyzer Overview EM-5 series flue gas analyzer is a series of self-developed gas analysis products for online gas analysisof domestic as well as foreign en...
EM-5P Portable Flue Gas Analyzer
EM-5P Portable Flue Gas Analyzer Overview Of EM-5P Portable Flue Gas Analyzer EM-5P portable flue gas analyzersystem mainly consists of gas monitoring sensors (SO2, NOx, CO, CO2), ...
EQMS-3000B Environmental Quality Monitoring System
EQMS-3000B Environmental Quality Monitoring System Overview According to the status and characteristics of construction sites and production industry generated pollution, applying more sophisticated...
WDet-5000 Total Phosphorus Online Automatic Analyzer
WDet-5000 Total Phosphorus Online Automatic Analyzer HANGZHOU ZETIAN TECHNOLOGY CO., LTDwas established in 2008, and as a fast-growing company specialized in environmental pollution monitoring and i...
供应产品

Same products

PH-A8防水笔式土壤PH计
PH-A8防水笔式土壤PH计 卖方: 科立龙电子股份有限公司 大屏幕背光,清晰显示,同时显示温度带自动温度补偿 主机防水等级IP65 两点按键校准,自动识别pH标准缓冲液;自动电源关闭,有效节省电池能耗。 两节七号大容量供电。适用于土壤、面团等。 ...
JWL-V6 Portable Physicochemical Meter
JWL-V6 Portable Physicochemical Meter 卖方: Shenzhen Joinwe Smart Co.,Ltd JWL-V6Portable PhysicochemicalMetercan detect the physical and chemical indicators of oil product...
SA-917 折叠盐度计
SA-917 折叠盐度计 卖方: 科立龙电子股份有限公司ctron Co.Ltd 范围 0~100.0ppt 准确性 ±2℅F·S 分辨率 0.1幻灯片 ...
SA-3287 防水盐度计
SA-3287 防水盐度计 卖方: 科立龙电子股份有限公司ctron Co.Ltd 范围 0~100.0‰ 准确性 ±2℅F·S 分辨率 0.1&perm...
SA-1397 防水盐度和温度计
SA-1397 防水盐度和温度计 卖方: 科立龙电子股份有限公司ctron Co.Ltd 自动温度补偿(带温度显示。)*IP55防水等级。