Advanced heat insulation test furnace for fireproof coatings

Whats/App: +86 156 1874 6768

Web: standard-groups.com

Application

This test furnace is suitable for determining:

1. Thermal insulation efficiency of steel structure fireproof coatings

2. Fire resistance limit of coated steel components

3. Back-surface temperature rise under standard fire exposure

4. Heating and furnace-pressure stability during fire resistance tests

It is widely used in:

5. Fire protection coating manufacturers

6. Building material testing laboratories

7. National and third-party inspection institutes

8. Research universities and engineering research centers

9. Fireproof material certification and quality evaluation agencies

Standards

This instrument is developed and manufactured in accordance with the following standards:

Primary Standards

GB 14907-2018 – Fire-Resistant Coatings for Steel Structures

GB/T 9978-2008 – Fire-Resistance Tests for Building Construction Elements

International Standards

Although the furnace is primarily based on Chinese national requirements, its fire-curve control and measurement systems align with major global fire-test methodologies, including:

ISO 834 – Standard Fire Resistance Test Curve

EN 13381 Series – Test methods for determining the contribution of fire protection materials to structural elements

ASTM E119 – Standard Test Methods for Fire Tests of Building Construction and Materials

BS 476 (Part 20–24) – Fire tests on building materials and structures

This compatibility allows the furnace to support laboratories that reference or compare international fire-test systems.

Parameters

ItemSpecification
Working Voltage AC 220V ± 10%, 50 Hz
Power Consumption 3 kW
Furnace Pressure Range 0–100 Pa
Pressure Control Accuracy 20 ± 3 Pa
Airflow Range 0–50 m³/min
Airflow Accuracy ≤ ±0.5 m³/min
Fuel Type 0# or −10# diesel
Standard Fire-Curve Temperature Program 60 min: RT–925°C; 120 min: RT–1050°C; 180 min: RT–1115°C; 240 min: RT–1150°C; 300 min: RT–1180°C; 360 min: RT–1210°C
Furnace Temperature Sensors 4 pcs K-type armored thermocouples
Back-Surface Temperature Sensors 2 pcs, accuracy ±4°C
Temperature Control Mode Average or independent control
Heating Mode Fully automatic diesel spray burner
Cooling Mode Automatic exhaust and cooling
Timing Range 0–360 min
Data Storage Capacity ≥100 sets
Sample Size 500 × 500 × 6 mm
Overall Dimensions 1.6 m × 1.6 m × 1.6 m

Test Procedures

1. Cut the sample using a 250mm × 90mm template. If the sample is a melt-shrinkable product, use a 250mm × 180mm template.

2. Clamp the sample in a 250mm × 90mm sample holder, tighten the wing nut, and secure the sample. The sample holder size for melt-shrinkable products is 250mm × 180mm. At this point, do not yet mount the sample holder onto the horizontal rod.

3. Turn on the power switch and set the ignition time to a slightly longer duration, such as 5 minutes. Select the appropriate positioning device according to the ignition method. After pressing "Start Test" and the burner stops at the 45° position, insert one end of the positioning device into the burner. Now mount the sample holder onto the horizontal rod and adjust the sample holder's up, down, left, and right positions to meet the test requirements. Remove the positioning device. Press the return key. After the burner resets, press "Test Reset".

4. Press and hold the ignition button, and simultaneously adjust the flame adjustment valve on the instrument panel. After igniting the gas, adjust the flame height to 20 ± 1mm. When adjusting the flame from large to small, it is easier to first tighten the flame adjustment knob on the panel. The height gauge should be at 20mm. Close the front door.

5. Place two pieces of filter paper in the collection tray and then place it under the sample.

6. Set the ignition time, press "Start Test," and the burner will automatically rotate to the 45° position and remain there for the set ignition time, then reset. The afterglow time starts counting simultaneously. If the sample stops afterglow before reaching the marked line, press "Stop Afterglow." If the sample afterglow reaches the marked line, immediately press "Stop Afterglow." Record the burning time.

7. Press "Test Reset" to stop the test. To continue the test, repeat the above steps; otherwise, turn off the power. Provided by Shandong Provincial Textile Science Research Institute.


Отправить запрос, связаться с поставщиком

Кому: Standard International Group (HK) Limited
Ваш E-mail:
Текст письма:


Send to other suppliers

Другие товары поставщика

 Direcct flame impingement Tester
Direcct flame impingement Tester Introduction The testerisdesignedaccording to the BS476-12:1991.Method of test for ignitability of products by direct flame impingement.The accu...
Walk In Pharmaceutical Stability Chamber
Walk In Pharmaceutical Stability Chamber Applications Pharmaceutical Stability Testing Laboratory is designed for conducting accelerated humidity and heat testing, long-term stability tes...
Diaper impurity testing machine
Diaper impurity testing machine The Diaper Impurity Testing Machine is designed for visual inspection of disposable diapers under controlled lighting conditions to detect residual...
Battery Electrode Resistance Tester
Battery Electrode Resistance Tester Standard ASTM B193: Standard Test Method for Resistivity of Electrical Conductor Materials ISO 18516: Conductive Coatings – Measurement o...
Residual Ethylene Oxide Gas Chromatography System
Residual Ethylene Oxide Gas Chromatography System The Residual Ethylene Oxide Gas Chromatography System is a professional analytical instrument designed for the detection and quantification of resi...
Все товары поставщика

Похожие товары

Tabletop Xenon Arc Weathering Tester (SunMaster)
Tabletop Xenon Arc Weathering Tester (SunMaster) Продавец: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Water-Cooled Xenon Lamp Weat...
Blow-By Test Rig
Blow-By Test Rig Продавец: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test system for the characterisa...
Air filter System(ISO 16890/EN 779)
Air filter System(ISO 16890/EN 779) Продавец: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Filter test system for general a...
Oil Thread Leak Test System
Oil Thread Leak Test System Продавец: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com Test System for oil thread leak ...
Manual HEPA/ULPA Filter Scanning Test System
Manual HEPA/ULPA Filter Scanning Test System Продавец: Standard Groups Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com The Manual HEPA/ULPA Filter Scan...