Digital Twelve Channels ECG RSD912A cheap
NAME Digital Twelve Channels ECG
STYLE NO RSD912A
DESCRIPTIONS ●Twelve Channel digital ECG machine with Interpretive;
●Self-test upon powering on;
●Simultaneous 12 lead acquisition;12-channel, 6-channel
printing form at choice; Date, time and parameters, include
ng paper speed,heart rate sensitivity, filter status and etc;
●Designed with a high-resolution thermal printer to print out ECG waveform, annotation as well as related parameters for diagnosis reference;
●Recording paper for ECG-912A is 210mm in width;
●Visual display of operation menu, ECG waveform and parameters;
●Digital isolation technology is adopted to minimize time/temperature drift of components so as to promote adaptability of the unit to circumstance ;
●Digital filter for effective inhibition of baseline drift, interference, as well as automatic position of baseline;
●Function of rhythm lead for observing abnormal ECG waveform & heart rate;
●Automatic measurement and interpretation of ECG waveform to relieve doctors from strain;
●Multiple print formats include 6-channels and 12-channels;
●AC/Rechargeable battery for continuous examination when AC power supply is unavailable. For battery operation, ECG-912A is equipped with a charger & system for battery capacity management and protection itself from excessive discharge;
●Safety level for ECG-912A corresponds to Class I, type CF.
SPECIFICATION
Lead Standard 12 leads CMMR >60dB
Acquisition Simultaneous 12 leads >100dB(with AC filter)
Resolution 18 bit/1000Hz Filter AC filter, EMG filter
Operating Mode Automatic or manual at choice (with measurement report ) Drift filter Anti-Drift System
Patient Leakage Current 10A
Input Circuit Floating input circuit; Protection against defibrillation and pace making Threshold ≤20V
Recorder High resolution thermal printer
Paper Specification Folding210x140x20mm
Input Circuit Current Isolated ,<=0.1uA Rolling 210X30mm
LCD display Display menu, operation status, heart rate(digit and words)
Sampling Rate 1ms
Input Impedance >50MQ
Calibration Voltage 1mV±3% Safety level IEC60601-1 class I , type CF
Sensitivity Auto,2.5,5, 10, 20 mm/mV Power Supply AC:220V/110V+-10%;50/60Hz+-1Hz,30VA ; DC: 14.4V(1600mAH );
rechargeable Ni-MH battery
Paper speed 25,50mm/s (±3%)
Time Constant > 3.2 s
Noise Level <15uVp-p Dimension and weight 400mm*325mm*84.5mm
Frequency Response 0.05-150Hz NW 6kg
ACCESSORIES
Main unit 1 PC; Patient cable 1 PC; Limb electrode 4 PCS; Roll ECG paper 1 roll; Three plugged power cable 1 PCS
Chest electrode 6 PCS; Backup AC fuse 2 PCS; Operating instructions 1 Copy; Earth connecting line 1 PC
Other supplier products