Parameters
| Parameter | Specification |
|---|---|
| Dimensions | 950 × 600 × 870 mm |
| Weight | 105.5 kg |
| Heating Devices | 2 × 1000 W quartz heaters providing thermal radiation |
| Burner Type | T-shaped tubular burner |
| Sample Holders | 3 types, stainless steel wrapped |
| Thermocouples | Continuously measure temperature difference between chimney and room |
| Power Output Control | Automatic adjustment according to test timeline |
| Data Acquisition | Computer-based, report generation |
Features
Durable stainless steel frame: Ensures structural stability and long-term reliability.
Silicate combustion chamber: Provides uniform and controlled testing environment.
Multiple sample holders: Accommodate different test configurations securely.
Quartz heaters: Deliver consistent thermal radiation to samples.
T-shaped burner: Provides precise direct flame for small flame spread tests.
Continuous temperature monitoring: Thermocouples record chimney versus room temperature.
Automatic power control: Heater output adjusts automatically according to test timeline.
Integrated data acquisition: Computer automatically processes data and generates test reports.
Accessories
Quartz radiation heaters (2 × 1000 W)
T-shaped tubular burner
Stainless steel sample holders (3 types)
Silicate combustion chamber
Thermocouples and mounting accessories
Integrated computer and printer with data acquisition system
Test Procedures
Install the sample in the combustion chamber using the appropriate sample holder.
Connect the G112 test gas supply.
Start the quartz heaters and T-shaped burner, providing thermal radiation and flame exposure.
Adjust heater power automatically according to the test schedule:
First 2 min 45 s: initial burner exposure at 530 J/s
At 2:45 min: increase total heater power to 1800 W
At 5 min: reduce power to 1500 W and maintain until test end
Record flame propagation, temperature, and FSI using the thermocouple and data acquisition system.
Total test duration is 20 minutes. Upon completion, safely remove the sample and review results.
Maintenance Information
Regularly clean the combustion chamber and burners to ensure consistent heat flux.
Calibrate thermocouples and power output control systems periodically.
Inspect sample holders and chamber integrity before each test.
Ensure gas supply is stable and leak-free.
Maintain ambient temperature between 10–35°C for consistent testing conditions.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Industrial Oil Separator Test Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards ISO 11155 series – air filtration and separator standards DIN 71460-1/2 &... | |
| Toyota Abrasion Tester for Automotive Materials | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Toyota Special Abrasion Tester is a precision wear testing instrument des... | |
| Carpet tuft length measuring instrument | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard BS 4223:1989 (British Standard): Specifies methods for determining the pile len... | |
| Advanced suture needle cutting force apparatus | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards YY/T 0043-2016:Medical suture needle YY 0166-2002:Surgical suture needles with ... | |
| Gas Hazardous Characteristics Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction This tester is used to test the flammable concentration range of gase... |