Rubber sleeve rebound elasticity tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Application
The tester is applicable to a variety of medical and laboratory products, including:
(1) Collection tubes and blood sampling needles – assessing rubber sleeve rebound force to ensure secure sealing and consistent operation.
(2) Pre-filled syringes – verifying the compression and resilience of protective rubber stoppers.
(3) Catheter and infusion system components – evaluating the elasticity and durability of rubber components.
(4) Quality control in production – ensuring each batch meets predefined resilience specifications.
(5) Research and development – optimizing rubber formulations for medical applications.
Standards
The Rubber Sleeve Rebound Tester is designed to comply with the following standards:
(1) YY/T 1618-2018: Collection Needle Rubber Sleeve Resilience Test (China)
(2) DIN 53 512: Testing of Rubber – Rebound Resilience (Germany)
(3) ISO 4662: Rubber, Vulcanized or Thermoplastic – Determination of Rebound Resilience
(4) ASTM D7121: Standard Test Method for Rubber Compression Set
Parameters
| Parameter | Specification |
|---|---|
| Pressure test range | 0 – 50 N, error ≤ ±1%, resolution 0.01 N |
| Test time | 1 – 999 s, error ≤ ±0.5 s |
| Test speed | 50 mm/min, 80 mm/min, 100 mm/min, error ≤ ±2% |
| Measuring distance | 0 – 50 mm, error ≤ ±0.1 mm |
| Rebound force range | 0.5 – 1.5 N |
| Printing capability | 8. 10. or 13 test data points, including average and deviation |
| Machine dimensions | Customizable depending on model and fixture |
| Power supply | AC220 V, 50 Hz (standard) |
Accessories
(1) Main test unit – 1 set
(2) Needle clamping fixtures – 1 set
(3) Printer paper roll – 1 piece
(4) Operation manual – 1 set
(5) USB storage (optional)
Test Procedures
(1) Insert the needle at the butt end into the die hole of the clamping fixture, 2 mm above the surface.
(2) Start the test device to compress the rubber sleeve.
(3) Measure the rubber sleeve resilience at distances of 5 mm and 15 mm between the needle tube and the needle base.
(4) The instrument displays and records the rebound force. Each test can print 8. 10. or 13 data points, including average and deviation.
(5) Confirm that the measured resilience falls within the standard range of 0.5 N ≤ resilience ≤ 1.5 N.
Send product request
Other supplier products
| Photovoltaic module outdoor exposure test system | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Photovoltaic Module Outdoor Exposure Test System is designed to simulate real-world env... | |
| SUGA GX25 Xenon Lamp Weathering Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications Traditional Automotive: Brakes, steering wheels, manual transmissions, interi... | |
| Luggage Walked Wear Test Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com This product is used for suitcases, bags, strollers with wheels… Through the test mi... | |
| Incline Impact Tester,ISO 13355 angle impact tester for shipping materials | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Incline Impact Tester is designed to evaluate the impact resistance of products, packag... | |
| High-precision heat flow meter for insulation materials | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Measuring thermal conductivity and resistance of insulation materials (rigid a... |