Laboratory Fire Smoke and Toxic Gas Test Equipment
Whats/App: +86 156 1874 6768
Web: standard-groups.com
Standards
The system complies with the following standards and test methods:
(1) GB/T 19666 – General Rules for Flame-Retardant and Fire-Resistant Wires, Cables or Optical Cables
(2) EN 50305 – Railway Applications – Test Methods for Fire Behaviour of Electric Cables
(3) BS 6853 Annex B – Code of Practice for Fire Precautions in the Design and Construction of Passenger Carrying Trains
(4) ISO 19700 – Controlled Equivalence Ratio Method for the Determination of Toxic Product Yields in Fire Effluents (Steady-State Tube Furnace Method)
(5)ASTM E1678
(6)NFPA 269
Parameters
| Item | Specification |
|---|---|
| Tube furnace heating length | 600 mm |
| Temperature range | 0–1100 °C |
| Temperature accuracy | ±0.1 °C |
| Quartz tube | Heat-resistant quartz, thickness ≥ 2 mm |
| SO₂ measurement range | 0–1000 ppm |
| SO₂ resolution | 1 ppm |
| SO₂ accuracy | ±30 ppm |
| NOₓ measurement range | 0–3000 ppm |
| NOₓ resolution | 1 ppm |
| NOₓ accuracy | ±30 ppm |
| Gas sampling bag | PTFE-coated, 80 L capacity, with imported manual sampler |
| CO measurement range | 0–10 % |
| CO resolution | 0.01 % |
| CO accuracy | Absolute ±0.1 % or relative ±3 % |
| CO₂ measurement range | 0–20 % |
| CO₂ resolution | 0.01 % |
| CO₂ accuracy | Absolute ±0.2 % or relative ±3 % |
| Spectral analysis device | Spectrometer at 500 nm wavelength, membrane thickness 10 mm |
| Absorption solution | 0.1 N NaOH |
| Titration reagents | Picric acid aqueous solution (3 g/dm³) and sodium carbonate aqueous solution (50 g/dm³) |
Features
Sealed chamber design with the smoke generation system installed internally, providing stable operation and effective protection of core components
Large observation window allowing clear visual monitoring of test status and specimen movement
Integrated industrial design with panel-mounted structure, offering intuitive operation and ergonomic user interface
Tube furnace protected by perforated outer housing to enhance heat dissipation and ensure uniform temperature rise to the target value
High-temperature K-type thermocouple for accurate measurement of specimen zone temperature
Ceramic combustion boat driven by strong magnetic coupling and transmission mechanism for stable and repeatable specimen delivery into the furnace core
Precision lead-screw transmission system enabling accurate positioning and reliable sealing of the quartz tube, reducing the risk of damage due to manual operation
Dual filtration columns filled with activated carbon and color-changing silica gel for effective carrier gas purification and drying
Branded computer with dedicated control software for automatic test process control
Software-based temperature regulation and stabilization with real-time curve recording, automatic data display, storage, and report generation
Accessories
(1) Heat-resistant quartz tube
(2) Ceramic combustion boat
(3) PTFE-coated gas sampling bag (80 L)
(4) Activated carbon filter column
(5) Color-changing silica gel filter column
(6) K-type thermocouple
(7) Dedicated control software and industrial computer
Test Procedures
1. Pre-Test Preparation
Ensure the tester is installed in a well-ventilated laboratory environment. Verify that the tube furnace, quartz tube, and filtration system are clean and properly assembled. Check that the absorption solution (0.1 N NaOH) and titration reagents are prepared according to specified concentrations. Confirm normal operation of the temperature control system, gas analysis instruments, and data acquisition software.
2. Specimen Loading
Place the prepared test specimen into the ceramic combustion boat. Using the magnetic drive and transmission mechanism, steadily convey the boat into the center of the tube furnace. The lead-screw positioning system ensures accurate alignment and reliable sealing at the quartz tube interface, minimizing gas leakage during the test.
3. Heating and Smoke Generation
Set the target temperature according to the selected standard. The furnace temperature is automatically controlled by the software, with uniform heating achieved through regulated power output. As the specimen undergoes thermal decomposition or combustion, the generated smoke is carried by a controlled gas flow toward the sampling and absorption system.
4. Gas Collection and Analysis
The combustion gases are either introduced into the NaOH absorption solution via bubbling or collected in PTFE-coated gas sampling bags for subsequent analysis. Online infrared gas analyzers continuously measure CO and CO₂ concentrations, while SO₂ and NOₓ are determined using dedicated measurement channels or chemical analysis methods. All test data are automatically recorded and stored by the control software.
Maintenance Information
Regularly inspect the quartz tube, ceramic combustion boat, and furnace interior for residue buildup and clean them as required. Check the condition of the activated carbon and silica gel in the filter columns and replace them when saturation or discoloration is observed. Verify calibration of gas analyzers and temperature sensors according to laboratory quality procedures. Ensure the lead-screw transmission and magnetic drive mechanisms operate smoothly to maintain positioning accuracy.
Отправить запрос, связаться с поставщиком
Другие товары поставщика
| Taber Linear Abraser | Whats/App: +86 156 1874 6768 Web: standard-groups.com TheTABER Linear Abraser (Abrader) - Model 5750has been engineered to evaluate product&rsquo... | |
| Standard Helmet Visual Field Tester | The Standard Helmet Visual Field Tester is a precision instrument designed to evaluate the peripheral vision and visor obstruction of safety helmet... | |
| Photovoltaic module damp heat test chamber | The Photovoltaic Module Damp Heat Test Chamber is designed to simulate long-term tropical aging conditions for photovoltaic modules and related ele... | |
| Drawer slide durability meter | Product Introduction Drawer slides are fundamental components in furniture, cabinetry, and various storage solutions, providing smooth and reliabl... | |
| Digital display electronic ink viscosity meter | How it works: The working principle of the digital display electronic ink viscosity meter is based on two tin (usually a metal roller and a rubber... |
Похожие товары
| ISO 29463 HEPA/ULPA filter performance testing System | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system supports and compli... | |
| High-Precision ISO 16890 Filter Media Test Rig for R&D and QC | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system can execute and com... | |
| Mobile Floor Scrubber Aging Life Tester IEC 60312 | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com TheMobile Floor Scrubber Aging L... | |
| ISO 12103 Cleanable Filter Media Test System | Продавец: Standard International Group (HK) Limited | Whats/App: +86 156 1874 6768 Web: standard-groups.com Standards The system complies with: (1)... | |
| Robotic Vacuum Cleaners Edge-Following Tester | Продавец: Standard Groups | Get a quote: Whats/App: +86 185 0176 3615 Web: qinsun-lab.com TheEdge-Following Test Equipment... |