ISO fabric pilling tester for sale laboratory QC machine

Get a quote:

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Features

Modular Test Chambers: 4 dedicated pilling boxes (octagonal) and 2 snagging boxes for comprehensive analysis.

Rapid Mounting Jig: Simplifies sample fixation with quick-release clamps.

Accessories Kit: Includes cork liners, snagging points, and a pilling photograph guide for visual grading.

Motor Safety System: Overload protection and thermal cutoff prevent equipment damage.

Customizable Cycles: Users define test duration and reversal intervals via the LCD interface.

Compact Footprint: Space-saving design with integrated storage for accessories.

Technical Parameters

Technical ParametersSpecifications
Test Mode Automatic
Rotation Direction Unidirectional / Bidirectional (Reversing)
Rotation Speed Options 20rpm, 30rpm, 40rpm, 45rpm, 50rpm, 60rpm, 65rpm, 70rpm
Standard Speed Modes 1. 60rpm Unidirectional
2. 30rpm Unidirectional
3. 30rpm Bidirectional
Working Chamber 4 Chambers
End Alarm Buzzer
Steel Nails Material Stainless Steel
Power Supply 220V, 50Hz
Weight 75 kg
Dimensions

900mm × 800mm × 1100mm

Test Procedures

1. Pre-operation Preparation

Environmental Adjustment: Adjust the test environment to standard atmospheric conditions for textiles (in accordance with ISO 105-A01/GB/T 6529): temperature 20±2℃, relative humidity 65±3%; condition the instrument, samples, and consumables (such as backing fabric) in this environment for at least 4 hours (if conditioning is not required for the test, follow the corresponding standard).

Instrument Inspection: Connect the appropriate power supply (e.g., AC 220V/50Hz), start the instrument without load, and confirm that the heating plate temperature stability, pressure device (weight) accuracy, and timing module function are all normal, without any abnormal noise or fault indications.

Consumables Preparation: Prepare the auxiliary materials according to the corresponding standards, such as ISO 105-F02 standard cotton backing fabric, anti-staining backing fabric, sample fixing clamps, etc.

2. Sample Preparation

Cut the samples according to the test item (ensure no wrinkles, damage, or stains):

Heat Pressing/Sublimation Color Fastness Test: Cut into 100mm×40mm rectangles, taking one representative sample each from the warp and weft directions;

Bleached Fabric Yellowing Test: Cut into 150mm×150mm squares;

Reactive Dye Water Fastness Test: Cut into 100mm×40mm, and after wetting according to GB/T 5718 requirements, squeeze out excess water to the specified moisture content. 3. Test Parameter Settings

Set the parameters on the control panel according to the target standard (e.g., ISO 105 X11. GB/T 5718):

Mode Selection: Corresponding to the test type (e.g., "Dry Heat Sublimation," "Heat Press Resistance," "Yellowing Test");

Temperature: Set according to the standard (e.g., sublimation color fastness 180±2℃, heat press resistance 150±2℃, yellowing test 180±2℃);

Pressure: Set to 4 kPa for heat press resistance testing (matching the weight of the standard weight in ISO 105 X11), no additional pressure is required for sublimation/yellowing tests;

Time: Set according to the standard (e.g., sublimation 30±1s, heat press resistance 15±0.5s, yellowing test 10±0.5min).

4. Sample Loading and Test Operation

Heat Press Resistance Test: Place the sample between two layers of standard cotton lining fabric, lay it flat on the center of the heating plate, close the press head to apply pressure, and start the timing program;

Sublimation Color Fastness Test: Fix the sample and standard lining fabric separately on the sample holder and lining fabric holder (maintaining the standard distance), close the test chamber door, and start the dry heat program;

Yellowing Test: Lay the sample flat on the heating plate (without lining fabric), ensure sufficient contact, and then start the dry heat timing;

Water Staining Fastness Test: Stack the moistened sample and lining fabric, load according to the heat press resistance procedure, adjust to the corresponding water staining temperature, and then start the test.

5. Post-processing and Result Evaluation

After the test, wait for the instrument to cool to room temperature, and then remove the sample and lining fabric;

Use the corresponding gray scale (e.g., ISO 105-A02 for evaluating sample discoloration, ISO 105-A03 for evaluating lining fabric staining), and rate the results on a scale of 1-5 (5 being the best);

Record the sample information, test standard, parameters (temperature/pressure/time), and rating.

6. Instrument Cleaning and Maintenance

After turning off the power and allowing the instrument to cool, wipe the heating plate, press head, and sample holder with a dry, soft cloth to remove any residual fibers;

Calibrate the temperature and pressure modules according to the calibration schedule (e.g., every 6 months) to ensure that the testing accuracy meets the standards.


Send product request

To: Standard International Group (HK) Limited
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

Cloth Material Comfort Evaluation Device
Cloth Material Comfort Evaluation Device Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The instrument is widely used in: Textile R&D: Quantifying fabric hand fe...
Alternative to PV Hail Impact Tester
Alternative to PV Hail Impact Tester Whats/App: +86 156 1874 6768 Web: standard-groups.com Application The Hail Impact Test Machine is suitable for a variety of testing and research...
Multi function Vacuum Tribometer
Multi function Vacuum Tribometer Whats/App: +86 156 1874 6768 Web: standard-groups.com Application This tester is designed for friction and wear evaluation in vacuum or reduced-...
Battery combustion tester
Battery combustion tester Whats/App: +86 156 1874 6768 Web: standard-groups.com The Battery Combustion Tester is designed for safety evaluation of lithium-ion batteries by...
Textile Adhesion Testing Machine Supplier
Textile Adhesion Testing Machine Supplier Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Fabric Wet Adhesion Performance Tester is a precision instrument designed...
All supplier products

Same products

MMA-120 Handheld Arc Welder
MMA-120 Handheld Arc Welder Seller: 871183 The MMA-120 Handheld Arc Welder from ChuangLi Electronic Technology Co., Ltd. is designed for con...
SHX Series Circulating Fluidized Bed Steam Boiler
SHX Series Circulating Fluidized Bed Steam Boiler Seller: Hebei Genjux Boiler Co.,Ltd Capacity: 10-35 t/hWorking pressure: 1.25-2.5 MPaOperating temperature: 194-400 ℃Applicable to bi...
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler
YQ(Y)W Series Oil/Gas Fired Vertical Thermal Oil Boiler Seller: Hebei Genjux Boiler Co.,Ltd Capacity: 7000-29200 KWOperating pressure: 0.8-1.0 MPaOperating temperature: 320℃Operating mode: ...
DHL Series Coal-Fired Steam Boilers
DHL Series Coal-Fired Steam Boilers Seller: Hebei Genjux Boiler Co.,Ltd Capacity: 35-75 t/hOperating pressure: 1.25-5.4 MPaOperating temperature: 194-485℃Applicable to b...
SZL Series Biomass Steam Boiler
SZL Series Biomass Steam Boiler Seller: Hebei Genjux Boiler Co.,Ltd Capacity: 6-35 t/hWorking pressure: 1.0-2.5 MPaOperating temperature: 184-350 ℃Applicable fuel bi...