Laboratory Flame Spread Tester for Insulation Panels

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

This tester conforms to the following national and international standards:

GB/T 27904‑2011 — Test method for flame spread of interior decoration materials

GB/T 25207‑2010 — Classification of combustion performance of building decorative materials

GB 8624‑2012 — Classification of burning behavior of building materials and products

GB/T 16172‑2007 — Determination of heat release rate and total heat release of building materials and products (Cone Calorimeter Method)

BSS 7365 — UK aerospace standard for combustion and smoke density of aerospace composite and interior materials

AITM 2.0053 — European aerospace industry method for heat release rate and smoke evaluation

FAR Part 25 Appendix F Part VI — FAA regulatory requirement for flame propagation, smoke density, and toxicity in civil aircraft cabin materials

Parameters

ParameterSpecification
Dimensions 1220 × 570 × 520 mm; 1200 × 1150 × 1350 mm; 800 × 800 × 400 mm
Weight 185 kg
Thermal Radiation Source 400 VAC, 3-phase electric heating plate, PID control, ±0.1°C accuracy
Radiation Strip Width 76 mm, six strips, operational up to 704°C
Sample Angle 30° relative to horizontal plane
Ignition System Axially symmetric propane burner, 0.15 mm nozzle
Burner Height Flame at least 51 mm above sample plane
Flame Spread Measurement Laser pointer for accurate monitoring
Thermocouple Placement Back wall 279 mm, right wall 292 mm, below radiation plate 51 mm, ±2°C accuracy
Heat Flux Range ≥60 kW/m², with self-circulating cooling water
Heat Flux Mounting Steel plate 3.2 mm thick, with 3 × 25.4 mm insertion holes
Sample Holders Standard and short versions (for hook-and-loop tests)
Data Acquisition Computer and printer, automatic test timing ±0.1 s

Features

Enclosed test chamber: Includes integrated chimney and controlled radiation environment.

Precision thermal radiation source: PID-controlled electric heating plate ensures uniform, stable heat flux.

Automated sample handling: Pneumatic sliding drawer allows easy sample insertion and removal.

Accurate ignition and flame monitoring: Propane burner with laser pointer ensures precise flame spread measurement.

Thermocouple monitoring: Accurately records sample and chamber temperatures at multiple points.

High-performance heat flux system: Measures critical heat flux for ignition and flame propagation.

Data acquisition and automation: Test time, flame spread, and other parameters are recorded automatically and with high precision.

Safety and repeatability: Controlled combustion and automated monitoring ensure consistent and safe testing conditions.

Accessories

Enclosed radiation test chamber with integrated chimney

3-phase electric heating plate with PID temperature control

Pneumatic sliding drawer for sample insertion and removal

Propane burner with 0.15 mm nozzle

Laser pointer for flame spread measurement

Thermocouples and mounting accessories

Water-cooled heat flux measurement system

Standard and short sample holders

Computer and printer with data acquisition system

Test Procedures

Install the sample in the test chamber using the pneumatic drawer and proper sample holder.

Connect the propane supply and compressed air system.

Activate the radiation source to achieve the desired thermal flux and preheat the sample.

Ignite the sample using the propane burner positioned 51 mm above the sample plane.

Monitor flame spread using the laser pointer and record temperature, heat flux, and ignition parameters via the data acquisition system.

Test duration is automatically recorded and ends at the set time or upon completion of flame propagation.

After testing, remove the sample and inspect the chamber for maintenance and cleaning.

Maintenance Information

Regularly clean the test chamber, radiation plate, and propane burner to ensure uniform heating and accurate measurements.

Calibrate thermocouples, heat flux system, and data acquisition equipment periodically.

Verify the pneumatic drawer and sample holders are operating smoothly.

Check propane and compressed air supplies before each test.

Maintain ambient temperature between 10–35°C for consistent testing conditions.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

BS5852 Flammability Tester
BS5852 Flammability Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ◆UK Regulation British furniture (fire and safety) regulations test ◆BS 5852 Me...
Photovoltaic Acid Corrosion Test Chamber
Photovoltaic Acid Corrosion Test Chamber Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The Photovoltaic Acid Mist Test Chamber is a specialized device designed to s...
Safety Helmet Impact Acceleration Tester
Safety Helmet Impact Acceleration Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Purpose This instrument is suitable for impact - collision tests of helmets. It is a tes...
Bonding Strength Measuring Instrument
Bonding Strength Measuring Instrument The tack - holding tester, also known as the tack - holding property tester, tack - holding force tester, or holding - power tester, is a device sp...
Dusting solar cell modules Test System
Dusting solar cell modules Test System Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications tests according to VDI 3956 temperature-dependent dusting and cleaning tests...
All supplier products

Same products

ISO 29463 HEPA/ULPA filter performance testing System
ISO 29463 HEPA/ULPA filter performance testing System Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system supports and compli...
High-Precision ISO 16890 Filter Media Test Rig for R&D and QC
High-Precision ISO 16890 Filter Media Test Rig for R&D and QC Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system can execute and com...
Mobile Floor Scrubber Aging Life Tester IEC 60312
Mobile Floor Scrubber Aging Life Tester IEC 60312 Seller: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com TheMobile Floor Scrubber Aging L...
ISO 12103 Cleanable Filter Media Test System
ISO 12103 Cleanable Filter Media Test System Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system complies with: (1)...
Robotic Vacuum Cleaners Edge-Following Tester
Robotic Vacuum Cleaners Edge-Following Tester Seller: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com TheEdge-Following Test Equipment...