Medical Suture Line Diameter Tester
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
Quality Control: The medical suture thread diameter tester plays a crucial role in the quality control of medical devices. By accurately measuring the diameter of suture threads, it ensures that products meet relevant medical standards and specification requirements, thereby improving the quality and safety of medical devices. Manufacturer Inspection: During the production process, manufacturers need to conduct strict quality inspections on each batch of suture threads. The thread diameter tester can monitor the changes in thread diameter in real - time, promptly detect and correct production deviations, and improve production efficiency and product quality. Third - Party Inspection: Independent third - party testing laboratories are responsible for conducting random quality inspections and verifications on suture threads sold in the market. The high - precision data provided by the thread diameter tester can serve as a reliable basis for test reports, ensuring the safety and effectiveness of products during use. Product R & D: In the process of developing new suture thread products, R & D personnel need to repeatedly verify and adjust product specifications. The thread diameter tester can provide accurate thread diameter measurement data, helping R & D personnel optimize the design and improve product performance.
Standards
The medical suture thread diameter tester needs to comply with relevant medical industry standards, such as YY 1116 - 2020Absorbable Surgical Suture and YY 0167 - 2020 Non - absorbable Surgical Suture. These standards specify in detail the requirements for the technical parameters, testing methods, and measurement accuracy of the suture thread diameter tester, ensuring the accuracy and reliability of measurement results.
Features
1. Compliance
Execution Standards: Designed and manufactured according to the wire diameter testing clauses in YY1116-2020 "Absorbable Surgical Sutures" and YY0167-2020 "Non-absorbable Surgical Sutures".
Allows for measurement at multiple positions, including 1/4. 1/2. and 3/4 of the length; for multi-strand sutures with specifications greater than metric 2. the suture can be rotated 90° for measurement without disassembly.
2. Stability
Uses a grating displacement sensor instead of a micrometer, providing accuracy to one ten-thousandth and excellent stability.
3. Convenience
Manual sample loading, electric tensioning, and multi-point automatic measurement.
To address the issue of changing weights for each specification, it innovatively uses a high-precision force sensor to automatically set the tension force, making operation simple.
Equipped with quick mechanical buttons for testing and stopping, allowing for one-button testing and pausing.
4. Intelligence
A 7-inch color touch screen combined with intelligent data processing functions provides users with a comfortable and smooth operating experience.
Touch screen operation: Nominal specification, test load, printing, testing, left movement, right movement, time, calibration.
Technical Parameters
| Serial No. | Parameter Item | Parameter Details |
|---|---|---|
| 1 | Minimum Division Value | Not exceeding 0.002mm (some high-precision devices can reach 0.001mm), ensuring precise suture thread diameter measurement |
| 2 | Base Diameter | Usually greater than 50mm, providing stable basic support for measurement |
| 3 | Pressing Plate Diameter | 12.70mm±0.02mm, used to apply pressure to the suture thread for diameter measurement |
| 4 | Presser Foot Load | Varies by suture thread specifications:- For specifications > nominal size 0.4 (or 8-0): 210g±3g- For specifications ≤ the above size: ≤60g |
| 5 | Calibration Difference and Parallelism Difference | Not greater than 0.005mm (between pressing plate surface and base), ensuring measurement accuracy |
Accessoriess
One main unit, one certificate of conformity, accompanying fixtures, one instruction manual, and one roll of printing paper.
Send product request
Other supplier products
| Iso 6356 static electrical propensity | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com The Carpet Static Electricity Tester is a precision instrument designed to me... | |
| Spinal Internal Fixation Components System Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standard YY/T 0961-2014 (China): Specifies static/fatigue testing protocols for spinal fix... | |
| OEM Walk-in Climate Test Chamber Factory Direct | Temperature & Humidity Simulation Solutions | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards GB/T 2423.1: Environmental testing methods for electrical and elec... | |
| Dual-Purpose Manual & Automatic Tear Tester | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com G008 Manual Tearing Tester, The Elmendorf tear method can be used to test the... | |
| Paper roughness and air permeability tester for ISO paper quality testing | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards Surface Roughness: ISO 8791-2 Paper and paperboard - Determinatio... |
Same products
| YGD03 Stainless steel electric operation table | Seller: Jiangsu Yigao Medical Equipment Technology Co., Ltd. | YGD03 Electric Operating TableProfessional Multi‑Purpose Surgical Table for Operating RoomsⅠ. Pro... | |
| Prophy Brush(hard/soft) | Seller: Huanghua Promisee Dental Co.,Ltd | Prophy Brush(hard/soft) Product Description Prophy Brushis composed of a metal handle and a wor... | |
| Micro Applicator(Ultrathin) | Seller: Huanghua Promisee Dental Co.,Ltd | Micro Applicator(Ultrathin) Product Description Micro applicator bulk pack supplieroffering 100... | |
| Dental Retractor(M type) | Seller: Huanghua Promisee Dental Co.,Ltd | Dental Retractor(M type) Product Description ThisDental Retractor(M type)is M type design. In o... | |
| Dental Dam Kit | Seller: Huanghua Promisee Dental Co.,Ltd | Dental Dam Kit Product Description The Dental Dam Kit, a type of disposable dental supplies, is... |