Balloon Catheter Leak and Compliance Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
ISO 13988-1:2017: Intravascular catheters - Part 1
ISO 8536-4:2016: Medical infusion equipment
YY/T 0285.4-2017: Intravascular catheters - Part 4
YY/T 1551.1-2017: Non-intravascular catheters
Features
Color LCD large screen display: Real-time display of test data and statistical values, intuitive and clear.
Intelligent touch control operation: Simple human-machine interface for quick experiment switching.
High-precision electronic flow meter: Achieves precise flow control, avoiding testing errors.
Top-tier sensor system: High-sensitivity data acquisition ensures stable and consistent test results.
Data security management: Four-level access control and encrypted storage prevent unauthorized operations and ensure the integrity and traceability of results.
Convenient result output: Equipped with a micro printer, supporting instant printing of single test results and statistical data.
Technical Parameters
| Parameter | Specification |
|---|---|
| Pressure Range | Determined by test object & application scenarios:- e.g., 0 - 300kPa for medical breathing circuits- Narrower range (e.g., 0 - 50kPa) for routine compliance testing |
| Pressure Accuracy | Key for test accuracy; generally requires ±0.1% or higher (e.g., 0 - 50kPa pressure sensor can reach one-thousandth accuracy) |
| Flow Range | Covers various flow situations of tested catheters; common flowmeter range: 0 - 100L/min |
| Flow Accuracy | Important indicator; generally requires ±1% or higher (e.g., ±1%FS for 2L/min mass flow controller) |
| Volume Measurement Range and Accuracy | - Range: From a few milliliters to several liters (adapts to different catheter volumes)- Accuracy: Generally needs to reach ±0.1ml or higher (to accurately calculate catheter compliance) |
| Test Time Range | - Leakage tests: Flexible setting (a few seconds to several hours)- Compliance tests: Controls pressure change rate & time interval to capture catheter volume changes at different pressure stages |
Accessoriess
Standard accessories include: specialized connectors for different sizes of medical tubing (such as ventilator/anesthesia machine breathing circuits), pressure/flow test tubing, tubing fixing clamps (to ensure sample stability during testing), standard test samples for equipment calibration (including standard leakage and compliance components), operating manual, and calibration certificate.
Test Procedures
Sample Pretreatment
Tubing Aging Treatment: Immerse in physiological saline at 37°C for 24 hours to relieve material stress;
Connection Sealing Inspection: Apply soapy water to the interfaces to confirm no initial leakage.
Typical Test Modes
Static Compliance Test:
Pressurize to 100kPa, maintain pressure for 30 seconds, and record the volume change ΔV1;
Release pressure to 0kPa, and record the volume recovery ΔV2;
Compliance calculation: C = (ΔV1 - ΔV2)/ΔP (unit: mL/kPa).
Dynamic Leakage Test:
Apply a sinusoidal pressure wave with a frequency of 1Hz and an amplitude of ±50kPa for 10 minutes;
Collect the leaked fluid and calculate the leakage rate using the weighing method (≤0.5mL/10min is considered acceptable).
Result Evaluation Standards (Refer to YY/T 0285.4-2017)
Compliance Deviation: ≤±15% compared to the nominal value;
Leakage: ≤1mL/minute under positive pressure of 100kPa, and no visible bubbles under negative pressure of -50kPa.
Send product request
Other supplier products
| Leather edge oil fatigue testing machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Leather Edge Oil Fatigue Testing Machine is designed to evaluate the fatigue resistance... | |
| UV & Carbon Arc Light Exposure Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The Carbon Arc Aging Chamber employs alternating current to ignite carbon rod... | |
| Toxicity index tester manufacturer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Toxicity Index Tester is a specialized instrument designed for evaluating... | |
| Medical glove water-impermeability tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Applications The impermeability tester for medical gloves is mainly used to determine the ... | |
| Helmet fit height measurement system for EN 397 compliance certification | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Measure the wearing height of helmets to ensure the accuracy and... |
Same products
| Common Rail Injector Valve F00R J02 012 | Seller: 868983 | Common Rail Injector Valve F00R J02 012 David Chen | |
| Common Rail Injector Valve F00R J02 005 | Seller: 868983 | Common Rail Injector Valve F00R J02 005 David Chen | |
| Common Rail Injector Valve F00R J01 945 | Seller: 868983 | Common Rail Injector Valve F00R J01 945 David Chen | |
| Common Rail Injector Valve F00R J01 941 | Seller: 868983 | Common Rail Injector Valve F00R J01 941 David Chen | |
| Common Rail Injector Valve F00R J01 924 | Seller: 868983 | Common Rail Injector Valve F00R J01 924 David Chen |