Benchtop Thermal Conductivity Tester for Insulation Materials

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Standards

The instrument complies with and references the following international and national standards:

ASTM D5470-2017

Standard Test Method for Thermal Transmission Properties of Thin Thermally Conductive Solid Electrical Insulation Materials

MIL-I-49456A

Insulation Sheet Materials, Thermally Conductive, Resin-Based and Glass Fiber Reinforced

GB/T 5598-2015

Method for Determination of Thermal Conductivity of Beryllium Oxide Ceramics

GB/T 29313-2017

Test Method for Thermal Conductive Properties of Electrical Insulating Materials

Parameters

ItemSpecification
Thermal conductivity range 0.01–50 W/m·K; 5–500 W/m·K (high-conductivity mode, automatic switching)
Thermal resistance range 0.05–500 cm²·K/W
Sample size (standard) Φ30 mm or 20 × 20 mm
Optional sample sizes Φ15 mm, Φ50 mm, 25.4 × 25.4 mm (customized by contract)
Sample thickness range 0.001–50 mm (typical: 0.02–20 mm)
Hot plate temperature range Room temperature to 99.99 °C (standard)
Optional hot plate ranges Room temperature to 299.9 °C / 500 °C
Cold plate temperature range 0–99.99 °C
Cold plate control accuracy 0.01 °C
Pressure range 0–1000 N
Pressure control Servo motor control, accuracy 0.1 N
Thickness measurement range 0–50.00 mm
Thickness resolution 0.001 mm
Number of samples 1 piece (thin films or stacked layers supported)
Measurement error (thermal conductivity & resistance) ≤3%
Measurement error (contact thermal resistance) ≤5%
Test operation Fully automatic computer-controlled testing
Power supply AC 220 V / 50 Hz
Power consumption 1000 W

Features

Fully designed and manufactured in accordance with ASTM D5470-2017

Benchtop configuration with the same testing performance as floor-standing models

Servo motor automatic loading with precise pressure control

Automatic thickness measurement for improved accuracy and repeatability

Six-point temperature gradient detection to enhance measurement precision

Thermal protection structure around the test rod to reduce environmental influence

Measurement of thermal conductivity, thermal resistance, and interfacial contact thermal resistance

Capability to generate thermal resistance curves under different pressures and temperatures

Optimized mathematical models for high-conductivity and interface testing

Automatic cold-end temperature compensation without ice-water compensation

Fully automated data acquisition, analysis, and report output

Accessories

Main testing unit

Thermal analysis software (Chinese and English versions)

HX-1008 precision low-temperature constant-temperature water bath (0.01 °C accuracy)

Computer (user-selectable)

Calibration reference samples (2 pieces)

Sample fixtures and accessories (solid sample molds, powder, paste, and grease sample cells)

Test Procedures

Select the appropriate test mode according to material type and test objective.

Place the solid sheet sample between the hot and cold plates.

Install sample frames or test cells when testing powder or paste materials.

Set test parameters, including temperature and pressure, via the software.

Apply pressure automatically using the servo-controlled loading system.

Measure sample thickness automatically before testing.

Record temperature gradients and heat flow data during steady-state conditions.

Calculate thermal conductivity, thermal resistance, and contact thermal resistance.

Save, print, and export test results and reports.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

Keyboard Key Life Tester
Keyboard Key Life Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com standards GB/T4967-95 silicone rubber and keypad life testing equivalent to JIS and IEC st...
Automated suture line diameter tester
Automated suture line diameter tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The medical suture thread diameter tester needs to comply with relevant medical ...
Battery Diaphragm Gas Permeability Tester
Battery Diaphragm Gas Permeability Tester The Battery Diaphragm Gas Permeability Tester is a precision instrument designed to evaluate the gas permeability of battery diaphragms and separat...
Rubber Low-Temperature Brittleness Evaluation Machine
Rubber Low-Temperature Brittleness Evaluation Machine Whats/App: +86 156 1874 6768 Web: standard-groups.com Product Introduction: Plastic low temperature embrittlement testing machine is to determin...
Photovoltaic module backplane peeling strength testing machine
Photovoltaic module backplane peeling strength testing machine The Photovoltaic Module Backplane Peeling Strength Testing Machine is a specialized device designed to evaluate the peeling strength of photovoltai...
All supplier products

Same products

Benchtop Thermal Conductivity Tester for Insulation Materials
Benchtop Thermal Conductivity Tester for Insulation Materials Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument complies with a...
High-Accuracy Guarded Heat Flow Meter Thermal Conductivity Measurement System
High-Accuracy Guarded Heat Flow Meter Thermal Conductivity Measurement System Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument is designed and...
Heat Flow Method for Thermal Conductivity Measurement
Heat Flow Method for Thermal Conductivity Measurement Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument complies with a...
Benchtop Thermal Conductivity Tester
Benchtop Thermal Conductivity Tester Seller: Standard International Group (HK) Limited Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The instrument is designed and...
Guarded Heat Flow Measurement System