Best Guide Wire Tip Softness Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
FDA-16007 FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires –Performance Tests and Recommended Labeling
ISO 11070:Cardiovascular implants and extracorporeal systems — Balloon dilatation catheters
YY 0450.1-2003:Disposable sterile intravascular catheters — Part 1: General requirements
YY/T 1554-2017:Medical devices — Surface lubricious coatings — Test method for coefficient of friction
Features
1. Intelligent Operation: A large LCD screen displays the testing process in real time. Built-in standard testing programs enable one-click operation and automatic completion of the experiment.
2. High-Precision Sensors: Utilizes advanced displacement sensors to ensure accurate speed and displacement transmission, making testing more scientific and reliable.
3. Secure and Compliant Design: Supports hierarchical access control and password protection to prevent unauthorized operations and ensure data traceability.
4. Complete Data Output: Equipped with a mini printer, it can directly print information on test personnel, testing time, and results, ensuring complete retention of experimental records.
5. Personalized Upgrades: The system program supports ISP online upgrades and can provide customized services based on user needs.
Technical Parameters
| Technical Parameter | Specification |
|---|---|
| Testing Speed | It can be adjusted within a certain range, such as 3mm/min or 8mm/min. Some testers can reach 100mm/s and can be set arbitrarily with stepless speed regulation, and the speed accuracy can reach 0.01mm/s. |
| Displacement Range and Resolution | The displacement range is generally 0.001 - 40.000mm, and the displacement resolution is 0.001mm. |
| Force - measuring Mechanism | Mostly built - in, using high - precision force sensors. |
| Control System | Commonly, there is a PLC (Programmable Logic Controller). The operation interface is generally a color touch screen, which can be switched between Chinese and English. |
| Clamping Device | It is used to fix the guidewire tip to ensure the stable position of the guidewire during the testing process. It is generally made of high - quality materials such as 304 stainless steel. |
Accessoriess
The package includes: 1 main unit, 1 set of fixtures, 1 set of testing software, 1 set of sensors, 1 power cord, 1 instruction manual, 1 operation video, and 1 certificate of conformity.
Test Procedures
1.Device Preparation
Check the instrument’s power supply, connection status, and external components (e.g., clamps, sensors) for integrity.
Power on the device and preheat it to the required working state (if specified by the instrument).
2.Sample Preparation
Take the guidewire to be tested, confirm its model/specification meets the test requirements, and clean the guidewire surface (to avoid contamination affecting test results).
3.Sample Installation
Fix the guidewire on the instrument’s clamping device (use 304 stainless steel clamps as mentioned), and accurately position the tip part of the guidewire at the test station (ensure stable fixation without affecting the tip’s flexibility).
4.Parameter Setting
Operate the color touch screen (PLC control system) to set test parameters (e.g., testing speed, displacement range, data storage quantity) according to the test standard (e.g., YY/T 1554-2017) or test requirements.
5.Initiate Testing
Start the test via the operation interface; the instrument will automatically apply force/bending action to the guidewire tip according to the set parameters.
6.Data Acquisition & Recording
The instrument will automatically collect and store test data (e.g., flexibility, bendability data) during the process (supports storing multiple groups like 200/500 groups).
7.Post-Test Operation
After the test is completed, stop the instrument, safely remove the tested guidewire, and check the integrity of the collected data.
8.Data Processing & Report Generation
Export the test data via the system, analyze the flexibility performance of the guidewire tip, and generate a test report (if the instrument supports report output).
Send product request
Other supplier products
| Automated Reaction Heat Meter | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB 51283 GB/T 22232 GB/T 17802 SN/T 3078.1 Product Features 1. It offers th... | |
| Small Parts Tester | Test Procedure: Usage Methods and Interpretations Applicable standards: ASTM F963 EN 71-1 16 CFR Part 1501 ISO 8124-1 Product Advantages: Eff... | |
| Laboratory PVT coolant pulse testing equipment | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The PVT Coolant Pulse Test Machine complies with a broad range of international ... | |
| Melt Mass-Flow Rate Test Machine, ASTM D1238-98 | Whats/App: +86 156 1874 6768 Web: standard-groups.com Introduction: This machine is deisgned according to ISO1133. ASTM D1238-98 ,ISO1133; Mel... | |
| High Frequency Reciprocating Rig | Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The High-Frequency Reciprocating Tester is used in standardized and research-l... |
Same products
| High precision continuous load surface tester | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for... | |
| Multi-function surface performance test system | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for... | |
| Laboratory tribology research equipment | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for... | |
| Tribology test system with adjustable load | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 6040-2008 Test Method for... | |
| Laboratory friction and wear tester with adjustable load | Seller: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 3960-2008 Metallic Materi... |