Laboratory Electrospinning Device for Nanofiber Membrane Fabrication
Get a quote:
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standard
T/YNIA 017-2023 "Electrospun Nonwoven Air Filter Materials"
Technical Parameters
| Parameter | Specification |
|---|---|
| Dimensions / Weight | W900 × D750 × H1080 mm / approx. 120 kg |
| Power Supply | AC100V, max. 80 W (must be grounded) |
| Operating Environment | Temperature 10–40°C, humidity 30–70% RH, non-condensing |
| DC High-Voltage Power Supply | 0–39 kV digital display, leakage current Max. 500 μA |
| Exhaust Fan | Multi-blade (scirocco) fan |
| Injection Pump Capacity | Standard 20 ml, up to 3 pumps |
| Injection Speed | 0.000–0.500 mm/min (stable range 0.030–0.250 mm/min) |
| Injection Movement Speed | 0–300 mm/min |
| Injection Target Range | 0–200 mm (standard) |
| Injection Distance | Horizontal 45°, vertical adjustable up to 150 mm |
| Drum Size | Diameter 100 mm, width 330 mm |
| Drum Circumferential Speed | 0–9.00 m/min |
| Safety Measures | Spark limiter (stops above 500 μA), valve interlock |
Features
Controlled Fiber Diameter: produces fibers from 50 to 800 nm with high uniformity and reproducibility
Multi-Polymer Compatibility: small polymer quantities can produce nanofibers rapidly
High Safety Design: spark limiter and valve interlock ensure safe high-voltage operation
Controllable Injection System: adjustable pump speed, movement speed, and injection distance
Drum Collection Device: enables continuous winding or batch collection
Broad Application Scope: suitable for automotive, energy, electronics, biomedical, catalysis, and textile research
Accessories
| Accessory / Consumable | Function |
|---|---|
| Injection Pump (20 ml) | Nanofiber material injection and speed control |
| Drum Collector | Nanofiber collection and winding |
| Electrode Capillary | Electrospun fiber formation |
| High-Voltage Power Supply | Provides electrospinning electric field |
| Exhaust Fan | Removes solvent vapor and volatile substances |
| Safety Device | Spark limiter, valve interlock for operational safety |
Send product request
Other supplier products
| Diaper Absorption Rate Testing Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Diaper Absorption Rate Testing Machine is a specialized laboratory device used to evalu... | |
| Shoe Peeling Strength Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Shoe Peeling Strength Tester is designed to evaluate the adhesive bond strength between... | |
| Automatic dynamic scratch and wear tester system | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Primary applications Scratch, wear, and impact resistance testing for materia... | |
| Cleveland Open Cup Flash Point Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The instrument can automatically test the flash point and ignition point of petroleum produ... | |
| Intramedullary Nails Four - point Bending Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0591-2011: Chinese standard specifying bending fatigue test methods for intr... |