Oxidation Induction Phase Analysis Tester
Get a quote:
Whats/App: +86 181 0160 0864
Web: gellowen.co.uk
Differential Scanning Calorimeter DSC is designed to determine the inner heat transition relating to temperature and heat flow, it is widely used in the field of polymer development, performance testing & quality control. DSC research and development includes the following field: glass transition temperature, melting point, cold crystallization, crystallization, phase transition, Enthalpy, oxidation induction time (OIT).
Application
It is used to determine the time of automatic catalytic oxidation of the sample at room temperature hydrogen, and to evaluate the thermal stability of the sample. It is suitable for plastics and rubber materials.
Standard
GB/T17391-1998:Polyolefin pipe systems for the conveyance of fluids under pressure - Test methods for the resistance to internal pressure
ISO/TR10837:1991:Technical Report on the use of plastics pipes and fittings for the conveyance of drinking water - Guidelines for the selection of materials and components
ASTM D3895-1998:Standard Test Method for Oxidative Induction Time of Polyolefins by Differential Scanning Calorimetry (DSC)
ASTM E 967:Standard Practice for Temperature Calibration of Thermocouples and Thermocouple Materials by Comparison Techniques
ASTM E 968:Standard Practice for Thermal Cycling of Electronic Components and Materials for Temperature Stress Screening
ASTM E 793:Standard Test Method for Measuring Surface Discontinuities on Metal Surfaces by Electromagnetic (Eddy-Current) Method
ASTM D 3895:Standard Test Method for Oxidative Induction Time of Polyolefins by Differential Scanning Calorimetry (DSC)
ASTM D 3417:Standard Test Method for Thermogravimetric Analysis (TGA) of Polymers
ASTM D 3418:Standard Test Method for Transition Temperatures and Enthalpies of Fusion and Crystallization of Polymers by Differential Scanning Calorimetry (DSC)
ISO 11357-6:Plastics - Differential scanning calorimetry (DSC) - Part 6: Determination of oxidation induction time (isothermal OIT) and oxidation induction temperature (dynamic OIT)
EN 728:Thermal insulating products for building applications - Determination of water vapour transmission properties
Features
The instrument has stable performance, high measurement accuracy and good repeatability;
Designed for measuring oxidation induction period in plastics and rubber industry:
* Integrated design, temperature control and furnace device are integrated, reducing signal loss and interference, beautiful and exquisite appearance;
The furnace body adopts the upper cover structure, no need to lift, convenient loading sample;
* High-resolution data acquisition system, real-time data acquisition;
Intelligent adaptive single-chip microprocessor temperature control system, temperature control speed, high precision:
* Perfect two-way air valve control system, using the original imported flow valve from the United States, can realize the automatic switching of ammonia to oxygen during the measurement process;
Professional oxidation induction period test software with automatic temperature correction and automatic test program to automatically calibrate temperature, test oxidation induction time and plot curves:
Simple operation method, to achieve one-click testing, test results can be saved and printed:
Computer real-time display instrument status, test data and curves, easy to view.
Parameters
| Description | Details |
|---|---|
| Maximum tensile stress | 20MPa |
| Number of stations | 30 (can be customized) |
| Water bath temperature range | RT - 90°C |
| Temperature control accuracy | ±1°C |
| Switch sensitivity | 200g |
| Timing accuracy | 1s |
| Sample thickness | 0.75 - 3mm |
| Notch depth | 20% of the nominal thickness of the sample (6 - specified as an example value, if relevant) |
| Power supply | AC 380V 50HZ |
| Overall dimensions | 1420x640x1900mm |
| Weight | 380kg |
Send product request
Other supplier products
| Kernel Test Instrument | Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Kernel Test Instrument (also known as the Koenen Test Apparatus) is a speciali... | |
| Computer controlled horizontal tensile testing machine | Product Introduction: The microcomputer controlled horizontal tensile testing machine (hereinafter referred to as the testing machine) is mainly u... | |
| Air - tightness Testing Device for Automotive Connectors | Get a quote: Whats/App: +86 181 0160 0864 TheAir-Tightness Testing Device for Automotive Connectorsis designed to perform non-destructive airtigh... | |
| Car seat fatigue tester | Get a quote: Whats/App: +86 181 0160 0864 The Car Seat Fatigue Tester is a dedicated durability testing system used to evaluate the long-term fat... | |
| Civil Aviation Aircraft Interior Non-metallic Materials Flame Retardant Property Tester | Get a quote: Whats/App: +86 181 0160 0864 Web: gellowen.co.uk The Civil Aviation Aircraft Interior Non-metallic Materials Flame Retardant Proper... |
Same products
| HDPE High Speed Water Cooled Double Wall Corrugated Pipe Extrusion Line | Seller: Shandong HUASU Plastic Technology Co.,Ltd. | The molds adopt the double flow channel structure and nitriding polishing processed die head. Thi... | |
| PanEVAP & MVR system | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | A multiple-effect evaporator is a device that efficiently uses steam heat to evaporate water. Wat... | |
| PanGAP wide gap welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanGAP wide gap welded plate heat exchangerconsists of a welded plate pack with different channel... | |
| PanBloc welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanBloc welded plate heat exchangeris compact and efficient, although fully welded, is totally ac... | |
| PanBox welded plate heat exchanger | Seller: Yangzhou Panstar Heat Exchanger Equipment Co., Ltd | PanBox welded plate heat exchangerconsists of a laser welded plate pack with optimized corrugated... |