Adjustable Catheter Hydraulic Leakage Tester device
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Applications
The catheter hydrostatic pressure resistance and leakage tester is a device specifically designed to evaluate the sealing performance of catheters in a hydraulic environment. Its main uses include: Ensuring Catheter Safety: Through testing, it ensures that catheters have good sealing performance under various usage conditions, preventing liquid leakage and thus safeguarding the safety of patients. Improving Product Quality: It provides data support for medical device manufacturers, helping them to improve the design and manufacturing processes of catheters, and enhancing the quality and reliability of products. Promoting Technological Advancement: The application of the tester promotes technological progress and development in the medical field, and provides a powerful testing tool for the research and development of new - type catheters.
Standards
GB/T 15812.1: Disposable sterile catheters - Part 1: General requirements
Technical Parameters
| Parameter | Specification |
|---|---|
| Test Flow | Usually 30L/min, with an error ≤ ±5%; ensures stable and accurate flow during testing |
| Pressure Range | Generally 0 - 500KPa, with a resolution of 0.01KPa and an error ≤ ±1%; covers the pressure range most catheters withstand in actual use (ensures test comprehensiveness) |
| Measurement System Holding Time | Usually 1 minute (configurable); maintains a stable measurement environment during testing (improves test accuracy) |
| Test Duration | Generally 1 second - 99.9 minutes, with an error ≤ ±1 second; allows flexible setting of test time according to actual needs (ensures test adequacy and effectiveness) |
| Applicable Catheter Types | Applicable to various medical device products: infusion sets, blood transfusion sets, infusion needles, filters, pipelines, catheters, quick-connect fittings, etc. (meets testing needs of different products) |
Accessoriess
Main unit, hydraulic tank, miniature printer
Test Procedures
1. Pre-test preparation
Check the equipment power supply and hydraulic system sealing, and prepare the test tubing and adapter fittings;
2. Sample installation
Connect the tubing to the equipment's hydraulic circuit using a special connector, ensuring a sealed connection;
3. Parameter setting
Select the corresponding standard (e.g., GB/T 15812.1) via the touchscreen, and set parameters such as test flow rate, pressure range, and pressure holding time;
4. Start testing
The equipment automatically fills with fluid and pressurizes, maintaining the set parameters and monitoring for leaks;
5. Process monitoring
Observe pressure changes to confirm whether there are any leakage abnormalities;
6. Completion
After the equipment depressurizes, remove the tubing, record the results, and clean the pipeline.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| MCD scratch hardness tester for industrial metals | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards (1) GB/T 23989-2009 – Determination of Scratch Resistance of Coatings (2)... | |
| High Temperature Creep Testing Equipment ISO 204 | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Application (1) Tensile creep and endurance testing of metallic materials at... | |
| Bottle base and side wall thickness tester | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The design and manufacture of the wall and bottom thickness tester for bottles c... | |
| Carpet Electrostatic Decay Time Tester IEC 61340 Factory Direct Sale | Get a quote: Whats/App: +86 156 1874 6768 Web: standard-groups.com Features Accurately measures human-body electrostatic potential under standa... | |
| Shoe heat insulation testing machine -- sand bath | Whats/App: +86 156 1874 6768 Web: standard-groups.com The sand bath component of the shoe heat insulation testing machine is primarily used to te... |