Airflow conditioning environmental chamber manufacturer
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
Internal compliance with laboratory safety, electrical, and environmental simulation standards.
Air exchange rates, temperature, and humidity control meet industrial and research-grade environmental chamber specifications.
Custom configurations may comply with client-specific ventilation or airflow standards.
Parameters
| No. | Item | Technical Specifications |
|---|---|---|
| 1 | Chamber Dimensions | External: 5.1 m (L) × 5.6 m (W) × 3.2 m (H) Internal: 4.9 m (L) × 5.4 m (W) × 2.6 m (H) Standard insulated door: 0.8 m × 2.0 m with double-glass observation window Internal lighting: 1 daylight lamp; power outlet: 2.5 kW Digital display panel for temperature and humidity Optional additional window available at extra cost |
| 2 | Temperature Range | 10–40°C ±0.5°C adjustable |
| 3 | Humidity Range | 35–90% ±5% adjustable |
| 4 | Air Change Rate | 3–16 ACH; optional single-pass fresh air supply available (installation-dependent, extra cost) |
| 5 | Electrical Control System | Centralized control panel; Supply: AC 380V/50Hz, max power ≤28.5 kW; Equipped with short-circuit and over-temperature protection |
| 6 | Airflow Organization Modes | 9 modes: Mixed Ventilation:(1) Top diffuser with side-lower return, (2) Side supply-side return, (3) Side supply-top return Displacement Ventilation:(1) Top return, (2) Side-top return Floor Ventilation:(1) Top return, (2) Side-top return Stratified Ventilation:(1) Top return, (2) Side-lower return Optional quick-connect supply fittings available (extra cost) |
| 7 | Working Environment | Standard laboratory conditions |
| 8 | Other Requirements | Refrigeration unit piping and electrical lines default 5 m; site preparation including leveling, wall openings, rain shelter, water supply, drainage, and main power line to control box to be provided by user; chamber panels: 100 mm PU double-color steel |
| 9 | Special Notes | Default specifications are factory standards; custom requirements exceeding these parameters require additional consultation and pricing |
Features
High-Precision Environmental Control: Maintains stable temperature and humidity with ±0.5°C and ±5% RH accuracy.
Multiple Airflow Modes: Supports mixed, displacement, floor, and stratified ventilation with nine programmable configurations.
Observation and Monitoring: Equipped with observation window, internal lighting, and digital panel for real-time monitoring of chamber conditions.
Customizable Ventilation: Air change rates adjustable; optional quick-connect fittings allow flexible supply modifications.
Safety and Reliability: Electrical protections including short-circuit and over-temperature safeguards.
Modular Installation Support: Refrigeration, water, drainage, and power line preparation allow smooth on-site assembly.
Accessories
Digital display panel for temperature and humidity monitoring.
Standard insulated chamber door with double-glass observation window.
Internal lighting (daylight lamp).
Optional quick-connect supply fittings for airflow customization.
Test Procedures
Chamber Preparation: Confirm power, water, and drainage connections; ensure refrigeration system is operational (~10 min).
Set Environmental Conditions: Program desired temperature (10–40°C), humidity (35–90%), and airflow mode via control panel or software (~5 min).
Load Samples/Subjects: Place equipment, test materials, or participants inside chamber; ensure proper positioning (~10 min).
Run Test: Activate test program; monitor environmental conditions in real-time; adjust airflow mode if needed (~variable duration).
Data Recording and Export: Record chamber temperature, humidity, and airflow data; save and export for analysis (~5 min).
Completion and Retrieval: End program; allow chamber to stabilize at safe conditions; remove samples or subjects safely (~10 min).
Maintenance Information
Regularly clean observation window, internal lighting, and control panel.
Inspect refrigeration unit, piping, and electrical connections for wear or damage.
Verify airflow systems and sensors for calibration and accuracy.
Maintain chamber in a clean, dust-free, and temperature-stable environment.
Send product request
Other supplier products
| Aerosol Foams Flammability Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction This product complies with "Recommendations on the Transport of Dange... | |
| Electric pencil hardness tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction: This machine is suitable for all kinds of finished product spray har... | |
| Leather thickness meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction Leather thickness tester is a special high-precision instrument for l... | |
| Advanced heat insulation test furnace for fireproof coatings | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application This test furnace is suitable for determining: 1. Thermal insulation efficien... | |
| Low cost Mechanical load testing machine for photovoltaic modules | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Mechanical Load Testing Machine for Photovoltaic Modules is designed to v... |
Same products
| Gasket for Plate Heat Exchanger | Seller: Hengshui Hongcan Import and Export Trading Co., Ltd | The sealing gasketin plate heat exchangers primarily serves to seal between hot and cold media. ... | |
| Energy Saver | Seller: O&K GROUP LIMITED | Looking for Quality Products? From electronics to home essentials, fashion to industrial supplies... | |
| ER1100 Aluminum Welding Wire | Seller: 868776 | The ER1100 Aluminum Welding Wire from Hangzhou Kunli Welding Materials Co., Ltd. is designed for ... | |
| air conditioner | Seller: O&K GROUP LIMITED | Looking for Quality Products? From electronics to home essentials, fashion to industrial supplies... | |
| MMA-120 Handheld Arc Welder | Seller: 871183 | The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con... |