Import Infusion Pump Flow Rate Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
The infusion set flow rate tester is an instrument specifically designed for measuring and monitoring the flow rate of infusion sets. Its main uses include: - Flow Rate Measurement: Precisely measure the flow rate of the infusion set, usually displayed in milliliters per hour (ml/h) or drops per minute (gtt/min). - Flow Monitoring: Monitor the changes in flow rate in real - time during the infusion process, and promptly detect abnormal flow rate situations, such as the flow rate being too fast, too slow, or blocked, to ensure the safety and effectiveness of the infusion process. - Alarm Function: When the infusion flow rate exceeds the set normal range, the tester can emit audible and visual alarm signals to remind medical staff to take timely action and avoid potential medical risks. - Data Recording and Storage: Record and store information such as flow rate data and time during the infusion process, which is convenient for medical staff to view and analyze, and helps track and evaluate the patient's infusion treatment.
Standards
GB 8368 - 2018 : Disposable Infusion Sets
Technical Parameters
| Parameter | Specification |
|---|---|
| Testing Range | Wide flow rate testing range (e.g., 1 - 2000ml for some testers), adapting to different types/specifications of infusion sets (varies by catheter specifications) |
| Testing Accuracy | High accuracy for flow rate measurement (e.g., ±5ml/h or ±0.1gtt/min, etc.) to ensure test result accuracy |
| Static Water Level Height | Adjustable range: usually 0 - 1000mm, meeting requirements under different testing conditions |
| Temperature Control Range | Usually 5 - 90℃, simulating infusion situations under different temperature conditions |
| Fixtures | Adapt to various catheters; customized according to catheter specifications to firmly hold infusion sets of different specifications during testing |
| Flow Calculation | Adopts methods like the weighing method for flow calculation, ensuring accurate flow measurement |
| Alarm Threshold | Allows users to set the alarm threshold; triggers an alarm when the flow rate exceeds the set range |
Accessoriess
Standard accessories include: specialized fixing clamps for different types of infusion sets (for securely holding various infusion sets), connecting tubing (for setting up the flow path to complete flow rate testing), equipment calibration auxiliary parts (to ensure testing accuracy), operating manual, and calibration certificate.
Test Procedures
Infusion Flow Rate
For infusion sets with a drip rate of 20 drops/mL, the infusion set should deliver at least 1000 mL of sodium chloride solution [mass concentration p(NaCl) = 9 g/L] within 10 minutes under a static pressure head of 1 meter.
For infusion sets with a drip rate of 60 drops/mL, the infusion set should deliver at least 1000 mL of sodium chloride solution [mass concentration p(NaCl) = 9 g/L] within 40 minutes under a static pressure head of 1 meter.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Pressure Transmission Flow Meter Test Instrument | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications - Fluid System Performance Evaluation: The Pressure Transmission and Flow Tes... | |
| Oil Separator Test Rig,ISO 11155 oil droplet separation tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Oil Separator Test Rigis designed for automated testing procedures in order to obtain r... | |
| Photovoltaic module outdoor exposure test system | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Photovoltaic Module Outdoor Exposure Test System is designed to simulate real-world env... | |
| Burn rate evaluation equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com This product complies with "Recommendations on the Transport of Dangerous Goods - Manual of... | |
| Testing apparatus for back and butt forms evaluation | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com 5th Percentile Female SAE Butt Form The 5th Percentile Female SAE Butt Form ... |