Automatic catheter leakage and compliance apparatus
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
The catheter leakage and compliance tester is mainly used to detect the leakage and compliance of medical catheters (such as the breathing circuits of ventilators and anesthesia machines). Specifically, it can: Leakage Test: By applying a certain pressure inside the catheter and monitoring the rate or extent of pressure drop, it determines whether there is leakage in the catheter. This is crucial for ensuring the sealing and safety of the catheter during medical procedures. Compliance Test: Compliance refers to the ability of the catheter to deform when subjected to external forces. The tester applies a certain pressure inside the catheter and measures the volume change of the catheter to evaluate its compliance. This is of great significance for understanding the flexibility and elasticity of the catheter, as well as its applicability in medical applications.
Standards
ISO 13988-1:2017: Intravascular catheters - Part 1
ISO 8536-4:2016: Medical infusion equipment
YY/T 0285.4-2017: Intravascular catheters - Part 4
YY/T 1551.1-2017: Non-intravascular catheters
Technical Parameters
| Parameter | Specification |
|---|---|
| Pressure Range | Determined by test object & application scenarios:- e.g., 0 - 300kPa for medical breathing circuits- Narrower range (e.g., 0 - 50kPa) for routine compliance testing |
| Pressure Accuracy | Key for test accuracy; generally requires ±0.1% or higher (e.g., 0 - 50kPa pressure sensor can reach one-thousandth accuracy) |
| Flow Range | Covers various flow situations of tested catheters; common flowmeter range: 0 - 100L/min |
| Flow Accuracy | Important indicator; generally requires ±1% or higher (e.g., ±1%FS for 2L/min mass flow controller) |
| Volume Measurement Range and Accuracy | - Range: From a few milliliters to several liters (adapts to different catheter volumes)- Accuracy: Generally needs to reach ±0.1ml or higher (to accurately calculate catheter compliance) |
| Test Time Range | - Leakage tests: Flexible setting (a few seconds to several hours)- Compliance tests: Controls pressure change rate & time interval to capture catheter volume changes at different pressure stages |
Accessoriess
Standard accessories include: specialized connectors for different sizes of medical tubing (such as ventilator/anesthesia machine breathing circuits), pressure/flow test tubing, tubing fixing clamps (to ensure sample stability during testing), standard test samples for equipment calibration (including standard leakage and compliance components), operating manual, and calibration certificate.
Test Procedures
Sample Pretreatment
Tubing Aging Treatment: Immerse in physiological saline at 37°C for 24 hours to relieve material stress;
Connection Sealing Inspection: Apply soapy water to the interfaces to confirm no initial leakage.
Typical Test Modes
Static Compliance Test:
Pressurize to 100kPa, maintain pressure for 30 seconds, and record the volume change ΔV1;
Release pressure to 0kPa, and record the volume recovery ΔV2;
Compliance calculation: C = (ΔV1 - ΔV2)/ΔP (unit: mL/kPa).
Dynamic Leakage Test:
Apply a sinusoidal pressure wave with a frequency of 1Hz and an amplitude of ±50kPa for 10 minutes;
Collect the leaked fluid and calculate the leakage rate using the weighing method (≤0.5mL/10min is considered acceptable).
Result Evaluation Standards (Refer to YY/T 0285.4-2017)
Compliance Deviation: ≤±15% compared to the nominal value;
Leakage: ≤1mL/minute under positive pressure of 100kPa, and no visible bubbles under negative pressure of -50kPa.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Fabric puncture damage analysis instrument | Whats/App: +86 156 1874 6768 Web: standard-groups.com Compliance Standards ASTM D638 (Plastics Tensile Properties) ISO 13934-1 (Textile Tensile... | |
| Gloves Cut Resistance Testing Machine | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Gloves Cut Resistance Testing Machine is designed for evaluating the cutting resistance... | |
| ASTM D573 Geosynthetic Chemical Aging Analyzer | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application Test method for acid-resistant lye properties of various geosynt... | |
| Motor Air Tightness Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com The Motor Air Tightness Tester is designed to evaluate the sealing performance of electric ... | |
| Solids Spontaneously Igniting Tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Conforms to the UN Recommendations on the Transport of Dangerous Goods - Manual of Tests an... |
Похожие товары
| Laser Cutting Machines | Продавец: Chanxan (Changshu) Laser Technology Co., Ltd. | Laser Cutting Machines PrecisionLaser Cutting Systemsfor Film, Glass & Industrial Materials ... | |
| Precision Laser Micromachining Machine | Продавец: Chanxan (Changshu) Laser Technology Co., Ltd. | Precision Laser Micromachining Machine| Ultrafast Laser Micromachining Systems Chanxan provides ... | |
| Ultrafast Laser Systems | Продавец: Chanxan (Changshu) Laser Technology Co., Ltd. | Ultrafast Laser Systems Femtosecond Laser System Ultra-precision femtosecond laser systemfor hi... | |
| Industrial Laser Equipment Manufacturer | Продавец: Chanxan (Changshu) Laser Technology Co., Ltd. | Industrial Laser Equipment Manufacturer for Precision Cutting, Marking & Micromachining Chan... | |
| Common Rail Injector Valve F00R J02 012 | Продавец: 868983 | Common Rail Injector Valve F00R J02 012 David Chen |