General Ventilation Test Bench
The General Ventilation Test Bench is designed to evaluate the performance of general ventilation air filters. Based on ISO 16890-2016, the system measures filtration efficiency across particle sizes from 0.3 μm to 10 μm, using both oily and saline aerosols. Coarse aerosols (artificial test dust) are employed to determine dust holding capacity and final life efficiency. The test bench ensures accurate classification of air filters according to particulate matter efficiency (ePM).
Application
Testing general ventilation air filters in HVAC systems.
Evaluating filter efficiency across PM1, PM2.5, and PM10 ranges.
Determining filter resistance and airflow characteristics.
Measuring dust holding capacity and end-of-life performance.
Laboratory quality control for filter production and certification.
Research and development of ventilation and filtration products.
Standards
ISO 16890-2016 – Air filters for general ventilation.
Applicable international particulate measurement and testing protocols.
Features
Key components from imported brands ensure accurate and stable test data.
Pneumatic sealing fixtures allow fast and secure filter installation.
Rotating pipeline structure simplifies filter placement.
Fluid-dynamics-based pipeline design ensures realistic airflow simulation.
Supports testing with both oily and saline aerosols for comprehensive evaluation.
Capable of measuring dust holding capacity and generating flow-resistance curves.
User-friendly interface with real-time data acquisition.
Technical Parameters
| Parameter | Specification |
|---|---|
| Flow Range | 600–5400 m³/h |
| Measurement Error | ±2% or less |
| Resistance Range | 0–500 Pa |
| Accuracy of Measuring Device | 0–70 Pa ≤ ±2 Pa; >70 Pa ≤ ±3% |
| Aerosol Homogeneity | ≤ ±15% |
| Initial Efficiency Repeatability | ±30–40% |
| Particle Concentration Indication Error | 40–60% ≤ ±15%; 60–80% ≤ ±8%; 80–90% ≤ ±4%; 90–99% ≤ ±2% |
| Efficiency Range | ePM1 ≤ 99% (Comprehensive efficiency) |
| Power Consumption | 18 kW |
| Main Pipeline | Square, 610 × 610 mm, 304 stainless steel |
| Floor Area | 13,000 × 2,500 × 2,500 mm (L × W × H) |
| Filtration Efficiency | 1. Filter resistance 2. DEHS counting efficiency (PM1) 3. KCl counting efficiency (PM2.5, PM10) 4. Weighing efficiency |
| Filtration Resistance | Flow-resistance curve |
| Dust Holding Capacity | Final life test of the filter |
| Counter | PALAS Promo2000, sampling flow rate: 5 L/min ±5% |
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Helmet Projection and Surface Friction tester | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The helmet projection and surface friction tester is a device used to test the fri... | |
| Blow-By Test Rig | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Applications determination of oil content in blow-by on the engine test stand ev... | |
| Textile Drying Time Tester ISO 11092 | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Textile Drying Time Tester is a precision instrument designed to measure evapo... | |
| GB/T 17650 Halogen Acid Gas Release Test Apparatus for Electric Cables | ASTM ISO Testing Machine | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com This apparatus is widely used in laboratories and quality control departments invo... | |
| Automotive Wire Marking Durability Tester | ASTM ISO Testing Machine | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Automotive Wire Marking Durability Tester is designed to evaluate the durabili... |