HYGK-307 Circuit breaker tester

Circuit breaker analyzer introduction:

1) Measurement of closing time, opening time, time spread between poles, time spread between units of one pole, bounces etc.
2) Measurement of closing speed, opening speed and average speed.
3) Measurement of clearance between open contacts, over travel, overshoot travel and total travel.
4) Measurement of closing coil current time curve, opening coil current time curve and the maximum coil current
5) Test travel time curve, speed time curve and acceleration time curve
6) Measurement of pre-insertion resistance value and pre-insertion ion time.
7) Test contact resistance
8) Measurement of dynamic resistance time curve and length for arcing contacts
9) Vibration test, including vibration fingerprint curve acquisition, management, and vibration fingerprint curve automatic matching.
10) CO-t1-CO,O-t-CO-t1-CO,CO rated operating sequence verification test.
11) Drawing Envelopes for reference travel curve and compare travel curve with envelopes
12) Mechanical travel curve automatic compare with saved travel curve
13) Measurement of CO time (close-open time) and OC time (Opening-Closing time)
14) Graphite contact closing time, opening time, dynamic resistance curve measurement
15) Zoom, editors, statistics and analysis for measurement curves.
16)12.1-inch touch screen for control and data input
17) Built-in micro printer for test result printout.
18) Generate WORD format test report automatically (all reports and data can be exported through Flash disk).

Circuit breaker analyzer technical index:

1) 12 channels for time measurement (3-channel supports both the time contact and resistance contact time, 6-channel isolated contacts for cascaded unit measurement of pole)
2) Time measurement
Range: 4000ms
Error :< 0.1ms
Resolution: 0.01ms
3) Travel measurement according to sensor:
Linear resistance sensor:
Measurement range: 0~250mm; resolution: 0.01mm; error: <0.5mm
Measurement range: 0~25mm; resolution: 0.01mm; error: <0.05mm
Angle resistance sensor:
Measurement range: 0~360°; resolution: 0.01°; error: <0.5°
4) Pre-insertion resistance measurement
Range: 50~5000ohm; error :<1%RDG+2D
5) Dynamic resistance measurement
Test current: 25~100A
Measurement error :< 1%RDG+2D
6) Contact resistance measurement
Range:0--10mohm
Error <0.5%RDG+0.05%FS
Range: 0-2mohm;
Error: <0.5%RDG+2D
7) Speed measurement
250mm linear resistance scale: 0~20m/s; error: <0.5%RDG+2D
25mm linear resistance scale: 0~20m/s; error: <0.5%RDG+2D
Angle sensor: 0~20m/s; error: <0.5%RDG+2D
Acceleration sensor: 0~20m/s; error: <5%RDG+2D
8) Coil current measurement
Range: 0~20A; resolution: 0.001A; error: <0.01A
9) Vibration measurement
Range: 0~5000G; error: <3%
10) Internal DC power source: voltage: 12~265V; current:0~20A
11) Built-in 8G storage memory and 2 USB interface for data export and external keyboard or mouse connection
12) Power supply: AC220V±10%; 50Hz±10%
13)Working environment: temperature: -10~50°C; Humidity: <80%

Email

Miss Chloe Wang



Send product request

To: Wuhan Huaying Electric Power Tech & Science Co.,ltd
Your E-mail:
Message text:


Send to other suppliers

Other supplier products

HYLB-601 instantaneous signal recorder
HYLB-601 instantaneous signal recorder Introduction: HYLB-601 is an instantaneous signal recorder applied in industry signal analysis. It takes IPC as the core for data sample and proces...
HYPCIT Primary current injection tester
HYPCIT Primary current injection tester Introduction: HYPCIT series primary current injection tester is applied in power system for circuit breaker test, protection system test and curre...
HYBC-901 Transformer Ratio Meter
HYBC-901 Transformer Ratio Meter Transformer ratio meter Itistotestvoltageratio,vector,polarity,ratioerrorofcurrenttransformerandvoltagetransformer. Transformer turns ratio met...
HYBL-II MOA tester
HYBL-II MOA tester MOA arrester tester introduction 1 Large-screen LCD Display, English user menu, Easy to use.2 Use Precision Sampling and Fourier harmonic anal...
SLQ Primary Current Injection Test Set
SLQ Primary Current Injection Test Set Introduction: SLQ primary current injection test setis a high current generator instrument. It is applied in current transformer test, circuit brea...
All supplier products

Same products

YCS Microscope Swing Arm Bracket 144 LED Ring Light Source
YCS Microscope Swing Arm Bracket 144 LED Ring Light Source Seller: Phonefix YCS Adjustable 144 LED Ring Lamp and YCS Golden Dragon Arm 40cm 360° Rotation Microscope Swin...
Pneumatic fabric stiffness tester
Pneumatic fabric stiffness tester Seller: Standard International Group (HK) Limited The bending performance of the fabric is an important physical parameter that we study the proper...
Curl elasticity detector
Curl elasticity detector Seller: Standard International Group (HK) Limited Crimp elasticity detector is mainly used to determine the curl, curl elasticity, curl recovery ra...
Fiber thermal shrinkage tester
Fiber thermal shrinkage tester Seller: Standard International Group (HK) Limited Intelligent fiber heat shrinkable tester used to measure a variety of chemical fibers and natural...
Fiber impurity Tester
Fiber impurity Tester Seller: Standard International Group (HK) Limited Fiber impurity Tester is a special instrument for measuring the impurity content of fiber. It is ...